hplc column performance evaluation storage and destruction
1.0 Purpose
1.1 To describe a procedure for evaluation of HPLC column performance, storage and discard.
2.0 Scope
2.1 This standard operating procedure is applicable for evaluation of HPLC column performance, storage
and discard in Quality Control laboratory at ABC Ltd, India.
3.0 Responsibility
Officer QC/ Executive QC Preparation of SOP and follow the procedure as per SOP
Assistant Manager Reviewing and Training of SOP
Manager QC Implementation of SOP
4.0 Accountability
4.1 Manager Quality Control is accountable for implementation and compliance with SOP.
5.0 Procedure
5.1 All the HPLC column shall be procured from standard manufacturers.
5.2 At the time of receiving new HPLC column, check the pack of column physically as per indent
and certificate of efficiency of column certified by the manufacturer shall be checked and record by the Manager QC.
5.3 HPLC column performance shall be checked for every new C8 and C18 columns before
starting their actual usage for analysis.
5.4 Set following chromatographic conditions:
Mobile phase : Acetonitrile : Water (65 : 35)
Flow rate : 1.0 ml /min
Detection Wavelength : 254 nm
Column temperature : 40o C
5.5 Solution Preparation and Qualification
5.5.1 Benzene and Toluene: transfer 0.5 ml each of benzene and toluene to a 100 ml volumetric
flask, add 25 ml of Acetonitrile and dilute to volume with mobile phase (Solution A).
5.5.2 Napthalene: Transfer about 30 mg of Napthalene to a 100 ml volumetric flask, add 25 ml of
Acetonitrile, allow to dissolve and dilute to volume with mobile phase (Solution B).
5.5.3 Pipette out 5 ml and 10 ml of solution A and B respectively in a 50 ml volumetric flask and
dilute up to volume with mobile phase. (Solution C).
5.5.4 Inject 20 µl solutions C in five replicates and record the chromatogram.
5.5.5 Compare the chromatogram with that supplied along with the column.
5.5.6 Determine the tailing factor for the entire peak. The mean tailing factor for any component shall not be more than 1.5.
5.5.7 Determine the plate count for naphthalene peak. This shall be around 10,000 USP plates.
5.5.8 Determine the resolution between benzene and toluene peaks shall be NLT 2.0.
5.5.9 Calculate the mean and RSD of peak areas for the five replicate injections. The RSD for the
peak areas for each component shall not be more than 2.0%.
5.5.10 Record all the results in HPLC column performance report (Annexure I).
5.5.11 In case the column doesn’t meets as per established criteria, such columns shall not be used for analysis.
5.5.12 If column performance is not satisfactory by above procedure, the same may be tested using
the manufacturers recommended procedures (if different from the one given above). The matter
shall also be discussed with the column supplier accordingly.
5.6 Cleaning and storage
5.6.1 After completion of the analysis the column shall be cleaned as per the following procedure.
5.6.2 if any buffer solution use in mobile phase wash the column initially with water (HPLC grade)
at flow 1.0 ml/min. for one hour, than wash with methanol: water (50:50) or acetonitrile: water (50:50) for one hour.
5.6.3 If any buffer solution was not use in mobile phase wash the column methanol: water (50:50) or
acetonitrile: water (50:50) for one hour at flow 1.0 ml/min. for one hour.
5.6.4 Now remove the column and then close the column end properly store at specifies place.
5.7 Destruction of HPLC column
5.7.1 All the non-performing columns of HPLC (Not meeting the system suitability
requirements e.g. USP plates count, Tailing factor and Resolution), which shall be confirmed
and approved by the Quality Control-Manager, shall be segregated for destruction purpose Annexure-II.
5.7.2 After due permission from Quality Control-Manager, take out packing of the column,
deface it and cut it into three pieces before sending it to scrap yard for destruction.
6.0 Reference
In House Specification
7.0 Attachments
Annexure I : HPLC Column Performance Record
Annexure II : Destruction Record of HPLC column
8.0 Distribution
Distribution of the controlled copy to concerned departments.
Additional copies may be issued for specific purpose.
9.0 Abbreviations
QA : Quality Assurance
QC : Quality Control
SOP : Standard Operating Procedure
HPLC : High Performance Liquid Chromatography
10.0 Revision History
| Revision No. | Supersedes | Effective Date | Details of reason |
| 01 | Nil | New SOP |
Operating Procedure of Centrifuge
sop for preparation and storage of reagent solution
sop for Calibration and Maintenance of Laboratory Instruments and Equipment
Disposal of Residual Sample or Left Over Material
sop for for Laboratory Incident
standard operating procedure temperature monitoring
sop for operation of infrared moisture balance
sop for preparation of mobile phase
sop for Preparation and Issuance of Analysis protocol standard
sop of placebo and impurity stock solutions
sop for disposal of residual sample
sop for handling of pharmacopoeial changes
sop for procedure for operation of ultrasonic cleaner
difference between UPLC and HPLC
sop for for Emergency Eyewash and Shower
sop for operation and calibration of total organic carbon analyzers
sop for operation of cobb tester
sop for Operation and calibration of atomic absorption spectrophotometer
sop for Operation and calibration of gas liquid chromatograph
sop for operation of humidity oven
sop for operation and calibration of serological water bath
sop for monitoring of drain trap
sop for destruction of analytical samples after testing and control samples
sop for destruction of used chemicals
Sop for Operation of suction pump
sop for Operation and calibration uv cabinet
sop for Operation and calibration of bulk density apparatus
sop for operation and calibration of shore hardness tester
sop for operation of rub proofness tester
sop for monitoring of purified water
sop for Retesting of packaging materials
sop for Retesting and resampling of raw materials
sop for Control of issuance of record of analysis green sheets
sop for Control of computer passwords
sop for sampling of packaging materials PM
Operation and Calibration Procedure for pH Meter
Calibration of Melting Point Apparatus
Procedure For HPLC column washing procedure
Procedure For Retest Period of Raw Material
Procedure For Training Programme For Quality Control Personnel
Standard Operating Procedure is to provide guidelines for Preparation
Procedure For Analysis of Water rinsing
Procedure For Calibration of Glassware
Procedure For Procurement and storage of reference standards
Procedure For Preparation of buffer solution
safety guidelines for laboratory
responsibilities of quality control and microbiological department
responsibilities of quality control and microbiological department
handling of corrosive hazardous and flammable chemical
calibration of volumetric glassware
operation and cleaning of water bath
testing and approval rejection of bulk finish and packed drug products
procedure for preparation of general test procedures and standard test procedures