Operation and calibration of fuming chamber

 

 

Operation and calibration of fuming chamber

 

1.0 Purpose
1.1 To describe a procedure for operation and cleaning of fuming chamber.

2.0 Scope
2.1 This standard operating procedure is applicable for operation and cleaning of Fuming Chamber located

at Chemical Analysis Room in Quality Control laboratory at ABC Pvt. Ltd.

3.0 Responsibility
Officer QC/ Executive QC Preparation of SOP and follow the procedure as per SOP
Assistant Manager Reviewing and Training of SOP
Manager QC Implementation of SOP

4.0 Accountability
4.1 Manager Quality Control is accountable for implementation and compliance with SOP.

5.0 Procedure
5.1 Operation
5.1.1 Ensure that fuming chamber is clean and no other activity is under operation in chamber.
5.1.2 Open the glass door for the work by sliding up the door.
5.1.3 Push the green switch on the exhaust of the fuming chamber.
5.1.4 Connect the instrument (Muffle furnace etc.) with main power supply and others during the analysis.
5.1.5 Carry out the test as per standard test procedure in the chamber.
5.1.6 Fumigating solvent e.g. hydrochloric acid sulphuric acid and nitric acid on sand tray inside the fuming chamber.
5.1.7 Remove the test sample from chamber after analysis and close the front glass door.
5.1.8 After completion of work clean the chamber carefully.
5.1.9 Push the red switch off the exhaust of the fuming chamber.

5.2 Cleaning of Fuming Chamber
5.2.1 Proper functioning of fuming chamber is essential to maintain laboratory, pollution free operation.
5.2.2 Clean the exterior and top of the fume chamber, first by sweeping the top and dusting with cloth.
5.2.3 Shut off power supply from mains.
5.2.4 Clean the exhaust ducting with brush.
5.2.5 Clean the tube light and surface of fuming chamber with a piece of cloth.
5.2.6 Check the performance of these following point
5.2.6.1 Starter of exhaust fan
5.2.6.2 Performance of exhaust fan
5.2.6.3 Performance of tube light change if required
5.3 If the paint on interior surface has become dull or having stains, arrange for re-painting.

5.4 Calibration once in year
5.4.1 Fuming chamber calibrated by outside external calibration agencies.

5.5 Precaution
5.5.1 Ensure that there are no inflammable chemicals present in the chamber.
5.5.2 Clean the fume chamber after disconnecting all electrical connection.
5.5.3 Use strong Acid carefully in chamber.
5.5.4 Air exhaust should be on at the working time in fume chamber.

6.0 Reference
In House Specification

7.0 Attachments
Nil

8.0 Distribution
Distribution of the controlled copy to concerned departments.
Additional copies may be issued for specific purpose.

9.0 Abbreviations
QA : Quality Assurance
QC : Quality Control
SOP : Standard Operating Procedure
oC : Degree Centigrade
10.0 Revision History

Revision No. Supersedes Effective Date Details of Revision
01 Nil New SOP

 

Operating Procedure of Centrifuge

sop for preparation and storage of reagent solution

sop for Calibration and Maintenance of Laboratory Instruments and Equipment

Disposal of Residual Sample or Left Over Material

sop for for Laboratory Incident

standard operating procedure temperature monitoring

sop for drying of silica gel

sop for operation of infrared moisture balance

sop for preparation of mobile phase

sop for treated RO water

sop for Preparation and Issuance of Analysis protocol standard

sop of placebo and impurity stock solutions

sop for disposal of residual sample  

sop for handling of pharmacopoeial changes

sop for procedure for operation of ultrasonic cleaner

difference between UPLC and HPLC

sop for for Emergency Eyewash and Shower

sop for operation and calibration of total organic carbon analyzers

sop for operation of cobb tester

sop for Operation and calibration of atomic absorption spectrophotometer

sop for Operation and calibration of gas liquid chromatograph

sop for operation of humidity oven

sop for operation and calibration of serological water bath

sop for monitoring of drain trap

sop for destruction of analytical samples after testing and control samples

sop for destruction of used chemicals

Sop for Operation of suction pump

sop for Operation and calibration uv cabinet

sop for Operation and calibration of bulk density apparatus

sop for operation and calibration of shore hardness tester

sop for operation of rub proofness tester

sop for monitoring of purified water

sop for Retesting of packaging materials

sop for Retesting and resampling of raw materials

sop for Control of issuance of record of analysis green sheets

sop for Control of computer passwords

sop for sampling of packaging materials PM

Operation and Calibration Procedure for pH Meter

Calibration of Melting Point Apparatus

Procedure For HPLC column washing procedure

Procedure For Retest Period of Raw Material

Procedure For Training Programme For Quality Control Personnel

Standard Operating Procedure is to provide guidelines for Preparation

Procedure For Analysis of Water rinsing

Procedure For Calibration of Glassware

Procedure For Procurement and storage of reference standards

Procedure For Preparation of buffer solution

safety guidelines for laboratory

responsibilities of quality control and microbiological department

responsibilities of quality control and microbiological department

handling of corrosive hazardous and flammable chemical

cleaning of glassware

calibration of volumetric glassware

operation and cleaning of water bath

sampling of raw materials

testing and approval rejection of bulk finish and packed drug products

procedure for preparation of general test procedures and standard test procedures

disposal of samples chemicals glassware columns

hplc column performance evaluation storage and destruction

operation and calibration of uv cabinet

operation and calibration of refractometer

operation and calibration of digital micrometer

handling of analytical weight box

operation and calibration of melting point apparatus

Operation and calibration of fuming chamber

Exit mobile version