Calibration of Melting Point Apparatus

 

Calibration of Melting Point Apparatus

 

1.0 Objective
To lay down a guidelines for Calibration of Melting Point Apparatus in Quality Control Department.
2.0 Scope
This SOP applies to Calibration of Melting Point Apparatus used in Quality control Lab in abc company.
3.0 Responsibility
3.1 Executive – Quality Control.
3.2 In charge/Head.
4.0 Abbreviations and Definitions
SOP : Standard Operating Procedure
QC : Quality Control
CC NO : Change Control Number
cGMP : Current Good Manufacturing Practice
C : Calibration
5.0 Procedure
5.1 First check the instrument is free from dust and foreign matter.
5.2 Now on the Switch of main supply then on the switch of the instrument.
5.3 Fill the beaker with silicon oil / liquid paraffin.
5.4 Take the sample in melting point capillary to form a compact column.
5.5 Adjust the heating rate to Low / High according to the temperature specified.
5.6 Switch on the Lamp and Heater.
5.7 Switch on the “Boost Heater”
5.8 Introduce the capillary tube containing the sample when the temperature is 10° below the expected temperature.
5.9 When the Sample is start to the melts and complete melts note down the temperature using a calibrated thermometer.
5.10 Take standard raw material and check the melting point. Keep standard thermometer. As per list.
Sr. No. Reference Standard Result Temperature Range

Sr. No. Reference Standard Result Temperature Range
1 VANILLIN 81.0 to 83.0°C
2 SULFANILAMIDE 164.5 to 166.5°C
3 CAFFEINE 234.0  to 237.0°C

6.0 Forms and Records
6.1 Calibration of Melting Point – Annexure-1

7.0 Distribution
7.1 Master copy – Documentation Cell (Quality Assurance)
7.2 Control copy – Quality Control

8.0 History

Revision Number Details For Change
Reason for Revision
00 New SOP NA

Operating Procedure of Centrifuge

sop for preparation and storage of reagent solution

sop for Calibration and Maintenance of Laboratory Instruments and Equipment

Disposal of Residual Sample or Left Over Material

sop for for Laboratory Incident

standard operating procedure temperature monitoring

sop for drying of silica gel

sop for operation of infrared moisture balance

sop for preparation of mobile phase

sop for treated RO water

sop for Preparation and Issuance of Analysis protocol standard

sop of placebo and impurity stock solutions

sop for disposal of residual sample  

sop for handling of pharmacopoeial changes

sop for procedure for operation of ultrasonic cleaner

difference between UPLC and HPLC

sop for for Emergency Eyewash and Shower

sop for operation and calibration of total organic carbon analyzers

sop for operation of cobb tester

sop for Operation and calibration of atomic absorption spectrophotometer

sop for Operation and calibration of gas liquid chromatograph

sop for operation of humidity oven

sop for operation and calibration of serological water bath

sop for monitoring of drain trap

sop for destruction of analytical samples after testing and control samples

sop for destruction of used chemicals

Sop for Operation of suction pump

sop for Operation and calibration uv cabinet

sop for Operation and calibration of bulk density apparatus

sop for operation and calibration of shore hardness tester

sop for operation of rub proofness tester

sop for monitoring of purified water

sop for Retesting of packaging materials

sop for Retesting and resampling of raw materials

sop for Control of issuance of record of analysis green sheets

sop for Control of computer passwords

sop for sampling of packaging materials PM

Operation and Calibration Procedure for pH Meter

Calibration of Melting Point Apparatus

Leave a Comment

error: Content is protected !!