sop for operation of vernier calipers
1.0 Purpose
1.1 To describe a procedure for operation of Vernier Calipers.
2.0 Scope
2.1 This procedure is applicable for operation of Vernier Calipers located at Chemical analysis
room in Quality Control Department at ABC Pvt. Ltd., Roorkee (UK), India.
3.0 Responsibility
Officer QC/ Executive QC Preparation of SOP and follow the procedure of SOP
Assistant Manager Reviewing and Training of SOP
Manager QC Implementation of SOP
4.0 Accountability
4.1 Manager Quality Control is accountable for implementation and compliance with SOP
5.0 Procedure
5.1 Operation
5.1.1 Ensure that the instrument is clean.
5.1.2 Press ON/ZERO button, 0.00 mm display on the LCD screen of Digital Vernier calipers.
5.1.3 If there is any deviation, adjust the Vernier Calipers by slightly moving it, so that it
give ‘0.00 mm’ display on the screen of Vernier Calipers otherwise press ON/ZERO button to release the error.
5.1.4 Open the jaws slowly with the help of thimble and insert the object between it and then
close the jaws. Read the actual reading of the object directly on the LCD of vernier calipers.
5.1.5 The least count of the apparatus is 0.01mm.
5.1.6 Keep the apparatus in its box, when not in use.
5.2 Calibration
5.2.1 Vernier Calipers is calibrated by external calibration agencies on yearly basis.
5.3 Cleaning
5.3.1 To clean the instrument, use of a soft cloth soaked in a diluted neutral detergent.
5.3.2 Do not use any organic solvent. It may deform or damage the instrument.
5.4 Precaution
5.4.1 Do not use the Vernier Calipers at sites where it will be exposed to oil and water.
5.4.2 Do not store the Vernier Calipers at sites where it will be exposed to dust and moisture.
5.4.3 Do not apply an electric engraver to this instrument it may be damage.
6.0 Reference
In House
7.0 Attachments
Not Applicable
8.0 Distribution
Distribution of the controlled copy to concerned departments.
Additional copies may be issued for specific purpose.
9.0 Abbreviations
QA : Quality Assurance
QC : Quality Control
SOP : Standard Operating Procedure
LCD : Liquid Crystal Display
mm : Milli Meter
10.0 Revision History
| Revision No. | Supersedes | Effective Date | Details of reason |
| 01 | Nil | New SOP |
Operating Procedure of Centrifuge
sop for preparation and storage of reagent solution
sop for Calibration and Maintenance of Laboratory Instruments and Equipment
Disposal of Residual Sample or Left Over Material
sop for for Laboratory Incident
standard operating procedure temperature monitoring
sop for operation of infrared moisture balance
sop for preparation of mobile phase
sop for Preparation and Issuance of Analysis protocol standard
sop of placebo and impurity stock solutions
sop for disposal of residual sample
sop for handling of pharmacopoeial changes
sop for procedure for operation of ultrasonic cleaner
difference between UPLC and HPLC
sop for for Emergency Eyewash and Shower
sop for operation and calibration of total organic carbon analyzers
sop for operation of cobb tester
sop for Operation and calibration of atomic absorption spectrophotometer
sop for Operation and calibration of gas liquid chromatograph
sop for operation of humidity oven
sop for operation and calibration of serological water bath
sop for monitoring of drain trap
sop for destruction of analytical samples after testing and control samples
sop for destruction of used chemicals
Sop for Operation of suction pump
sop for Operation and calibration uv cabinet
sop for Operation and calibration of bulk density apparatus
sop for operation and calibration of shore hardness tester
sop for operation of rub proofness tester
sop for monitoring of purified water
sop for Retesting of packaging materials
sop for Retesting and resampling of raw materials
sop for Control of issuance of record of analysis green sheets
sop for Control of computer passwords
sop for sampling of packaging materials PM
Operation and Calibration Procedure for pH Meter
Calibration of Melting Point Apparatus
Procedure For HPLC column washing procedure
Procedure For Retest Period of Raw Material
Procedure For Training Programme For Quality Control Personnel
Standard Operating Procedure is to provide guidelines for Preparation
Procedure For Analysis of Water rinsing
Procedure For Calibration of Glassware
Procedure For Procurement and storage of reference standards
Procedure For Preparation of buffer solution
safety guidelines for laboratory
responsibilities of quality control and microbiological department
responsibilities of quality control and microbiological department
handling of corrosive hazardous and flammable chemical
calibration of volumetric glassware
operation and cleaning of water bath
testing and approval rejection of bulk finish and packed drug products
procedure for preparation of general test procedures and standard test procedures
disposal of samples chemicals glassware columns
hplc column performance evaluation storage and destruction
