cleaning of glassware
1.0 Purpose
1.1 To describe a procedure for Cleaning of Glassware’s.
2.0 Scope
2.1 This standard operating procedure is applicable for cleaning of glassware’s in Quality Control laboratory at ABC Pvt. Ltd.
3.0 Responsibility
Officer QC/ Executive QC Preparation of SOP
Assistant Manager Reviewing and Training of SOP
Manager QC Implementation of SOP
4.0 Accountability
4.1 Manager Quality Control is accountable for implementation and compliance with SOP.
5.0 Procedure
5.1 Cleaning of General Glassware
5.1.1 After completion of Analysis, put all the used Glassware in “To Be Clean Glassware” Tray.
5.1.2 The glassware shall be cleaned immediately after completion of analysis. If the cleaning is not possible
immediately soak the glassware in water.
5.1.3 Remove the contents from the glassware carefully. The corrosive chemicals shall be first diluted in water
and then discarded in a flow of water.
5.1.4 Rinse the glassware thoroughly with tap water before cleaning glassware.
5.1.5 Dip the glassware in solution of 0.5% Labolin or 5% Extran in water for 10-15 minutes. Scrub with
nylon brush under the flow of tap water.
5.1.6 Wash the glassware with tap water till it free from foam.
5.1.7 Finally rinse the glassware with purified water.
5.1.8 After washing, glassware shall be dried in glassware drying oven at 50±5°C before use.
5.1.9 During drying glass shall be placed in oven by keeping all the glassware in an inverted position and
allowing the water to come out.
5.2 Cleaning of pipette
5.2.1 After use the pipette rinse immediately with tap water to prevent the breaking the tip not hit in sink or water tap.
5.2.2 After rinse and follow the step 5.1.5 to 5.1.9.
5.3 Cleaning of HPCL vials
5.3.1 After completion of Analysis, put out the vials from the HPLC, and drain the solution from vials in running water.
5.3.2 Rinse the vials, and septum’s with tap water.
5.3.3 Dip the vials and septum`s in solution of 0.5% Labolin or 5% Extran in 1 liter glass beaker for 30 minutes.
5.3.4 Wash the vials and septum`s with tap water till it free from foam.
5.3.5 Boil the vials in purified water for 1 hour.
5.3.6 Finally rinse the Vials and septum`s with purified water.
5.3.7 After washing Vials and septum`s shall be dried in drying oven at 50±5°C before use.
5.4 Whenever required, use Chromic acid mixture for cleaning as follows.
5.4.1 Chromic acid mixture preparation: Dissolve 200 gms of Potassium dichromate in 100 ml of
water-cooled in an ice bath. To this add very slowly 1500 ml of sulphuric acid with stirring and mix thoroughly.
Note: Chromic acid is highly corrosive wear rubber gloves, goggles during preparation.
5.4.2 Add a little quantity of chromic acid mixture in glassware and shake than wash thoroughly
with plenty of tap water and then follow from step 5.1.5 to 5.1.9.
5.4.3 New glassware shall be soaked for 1 to 2 hours in 1% v/v hydrochloric acid before washing as per step no. 5.1.5 to 5.1.9.
5.4.4 Stick stains such as grease etc. can be cleaned by 10% sodium carbonate solution or with acetone.
Never use alkalis for cleaning.
5.4.5 Allow to Cool Down the Temperature of Clean and Dried Glassware Before use
6.0 Reference
In House
7.0 Attachments
Not Applicable
8.0 Distribution
Distribution of the controlled copy to concerned departments.
Additional copies may be issued for specific purpose.
9.0 Abbreviations
QA : Quality Assurance
QC : Quality Control
SOP : Standard Operating Procedure
% : Percentage
v/v : Volume per Volume
10.0 Revision History
| Revision No. | Supersedes | Effective Date | Details of reason |
| 01 | New SOP |
Operating Procedure of Centrifuge
sop for preparation and storage of reagent solution
sop for Calibration and Maintenance of Laboratory Instruments and Equipment
Disposal of Residual Sample or Left Over Material
sop for for Laboratory Incident
standard operating procedure temperature monitoring
sop for operation of infrared moisture balance
sop for preparation of mobile phase
sop for Preparation and Issuance of Analysis protocol standard
sop of placebo and impurity stock solutions
sop for disposal of residual sample
sop for handling of pharmacopoeial changes
sop for procedure for operation of ultrasonic cleaner
difference between UPLC and HPLC
sop for for Emergency Eyewash and Shower
sop for operation and calibration of total organic carbon analyzers
sop for operation of cobb tester
sop for Operation and calibration of atomic absorption spectrophotometer
sop for Operation and calibration of gas liquid chromatograph
sop for operation of humidity oven
sop for operation and calibration of serological water bath
sop for monitoring of drain trap
sop for destruction of analytical samples after testing and control samples
sop for destruction of used chemicals
Sop for Operation of suction pump
sop for Operation and calibration uv cabinet
sop for Operation and calibration of bulk density apparatus
sop for operation and calibration of shore hardness tester
sop for operation of rub proofness tester
sop for monitoring of purified water
sop for Retesting of packaging materials
sop for Retesting and resampling of raw materials
sop for Control of issuance of record of analysis green sheets
sop for Control of computer passwords
sop for sampling of packaging materials PM
Operation and Calibration Procedure for pH Meter
Calibration of Melting Point Apparatus
Procedure For HPLC column washing procedure
Procedure For Retest Period of Raw Material
Procedure For Training Programme For Quality Control Personnel
Standard Operating Procedure is to provide guidelines for Preparation
Procedure For Analysis of Water rinsing
Procedure For Calibration of Glassware
Procedure For Procurement and storage of reference standards
Procedure For Preparation of buffer solution
safety guidelines for laboratory
responsibilities of quality control and microbiological department
responsibilities of quality control and microbiological department