Procedure For Calibration of Glassware
1.0 Objective
To define a procedure for analysis of rinsing water samples received from production dept.
so as to as certain the equipments free from previous contamination.
2.0 Scope
Rinsing water samples received from production department at abc Pvt. Ltd.
3.0 Responsibility
3.1 Executive – Quality Control.
3.2 In charge/Head-Quality Control.
4.0 Abbreviations and Definitions
SOP : Standard Operating Procedure
QC : Quality Control
CC NO : Change Control Number
cGMP : Current Good Manufacturing Practice
G : General
5.0 Procedure
5.1 Nominal capacity of the volumetric glassware is determined gravimetrically in cubic
meters at the Indian Standard Reference Temperature 27°C using distilled water
over a temperature range 5°C to 40°C by using prescribed table of conversion.
5.2 Clean the glassware thoroughly following SOP for cleaning of glassware’s.
5.3 Make sure that the vessel or weighing bottle, water and the glassware are at room temperature.
5.4 Accurately weigh the vessel to be verified or a weighing bottle, if the vessel
is intended to deliver on an appropriate standardized balance.
5.5 Fill the vessel intended to contain ( volumetric flask) with distilled water to a
distance of a few millimeters above the graduation line to be tested and make the final
setting of the Meniscus to the line by withdrawing the surplus water by means of a pipette.
5.5.1 Setting of meniscus: Arrange the lighting so that the meniscus appears dark and
distinct in outline. For this purpose it should be viewed against white background and shaded
from undesirable illumination, this can be achieved by keeping black paper around the
vessel 1 mm below the level of setting.
5.5.2 Ensure that your body temperature does not affect volume of water by holdin the
article away from water level. Set the meniscus such that the top of the graduated line is just
touching the bottom of the meniscus. Parallax is avoided when graduation lines are of
sufficient length to be seen at the front and back of the vessel simultaneously.
5.5.3 After 2 minutes weigh the filled vessel for the same accuracy as in 1.3
and measure the temperature of the water.
5.5.4 Hold the vessel intended to deliver (pipettes) in a vertical position and fill to a
few millimeters above the graduation line to be tested. Remove any liquid remaining
on the outside of the jet. (Tip) by means of using tissue paper. Then make the setting by
running out the surplus water and adjusting to the graduation line. Remove any drop of
liquid adhering to the jet by bringing an inclined glass surface into contact with the tip
of the jet. Deliver the water into the tarred weighing bottle with the flow unrestricted.
5.5.5 The method of delivery described above leaves a small quantity of water remaining in
the jet of the pipette, no method of emptying such as blowing out or tapping should be used
which expel liquid completely from the jet. To ensure that the delivery is complete, a waiting
period of approximately 3 seconds be observed. Weigh the weighing bottle to the same
accuracy as in 1.3 and measure the temperature of the water.
5.6 For Burettes:
5.6.1 Hold the burette in a vertical position and fill it to a few millimeters above the
zero graduation. Remove any liquid remaining on the outside of the tip by means of
tissue Paper. Then open the stop cork and drain the liquid and set the meniscus to
zero graduation. Remove any drop of liquid adhering to the tip by bringing an inclined glass surface into contact.
5.6.2 Drain the liquid into a tarred weighing bottle, outflow being interrupted when
the meniscus has come to a few millimeters above the final graduation line. Note the
time taken in secs. Wait for 30 secs and set the meniscus to the graduation line. Weigh
the weighing bottle and measure the temperature of the water. Measure the total length
of the burette the scale length.
5.6.3 The difference between the results of the first 1.3 and the second 1.5 weighing is the
mass of water contained in or delivered by the vessel under verification.
5.7 Identify the calibrated vessel by an appropriate identification mark.
5.8 Enter the details of the calibration in the format provided. Make a file of the calibration formats.
5.9 If there is use of A grade glassware then make record file of the calibration
certificate proved by party with them. In that case there is no need of calibration.
5.10 The frequency of calibration of glassware will be once initially.
6.0 Forms and Records
6.1 None.
7.0 Distribution
7.1 Master copy – Documentation Cell (Quality Assurance)
7.2 Control copy – Quality Assurance, Quality Control
8.0 History
Revision Number | Details For Change |
Reason for Revision |
00 | New SOP | NA |
Operating Procedure of Centrifuge
sop for preparation and storage of reagent solution
sop for Calibration and Maintenance of Laboratory Instruments and Equipment
Disposal of Residual Sample or Left Over Material
sop for for Laboratory Incident
standard operating procedure temperature monitoring
sop for operation of infrared moisture balance
sop for preparation of mobile phase
sop for Preparation and Issuance of Analysis protocol standard
sop of placebo and impurity stock solutions
sop for disposal of residual sample
sop for handling of pharmacopoeial changes
sop for procedure for operation of ultrasonic cleaner
difference between UPLC and HPLC
sop for for Emergency Eyewash and Shower
sop for operation and calibration of total organic carbon analyzers
sop for operation of cobb tester
sop for Operation and calibration of atomic absorption spectrophotometer
sop for Operation and calibration of gas liquid chromatograph
sop for operation of humidity oven
sop for operation and calibration of serological water bath
sop for monitoring of drain trap
sop for destruction of analytical samples after testing and control samples
sop for destruction of used chemicals
Sop for Operation of suction pump
sop for Operation and calibration uv cabinet
sop for Operation and calibration of bulk density apparatus
sop for operation and calibration of shore hardness tester
sop for operation of rub proofness tester
sop for monitoring of purified water
sop for Retesting of packaging materials
sop for Retesting and resampling of raw materials
sop for Control of issuance of record of analysis green sheets
sop for Control of computer passwords
sop for sampling of packaging materials PM
Operation and Calibration Procedure for pH Meter
Calibration of Melting Point Apparatus
Procedure For HPLC column washing procedure
Procedure For Retest Period of Raw Material
Procedure For Training Programme For Quality Control Personnel
Standard Operating Procedure is to provide guidelines for Preparation