Operation and calibration of polarimeter
1.0 Purpose
1.1 To describe a procedure for operation and calibration of Polarimeter.
2.0 Scope
2.1 This standard operating procedure is applicable for operation and calibration
of Polarimeter located at Chemical Room in Quality Control laboratory at ABC Pvt. Ltd., Roorkee .
3.0 Responsibility
Officer QC/ Executive QC Preparation of SOP and follow the procedure as per SOP
Assistant Manager Reviewing and Training of SOP
Manager QC Implementation of SOP
4.0 Accountability
4.1 Manager Quality Control is accountable for implementation and compliance with SOP.
5.0 Procedure
5.1 Operation
5.1.1 Ensure that all the parts including optical lens, telescope lens, and Polarimeter tube are clean.
5.1.2 Switch on the instrument and warm up sodium lamp to its maximum brightness.
5.1.3 Set the alignment of the sodium lamp compartment to the optical bench.
5.1.4 Ensure that all the setting of scale should be in zero position.
5.1.5 Open the Polarimeter tube compartment and place the Polarimeter tube, which is filled
with solvent in which sample is prepared in such a way as to avoid air-bubbles.
5.1.6 Observing through telescope when the half darkened and half light portions of a
circle are in equilibrium does observation.
5.1.7 This half dark and half light portion is matched with a knob by rotating it in clockwise
or anti-clockwise direction. It is also defined as dextro (+) or levo (-) rotator respectively.
5.1.8 Note the observed readings and also perform a sample determination in the same manner.
5.1.9 Calculate the difference of angle of rotation of sample and blank.
5.1.10 Calculate the specific optical rotation of the sample at 250C by the following formula
α = angle of rotation,
L = length of Polarimeter tube,
C = concentration of solution in %
5.1.11 Record the observations of instruments on Usage of Logbook.
5.2 Calibration (Frequency: after 6 Months)
5.2.1 Switch on the instrument and warm up the sodium lamp to its maximum brightness.
5.2.2 Operate the instrument as per 5.1.2 to 5.1.7
5.2.3 Now fill the Polarimeter tube with 10% w/v solution of sucrose (dried) carry out the
determination in the similar manner as blank.
5.2.4 Repeat the step g. for 20, 30, 40 and 50% w/v concentration of sucrose solution.
5.2.5 The table containing concentration and corresponding angle of rotation in a 2-dm tube at 25°C are indicated below.
| Concentration gm/100 ml | Angle of rotation (+) at 25°C |
| 10.0 | +13.33o |
| 20.0 | +26.61o |
| 30.0 | +39.86o |
| 40.0 | +53.06o |
| 50.0 | +66.23o |
5.2.6 Document the activity in Calibration Record of Polarimeter Format (Annexure I).
5.2.7 Fill the Calibration Tag and affix it on the instrument.
5.2.8 Clean the instrument and keep under cover when not in use.
5.2.9 Report any discrepancy observed during calibration of instrument to Quality
Control- Manager or his designee for corrective and preventive action.
5.2.10 Quality Control- Manager or his designee informs the defect to Maintenance
Department service engineer to rectify the defect. Affix ‘Under Maintenance’ label on the instrument.
5.2.11 Calibration should be required after maintenance.
5.3 Cleaning
5.3.1 Clean the instrument with dry cloth to remove the dust before and after use.
6.0 Reference
Indian Pharmacopoeia
7.0 Attachments
Annexure I : Calibration Record of Polarimeter
8.0 Distribution
Distribution of the controlled copy to concerned departments.
Additional copies may be issued for specific purpose.
9.0 Abbreviations
QA : Quality Assurance
QC : Quality Control
SOP : Standard Operating Procedure
SOR : Specific Optical Rotation
°C : Degree Centigrade
° : Degree
gm : Gram
ml : Milli Liter
+ : Positive
10.0 Revision History
| Revision No. | Supersedes | Effective Date | Details of reason |
| 01 | Nil | New SOP |
Operating Procedure of Centrifuge
sop for preparation and storage of reagent solution
sop for Calibration and Maintenance of Laboratory Instruments and Equipment
Disposal of Residual Sample or Left Over Material
sop for for Laboratory Incident
standard operating procedure temperature monitoring
sop for operation of infrared moisture balance
sop for preparation of mobile phase
sop for Preparation and Issuance of Analysis protocol standard
sop of placebo and impurity stock solutions
sop for disposal of residual sample
sop for handling of pharmacopoeial changes
sop for procedure for operation of ultrasonic cleaner
difference between UPLC and HPLC
sop for for Emergency Eyewash and Shower
sop for operation and calibration of total organic carbon analyzers
sop for operation of cobb tester
sop for Operation and calibration of atomic absorption spectrophotometer
sop for Operation and calibration of gas liquid chromatograph
sop for operation of humidity oven
sop for operation and calibration of serological water bath
sop for monitoring of drain trap
sop for destruction of analytical samples after testing and control samples
sop for destruction of used chemicals
Sop for Operation of suction pump
sop for Operation and calibration uv cabinet
sop for Operation and calibration of bulk density apparatus
sop for operation and calibration of shore hardness tester
sop for operation of rub proofness tester
sop for monitoring of purified water
sop for Retesting of packaging materials
sop for Retesting and resampling of raw materials
sop for Control of issuance of record of analysis green sheets
sop for Control of computer passwords
sop for sampling of packaging materials PM
Operation and Calibration Procedure for pH Meter
Calibration of Melting Point Apparatus
Procedure For HPLC column washing procedure
Procedure For Retest Period of Raw Material
Procedure For Training Programme For Quality Control Personnel
Standard Operating Procedure is to provide guidelines for Preparation
Procedure For Analysis of Water rinsing
Procedure For Calibration of Glassware
Procedure For Procurement and storage of reference standards
Procedure For Preparation of buffer solution
safety guidelines for laboratory
responsibilities of quality control and microbiological department
responsibilities of quality control and microbiological department
handling of corrosive hazardous and flammable chemical
calibration of volumetric glassware
operation and cleaning of water bath
testing and approval rejection of bulk finish and packed drug products
procedure for preparation of general test procedures and standard test procedures
disposal of samples chemicals glassware columns
hplc column performance evaluation storage and destruction
operation and calibration of uv cabinet
operation and calibration of refractometer
operation and calibration of digital micrometer
handling of analytical weight box
operation and calibration of melting point apparatus
Operation and calibration of fuming chamber
Preparation standardization and storage of volumetric solution
Operation and calibration of polarimeter