operation and calibration of uv cabinet
1.0 Purpose
1.1 To describe a procedure for operation and calibration of UV Cabinet.
2.0 Scope
2.1 This procedure is applicable for operation and calibration of UV Cabinet located at Instrumentation
Room in Quality Control laboratory at ABC Pvt. Ltd.
3.0 Responsibility
Officer QC/ Executive QC Preparation of SOP and follow the procedure as per SOP
Assistant Manager Reviewing and Training of SOP
Manager QC Implementation of SOP
4.0 Accountability
4.1 Manager Quality Control is accountable for implementation and compliance with SOP.
5.0 Procedure
5.1 Operation
5.1.1 Ensure that the instrument is free from dust.
5.1.2 Switch on the mains.
5.1.3 Switch ON the instrument indicator light will be ON. After that off the switch.
5.1.4 Open the cabinet and place the TLC plate inside and close the cabinet.
5.1.5 Three switches are provided on the top right corner of the instrument.
Switch 1: For UV-Light at long wavelength
Switch 2: For UV-Light at short wavelength
Switch 3: For Visible
5.1.6 Select the required UV-Light and switch ON.
5.1.7 Observe the spot from the TLC plate.
5.1.8 After that switch off the instrument.
5.2 Calibration (Frequency: after 6 Months)
5.2.1 Ultra-Violet lamps used for viewing of Thin Layer Chromatography can be calibrated.
5.2.2 Prepare a 0.04% w/v solution of sodium salicylate in alcohol and apply a 5 micro litre spot
on a plate coated with silica gel GF-254 for lamps emitting emission band with maximum intensity at about 254 nm.
5.2.3 Place the TLC plate under the radiation so that the radiation falls perpendicularly on the plate and examine the spot.
5.2.4 The lamp should be capable of revealing a standard spot of a diameter of about 5 mm of the plate.
5.2.5 Prepare a 0.02% w/v solution of sodium salicylate in alcohol and apply a 5 micro litre spot on
a plate coated with silica gel GF-254 for lamps emitting emission band with maximum intensity at about 365 nm.
5.2.6 Place the TLC plate under the radiation so that the radiations fall perpendicularly on the plate and examine the spot.
5.2.7 The lamp should be capable of revealing a standard spot of a diameter of about 5 mm of the plate.
5.2.8 The distance between the lamp and the chromatographic plate under examination used in
a pharmacopoeia test should never exceed the distance used to carry out the above test.
5.2.9 Use the calibrated and certified UV light meter for measure the intensity of UV light.
5.2.10 Document the activity of Calibration in Annexure I.
5.2.11 Fill the Calibration Tag and affix it on the instrument.
5.2.12 Report any discrepancy observed during calibration of instrument to Quality Control In-Charge
or his designee for corrective and preventive action.
5.2.13 Quality Control In-Charge or his designee informs the defect to Maintenance Department
and service engineer to rectify the defect. Affix ‘Under Maintenance’ label on the instrument.
5.2.14 Calibration should be required after maintenance.
6.0 Reference
In House Specification
7.0 Attachments
Annexure I : Calibration record format
8.0 Distribution
Distribution of the controlled copy to concerned departments.
Additional copies may be issued for specific purpose.
9.0 Abbreviations
QA : Quality Assurance
QC : Quality Control
SOP : Standard Operating Procedure
TLC : Thin Layer Chromatography
UV : Ultra Violate
mm : Milli Meter
10.0 Revision History
| Revision No. | Supersedes | Effective Date | Details of reason |
| 01 | Nil | New SOP |
Operating Procedure of Centrifuge
sop for preparation and storage of reagent solution
sop for Calibration and Maintenance of Laboratory Instruments and Equipment
Disposal of Residual Sample or Left Over Material
sop for for Laboratory Incident
standard operating procedure temperature monitoring
sop for operation of infrared moisture balance
sop for preparation of mobile phase
sop for Preparation and Issuance of Analysis protocol standard
sop of placebo and impurity stock solutions
sop for disposal of residual sample
sop for handling of pharmacopoeial changes
sop for procedure for operation of ultrasonic cleaner
difference between UPLC and HPLC
sop for for Emergency Eyewash and Shower
sop for operation and calibration of total organic carbon analyzers
sop for operation of cobb tester
sop for Operation and calibration of atomic absorption spectrophotometer
sop for Operation and calibration of gas liquid chromatograph
sop for operation of humidity oven
sop for operation and calibration of serological water bath
sop for monitoring of drain trap
sop for destruction of analytical samples after testing and control samples
sop for destruction of used chemicals
Sop for Operation of suction pump
sop for Operation and calibration uv cabinet
sop for Operation and calibration of bulk density apparatus
sop for operation and calibration of shore hardness tester
sop for operation of rub proofness tester
sop for monitoring of purified water
sop for Retesting of packaging materials
sop for Retesting and resampling of raw materials
sop for Control of issuance of record of analysis green sheets
sop for Control of computer passwords
sop for sampling of packaging materials PM
Operation and Calibration Procedure for pH Meter
Calibration of Melting Point Apparatus
Procedure For HPLC column washing procedure
Procedure For Retest Period of Raw Material
Procedure For Training Programme For Quality Control Personnel
Standard Operating Procedure is to provide guidelines for Preparation
Procedure For Analysis of Water rinsing
Procedure For Calibration of Glassware
Procedure For Procurement and storage of reference standards
Procedure For Preparation of buffer solution
safety guidelines for laboratory
responsibilities of quality control and microbiological department
responsibilities of quality control and microbiological department
handling of corrosive hazardous and flammable chemical
calibration of volumetric glassware
operation and cleaning of water bath
testing and approval rejection of bulk finish and packed drug products
procedure for preparation of general test procedures and standard test procedures
disposal of samples chemicals glassware columns