operation and calibration of digital micrometer

 

operation and calibration of digital micrometer

 

1.0 Purpose
1.1 To describe a procedure for operation and calibration of Digital Micrometer.

2.0 Scope
2.1 This procedure is applicable for operation and calibration of Digital Micrometer located at

Chemical Analysis Room in Quality Control Department at ABC Pvt. Ltd.

3.0 Responsibility
Officer QC/ Executive QC Preparation of SOP and follow the procedure as per SOP
Assistant Manager Reviewing and Training of SOP
Manager QC Implementation of SOP

4.0 Accountability
4.1 Manager Quality Control is accountable for implementation and compliance with SOP.

5.0 Procedure
5.1 Operation
5.1.1 Ensure that the instrument shall be clean.
5.1.2 Press the ORIGIN button to set the origin
5.1.3 The origin value, ‘0.000’ is display and “P” flashes on LCD. Than press the ORIGIN button again.
5.1.4 The display value will not change while “P” is flashing.
5.1.5 If the “P” disappears indicating the origin has been set.
5.1.6 Open the jaws slowly with the help of thimble and insert the object between it and then

close the jaws. Read the actual reading of the object directly on the LCD of Micrometer.
5.1.7 The least count of the apparatus is 0.001mm.
5.1.8 Keep the apparatus in its box, when not in use.

5.2 Calibration
5.2.1 Digital Micrometer is calibrated by external calibration agencies.

5.3 Cleaning
5.3.1 To clean the instrument, use of a soft cloth soaked in a diluted neutral detergent.
5.3.2 Do not use any organic solvent. It may deform or damage the instrument.

5.4 Precaution
5.4.1 Do not use the micrometer at sites where it will be exposed to oil and water.
5.4.2 Do not store the micrometer at sites where it will be exposed to dust and moisture.
5.4.3 Do not apply an electric engraver to this instrument it may be damage.

6.0 Reference
In House

7.0 Attachments
Not Applicable
8.0 Distribution
Distribution of the controlled copy to concerned departments.
Additional copies may be issued for specific purpose.

9.0 Abbreviations
QA : Quality Assurance
QC : Quality Control
SOP : Standard Operating Procedure
LCD : Liquid Crystal Display
mm : Milli Meter

10.0 Revision History

Revision No. Supersedes Effective Date. Details of reason
01 Nil New SOP

Operating Procedure of Centrifuge

sop for preparation and storage of reagent solution

sop for Calibration and Maintenance of Laboratory Instruments and Equipment

Disposal of Residual Sample or Left Over Material

sop for for Laboratory Incident

standard operating procedure temperature monitoring

sop for drying of silica gel

sop for operation of infrared moisture balance

sop for preparation of mobile phase

sop for treated RO water

sop for Preparation and Issuance of Analysis protocol standard

sop of placebo and impurity stock solutions

sop for disposal of residual sample  

sop for handling of pharmacopoeial changes

sop for procedure for operation of ultrasonic cleaner

difference between UPLC and HPLC

sop for for Emergency Eyewash and Shower

sop for operation and calibration of total organic carbon analyzers

sop for operation of cobb tester

sop for Operation and calibration of atomic absorption spectrophotometer

sop for Operation and calibration of gas liquid chromatograph

sop for operation of humidity oven

sop for operation and calibration of serological water bath

sop for monitoring of drain trap

sop for destruction of analytical samples after testing and control samples

sop for destruction of used chemicals

Sop for Operation of suction pump

sop for Operation and calibration uv cabinet

sop for Operation and calibration of bulk density apparatus

sop for operation and calibration of shore hardness tester

sop for operation of rub proofness tester

sop for monitoring of purified water

sop for Retesting of packaging materials

sop for Retesting and resampling of raw materials

sop for Control of issuance of record of analysis green sheets

sop for Control of computer passwords

sop for sampling of packaging materials PM

Operation and Calibration Procedure for pH Meter

Calibration of Melting Point Apparatus

Procedure For HPLC column washing procedure

Procedure For Retest Period of Raw Material

Procedure For Training Programme For Quality Control Personnel

Standard Operating Procedure is to provide guidelines for Preparation

Procedure For Analysis of Water rinsing

Procedure For Calibration of Glassware

Procedure For Procurement and storage of reference standards

Procedure For Preparation of buffer solution

safety guidelines for laboratory

responsibilities of quality control and microbiological department

responsibilities of quality control and microbiological department

handling of corrosive hazardous and flammable chemical

cleaning of glassware

calibration of volumetric glassware

operation and cleaning of water bath

sampling of raw materials

testing and approval rejection of bulk finish and packed drug products

procedure for preparation of general test procedures and standard test procedures

disposal of samples chemicals glassware columns

hplc column performance evaluation storage and destruction

operation and calibration of uv cabinet

operation and calibration of refractometer

operation and calibration of digital micrometer

Leave a Comment

error: Content is protected !!