Site icon Pharma Dekho

Analytical method validation

 

Analytical method validation

Purpose
1.1 To describe a procedure for the validation of analytical methods.

1.0 Scope
2.1 This standard operating procedure is applicable for validation of analytical methods for Drug

products in the QC laboratory at ABC Pvt. Ltd., Roorkee (UK), India.

2.0 Responsibility
3.1 Officer QC/ Executive QC Preparation of SOP and follow the procedure as per SOP
Assistant Manager Reviewing and Training of SOP
Manager QC Implementation of SOP

3.0 Accountability
3.1 Manager Quality Control is accountable for implementation and compliance with SOP.

4.0 Procedure
4.1 Analytical methods used for the determination of assay, determination of related substances shall be validated.
4.2 Analytical methods shall be validated for the following parameters.
4.3 For assay of Active Pharmaceutical Ingredients (API) & Drug Products
4.3.1.1 Linearity
4.3.1.2 Accuracy
4.3.1.3 Precision– System, Repeatability & Intermediate Precision
4.3.1.4 Specificity
4.3.1.5 Robustness
4.3.2 For related substances of Active Pharmaceutical Ingredients (API) & Drug Products
4.3.2.1 Linearity
4.3.2.2 Accuracy
4.3.2.3 Precision – System, Repeatability & Intermediate Precision
4.3.2.4 Specificity
4.3.2.5 Limit of Detection
4.3.2.6 Limit of Quantification
4.3.2.7 Robustness
4.3.3 For Dissolution and drug release of the Drug products:
4.3.3.1 Linearity
4.3.3.2 Specificity
4.3.3.3 Precision – System, Repeatability & Intermediate Precision
4.3.3.4 Robustness
4.3.4 Revalidation shall be done if there is change in Master formula of drug product,

or any other change in the analytical method.
4.3.5 The ‘Analytical Method Validation Protocol’ shall address the following parameters:

4.3.6 Linearity
4.3.6.1 The linearity of an analytical method is its ability to elicit test results that are directly,

or by a well defined mathematical transformation proportional to the concentration

of analyte in sample in given range.
4.3.6.2 Linearity shall be established across the range above and below at minimum five

concentration level, of the analytical procedure. The linearity shall be established by the

examination of plot of signal verses the concentration of analyte. The plot should be a

linear line with regression co-efficient not less than 0.99.

4.3.7 Precision
4.3.7.1 Precision of the analytical is the degree of agreement among the individual test results

when the method is applied repeatedly to a multiple sampling of a homogeneous sample.

The precision shall be expressed as the related standard deviation of a series of measurements.

4.3.8 System Precision
4.3.8.1 The system precision shall be established for reproducible performance of instruments.

4.3.9 Repeatability
4.3.9.1 Make 09 independent sample solutions of a pooled sample at three different

levels (50%, 100% and 150%) and analyse by the method. Calculate the assay for each analysis and mean & RSD.

4.3.10 Intermediate Precision (Ruggedness)
4.3.10.1 Analysis of a 03 independent sample solutions (50%, 100% and 150%) of a pooled

sample by this method in triplicate by two different Analysts on different days. Calculate the

assay for each analysis. Calculate the mean and RSD.

4.3.11 Accuracy
4.3.11.1 The accuracy of an analytical method is the closeness of test results obtained by that

method to the true value. Accuracy shall be calculated as the percentage of recovery by the

assay of known added amount of analyte in the sample. The accuracy of method shall be

accessed by minimum of nine determinations over minimum of three concentration levels,

covering the specified range (i.e. three concentrations and three replicates of each concentration.

4.3.12 Specificity
4.3.12.1 Specificity is defined as the ability to access unequivocally the analyte in the

presence of components that may be expected present, such as impurities,

degradation products and matrix components.
4.3.12.2 The specificity of an analytical method is determined by comparing test results

from the analysis of samples containing impurities, degradation products, or placebo

ingredients with those obtained from the analysis of samples without impurities,

degradation products, or placebo ingredients. When impurities or degradation

products are unidentified or unavailable, specificity may be demonstrated by analysis

by the method in question of samples containing impurities or degradation product

and comparing the results to those from additional purity assay. Specificity of an analytical

method shall not be established if it is already established in analytical Development department.

4.3.13 Limit of Detection
4.3.13.1 The determination of limit of detection of an analytical method shall depend on

the method i.e. instrumental or instrumental procedure. For instrumental procedure

Signal to noise ratio shall be measured. The signal to ration 3:1 shall be considered as the limit of detection.
4.3.13.2 For non-instrumental methods limit of detection is generally determined by

analysis of samples with known concentration of analyte and by establishing a minimum

level at which the analyte can be reliably detected.

4.3.14 Limit of Quantification
4.3.14.1 The determination of limit of quantification of an analytical method shall vary

with the method i.e. instrumental or non-instrumental procedure. For instrumental

procedure measure the magnitude of analytical background response by analysing No.

of blank samples and calculating the standard deviation of the response. The standard

deviation multiplied by the factor 10 provides an estimate of limit of quantification.

The limit is subsequently validating by analyzing a suitable no. of samples known to be

near or prepared at the limit of quantification.
4.3.14.2 For non instrumental methods limit of quantification is generally determined by

analysis of samples with known concentration of analyte and by establishing a minimum

level at which the analyte can be reliably detected with acceptable accuracy and precision.

4.3.15 Robustness
4.3.15.1 As appropriate, make following changes in the experimental conditions — Variation

in mobile phase composition or flow rate.
4.3.15.2 Under each of the above variation of mobile phase composition or flow rate, perform

analysis of sample. Calculate the assay, and cumulative RSD of each condition with precision results.

 

5.0 Reference
ICH Guideline Q2 R2/ USP

6.0 Attachments
Not Applicable

7.0 Distribution
Distribution of the controlled copy to concerned departments.
Additional copies may be issued for specific purpose.

8.0 Abbreviations
QA : Quality Assurance
QC : Quality Control
SOP : Standard Operating Procedure

9.0 Revision History

Revision No. Supersedes Effective Date Details of Revision
01 Nil New SOP

 

Operating Procedure of Centrifuge

sop for preparation and storage of reagent solution

sop for Calibration and Maintenance of Laboratory Instruments and Equipment

Disposal of Residual Sample or Left Over Material

sop for for Laboratory Incident

standard operating procedure temperature monitoring

sop for drying of silica gel

sop for operation of infrared moisture balance

sop for preparation of mobile phase

sop for treated RO water

sop for Preparation and Issuance of Analysis protocol standard

sop of placebo and impurity stock solutions

sop for disposal of residual sample  

sop for handling of pharmacopoeial changes

sop for procedure for operation of ultrasonic cleaner

difference between UPLC and HPLC

sop for for Emergency Eyewash and Shower

sop for operation and calibration of total organic carbon analyzers

sop for operation of cobb tester

sop for Operation and calibration of atomic absorption spectrophotometer

sop for Operation and calibration of gas liquid chromatograph

sop for operation of humidity oven

sop for operation and calibration of serological water bath

sop for monitoring of drain trap

sop for destruction of analytical samples after testing and control samples

sop for destruction of used chemicals

Sop for Operation of suction pump

sop for Operation and calibration uv cabinet

sop for Operation and calibration of bulk density apparatus

sop for operation and calibration of shore hardness tester

sop for operation of rub proofness tester

sop for monitoring of purified water

sop for Retesting of packaging materials

sop for Retesting and resampling of raw materials

sop for Control of issuance of record of analysis green sheets

sop for Control of computer passwords

sop for sampling of packaging materials PM

Operation and Calibration Procedure for pH Meter

operation and calibration of muffle furnace

Operation of magnetic stirrer

Operation and cleaning of centrifuge

Operation and cleaning of sonicator

Operation and calibration of dissolution test apparatus

Operation and cleaning of hot air oven

Operation and calibration of bursting strength

Operating procedure of refrigerator

Operating procedure of stop watch

 

Exit mobile version