Operation and calibration of bursting strength

 

Operation and calibration of bursting strength

 

1.0 Purpose
1.1 To describe a procedure for operation and calibration of Bursting Strength.

2.0 Scope
2.1 This standard operating procedure is applicable for operation and calibration of Bursting Strength

apparatus located at Quality Control Laboratory at ABC Pvt. Ltd., Roorkee (UK), India.

3.0 Responsibility
Officer QC/ Executive QC Preparation of SOP and follow the procedure as per SOP
Assistant Manager Reviewing and Training of SOP
Manager QC Implementation of SOP
4.0 Accountability
4.1 Manager Quality Control is accountable for implementation and compliance with SOP.

5.0 Procedure
5.1 Operation
5.1.1 Ensure that the instrument is clean and free from dust.
5.1.2 Ensure that the instrument is duly calibrated before use
5.1.3 Connect the apparatus with main power supply. Switch ON from the mains.
5.1.4 Turn the clamping wheel in anti clock wise direction.
5.1.5 Take the sample to be tested and place it over the diaphragm plate.
5.1.6 Clamp the sample by applying adequate pressure by clockwise turn of the wheel.
5.1.7 Select the gauge to be used depending on the sample to be tested. For

lower value like in case of paper and low range boards use the lower range gauge

by turning the rotary switch to gauge-2 position in case of board, turn back the switch to gauge -1 position.
5.1.8 Bring the red indicator of the gauge to zero position.
5.1.9 Place sample to be tested over the Diaphragm between upper and lower

clamping plates. The sample should be at least 100×100 mm (4” x 4”) in size.
5.1.10 Lower Tripod Plate applying sufficient clamping pressure to prevent slip of

the sample between Lower and Upper Clamping Plates during test.
Note: While testing corrugated board, it is recommended that Tripod plate be

lowered, so that sample is properly held and does not slip during testing.
5.1.11 In case of double gauge model, select relevant Pressure Gauge by operating Selector knob.
5.1.12 Set the maximum hand of the Gauge at the zero or at point on the scale below

the point at which bursting strength is expected occur.
5.1.13 Keep ‘Push for Burst’ button pressed till sample bursts. The motor will rotate in

clockwise direction at constant speed till ‘Push for Burst’ button is pressed, pushing plunger

to build pressure. On release of ‘Push for Burst’ button on burst, the motor will automatically

reverse to retract plunger to its starting position. The motor will stop in this position.
5.1.14 Record the bursting pressure indicated by the maximum hand.
5.1.15 Repeat the procedure on several samples or in other areas of the sample as necessary to

obtain average bursting pressure. Avoid overlapping of the clamping areas and disregard any

single reading that is not consistent with those taken in neighboring areas.
5.1.16 Press the push for burst and keep it pressed till the sample bursts.
5.1.17 Record the maximum value in Kg/cm2.

5.2 Cleaning Procedure
5.2.1 Disconnect the oven from main power supply.
5.2.2 Clean the oven externally with the help of clean dry cloth daily.

5.3 Calibration (Frequency: after 12 Months)
5.3.1 For lower readings (0 to 7 Kg/cm2 by) aluminium foil on board tester heaving

standard burst value of 3.0 ± 0.2 Kg/cm2.
5.3.2 Higher readings (above 7 Kg/cm2) by aluminium foil on board tester having

standard burst value of 11.8± 0.3 Kg/cm2.
5.3.3 Document the activity in Calibration Record of Bursting Strength Annexure I.
5.3.4 Fill the Calibration Tag on the instrument.
5.3.5 Report any discrepancy observed during calibration of instrument to Quality Control Manager

or his designee for corrective and preventive action.
5.3.6 Report any discrepancy observed during operation of the instrument to Manager-QC or his

authorized representative, and notify the defect to Engineering Department/ Service Engineer

to rectify the defect. Affix ‘Under Maintenance’ label on the instrument.

6.0 Reference
In house specification

7.0 Attachments
Annexure I : Calibration Record of Bursting Strength
8.0 Distribution
Distribution of the controlled copy to concerned departments.
Additional copies may be issued for specific purpose.

9.0 Abbreviations
QA : Quality Assurance
QC : Quality Control
SOP : Standard Operating Procedure
Kg/cm2 : Kilogram per Centimeter Square

10.0 Revision History

Revision No. Supersedes Effective Date Details of Revision
01 Nil New SOP

 

Operating Procedure of Centrifuge

sop for preparation and storage of reagent solution

sop for Calibration and Maintenance of Laboratory Instruments and Equipment

Disposal of Residual Sample or Left Over Material

sop for for Laboratory Incident

standard operating procedure temperature monitoring

sop for drying of silica gel

sop for operation of infrared moisture balance

sop for preparation of mobile phase

sop for treated RO water

sop for Preparation and Issuance of Analysis protocol standard

sop of placebo and impurity stock solutions

sop for disposal of residual sample  

sop for handling of pharmacopoeial changes

sop for procedure for operation of ultrasonic cleaner

difference between UPLC and HPLC

sop for for Emergency Eyewash and Shower

sop for operation and calibration of total organic carbon analyzers

sop for operation of cobb tester

sop for Operation and calibration of atomic absorption spectrophotometer

sop for Operation and calibration of gas liquid chromatograph

sop for operation of humidity oven

sop for operation and calibration of serological water bath

sop for monitoring of drain trap

sop for destruction of analytical samples after testing and control samples

sop for destruction of used chemicals

Sop for Operation of suction pump

sop for Operation and calibration uv cabinet

sop for Operation and calibration of bulk density apparatus

sop for operation and calibration of shore hardness tester

sop for operation of rub proofness tester

sop for monitoring of purified water

sop for Retesting of packaging materials

sop for Retesting and resampling of raw materials

sop for Control of issuance of record of analysis green sheets

sop for Control of computer passwords

sop for sampling of packaging materials PM

Operation and Calibration Procedure for pH Meter

operation and calibration of muffle furnace

Operation of magnetic stirrer

Operation and cleaning of centrifuge

Operation and cleaning of sonicator

Operation and calibration of dissolution test apparatus

Operation and cleaning of hot air oven

Operation and calibration of bursting strength

 

Leave a Comment

error: Content is protected !!