operation and calibration of uv cabinet

operation and calibration of uv cabinet

 

1.0 Purpose
1.1 To describe a procedure for operation and calibration of UV Cabinet.

2.0 Scope
2.1 This procedure is applicable for operation and calibration of UV Cabinet located at Instrumentation

Room in Quality Control laboratory at ABC Pvt. Ltd.

3.0 Responsibility
Officer QC/ Executive QC Preparation of SOP and follow the procedure as per SOP
Assistant Manager Reviewing and Training of SOP
Manager QC Implementation of SOP
4.0 Accountability
4.1 Manager Quality Control is accountable for implementation and compliance with SOP.

5.0 Procedure
5.1 Operation
5.1.1 Ensure that the instrument is free from dust.
5.1.2 Switch on the mains.
5.1.3 Switch ON the instrument indicator light will be ON. After that off the switch.
5.1.4 Open the cabinet and place the TLC plate inside and close the cabinet.
5.1.5 Three switches are provided on the top right corner of the instrument.
Switch 1: For UV-Light at long wavelength
Switch 2: For UV-Light at short wavelength
Switch 3: For Visible
5.1.6 Select the required UV-Light and switch ON.
5.1.7 Observe the spot from the TLC plate.
5.1.8 After that switch off the instrument.
5.2 Calibration (Frequency: after 6 Months)
5.2.1 Ultra-Violet lamps used for viewing of Thin Layer Chromatography can be calibrated.
5.2.2 Prepare a 0.04% w/v solution of sodium salicylate in alcohol and apply a 5 micro litre spot

on a plate coated with silica gel GF-254 for lamps emitting emission band with maximum intensity at about 254 nm.
5.2.3 Place the TLC plate under the radiation so that the radiation falls perpendicularly on the plate and examine the spot.
5.2.4 The lamp should be capable of revealing a standard spot of a diameter of about 5 mm of the plate.
5.2.5 Prepare a 0.02% w/v solution of sodium salicylate in alcohol and apply a 5 micro litre spot on

a plate coated with silica gel GF-254 for lamps emitting emission band with maximum intensity at about 365 nm.
5.2.6 Place the TLC plate under the radiation so that the radiations fall perpendicularly on the plate and examine the spot.
5.2.7 The lamp should be capable of revealing a standard spot of a diameter of about 5 mm of the plate.
5.2.8 The distance between the lamp and the chromatographic plate under examination used in

a pharmacopoeia test should never exceed the distance used to carry out the above test.
5.2.9 Use the calibrated and certified UV light meter for measure the intensity of UV light.
5.2.10 Document the activity of Calibration in Annexure I.
5.2.11 Fill the Calibration Tag and affix it on the instrument.
5.2.12 Report any discrepancy observed during calibration of instrument to Quality Control In-Charge

or his designee for corrective and preventive action.
5.2.13 Quality Control In-Charge or his designee informs the defect to Maintenance Department

and service engineer to rectify the defect. Affix ‘Under Maintenance’ label on the instrument.
5.2.14 Calibration should be required after maintenance.

6.0 Reference
In House Specification

7.0 Attachments
Annexure I : Calibration record format

8.0 Distribution
Distribution of the controlled copy to concerned departments.
Additional copies may be issued for specific purpose.

9.0 Abbreviations
QA : Quality Assurance
QC : Quality Control
SOP : Standard Operating Procedure
TLC : Thin Layer Chromatography
UV : Ultra Violate
mm : Milli Meter

10.0 Revision History

Revision No. Supersedes Effective Date Details of reason
01 Nil New SOP

Operating Procedure of Centrifuge

sop for preparation and storage of reagent solution

sop for Calibration and Maintenance of Laboratory Instruments and Equipment

Disposal of Residual Sample or Left Over Material

sop for for Laboratory Incident

standard operating procedure temperature monitoring

sop for drying of silica gel

sop for operation of infrared moisture balance

sop for preparation of mobile phase

sop for treated RO water

sop for Preparation and Issuance of Analysis protocol standard

sop of placebo and impurity stock solutions

sop for disposal of residual sample  

sop for handling of pharmacopoeial changes

sop for procedure for operation of ultrasonic cleaner

difference between UPLC and HPLC

sop for for Emergency Eyewash and Shower

sop for operation and calibration of total organic carbon analyzers

sop for operation of cobb tester

sop for Operation and calibration of atomic absorption spectrophotometer

sop for Operation and calibration of gas liquid chromatograph

sop for operation of humidity oven

sop for operation and calibration of serological water bath

sop for monitoring of drain trap

sop for destruction of analytical samples after testing and control samples

sop for destruction of used chemicals

Sop for Operation of suction pump

sop for Operation and calibration uv cabinet

sop for Operation and calibration of bulk density apparatus

sop for operation and calibration of shore hardness tester

sop for operation of rub proofness tester

sop for monitoring of purified water

sop for Retesting of packaging materials

sop for Retesting and resampling of raw materials

sop for Control of issuance of record of analysis green sheets

sop for Control of computer passwords

sop for sampling of packaging materials PM

Operation and Calibration Procedure for pH Meter

Calibration of Melting Point Apparatus

Procedure For HPLC column washing procedure

Procedure For Retest Period of Raw Material

Procedure For Training Programme For Quality Control Personnel

Standard Operating Procedure is to provide guidelines for Preparation

Procedure For Analysis of Water rinsing

Procedure For Calibration of Glassware

Procedure For Procurement and storage of reference standards

Procedure For Preparation of buffer solution

safety guidelines for laboratory

responsibilities of quality control and microbiological department

responsibilities of quality control and microbiological department

handling of corrosive hazardous and flammable chemical

cleaning of glassware

calibration of volumetric glassware

operation and cleaning of water bath

sampling of raw materials

testing and approval rejection of bulk finish and packed drug products

procedure for preparation of general test procedures and standard test procedures

disposal of samples chemicals glassware columns

hplc column performance evaluation storage and destruction

operation and calibration of uv cabinet

Leave a Comment

error: Content is protected !!