Procedure For Receipt performance and handling of columns

 

Procedure For Receipt performance and handling of columns

 

1.0 Objective
To describe a procedure for receipt, performance and handling of columns.
2.0 Scope
This SOP is applicable to QC Department at abc Pvt. Ltd.
3.0 Responsibility
3.1 Executive – Quality Control.
3.2 In charge/Head-Quality Control.
4.0 Abbreviations and Definitions
SOP : Standard Operating Procedure
QC : Quality Control
CC NO : Change Control Number
cGMP : Current Good Manufacturing Practice
G : General
5.0 Procedure
5.1 RECEIPT AND HANDLING OF NEW COLUMNS
5.1.1 QC Officer shall maintain a record for chromatography columns used for HPLC.
5.1.2 The QC Officer shall have the list of the various types of chromatography

columns available or to be used. The list shall have following details:
5.1.3 S.No. Name and type of column, Manufacturer’s name, Column ID,

Specification (dimensions, particle size, packing material).
5.1.4 The list shall be updated if a new type of column is introduced or deleted from analysis programmed.
5.1.5 The QC Officer shall enter the record for each type of column in the

inventory as per current version of format. And give the unique number to each and every column.
5.1.6 The unique identification number for columns shall be assigned as per the procedure given below:
5.1.7 The manufacturer’s column performance certificate shall be filed by QC Officer and retained for future reference.
5.1.8 QC Officer shall issue column(s) for use only after making proper entries in the record.
5.1.9 The column performance shall be checked by the QC Officer as per the system

suitability requirements of in-house method.
5.1.10 In case of column’s performance is found to be unsatisfactory, the

Manager-QC Manager-QA shall reject and take up the matter with the supplier. The action shall be recorded in the remarks column
5.1.11 During the use, the system suitability parameters obtained on the column

may be recorded along with product analyzed, to track the use and performance of the column.
5.1.12 If during use, the column’s performance is found unsatisfactory, the QC Officer shall report to the QC Officer.
5.1.13 Only after QC Officer has confirmed that the column is not performing satisfactorily,

it shall be withdrawn from use and identified with appropriate sticker “REJECTED”
5.1.14 The date of disposal of the column shall be entered in the record by the QC Officer.
5.1.15 The Manager-QC / Manager-QA shall ensure that the laboratory is equipped with all requisite columns necessary for analysis.
5.2 WASHING OF COLUMNS(For C18, C8, CN, column)
5.2.1 Change the mobile phase to 50:50 % v/v filtered distilled water and appropriate solvent (which is used in the mobile phase).
5.2.2 Allow the water to flow through the column at the rate of 1ml/min. for 60 minutes.

Check that pressure does not exceed 6000 PSI.
5.2.3 Ensure that column contains appropriate solvent.
5.2.4 Disconnect the column from the HPLC unit, put the end fittings and store in designated place.
5.2.5 For protein pack column.
5.2.6 Change the mobile phase to 10% isopropyl alcohol.
5.2.7 Allow 10% Isopropyl Alcohol to flow through the column at the rate of 1ml/min. for 30minutes.

Check that the pressure does not exceed 6000 PSI.
5.2.8 Ensure that the column contains 10% Isopropyl Alcohol.
5.2.9 Disconnect the column from the HPLC unit, put the end fittings and store in designated place.
5.3 REGENERATION OF COLUMNS (For C18, C8, CN.)
5.3.1 First flush the column with 55°C hot distilled water with flow rate 1 ml/min. for 50 minutes.
5.3.2 Then flush the column with solvent in following order with flow rate of 1ml/min. for 50 minutes.
a) Methanol HPLC grade.
b) Chloroform HPLC grade
c) Methanol HPLC grade
5.3.3 Ensure that the column storage solvent is Methanol.
5.3.4 Disconnect the column from the HPLC unit fix the end caps and store it in an appropriate place.
5.4 USAGE OF COLUMNS
5.4.1 HPLC/GC columns shall be used as per the respective standard testing procedure.
5.4.2 Usage of the each and every column shall be recorded.
6.0 Forms and Records
6.1 None.
7.0 Distribution
7.1 Master copy – Documentation Cell (Quality Assurance)
7.2 Control copy – Quality Assurance, Quality Control

8.0 History

Revision Number Details For Change
Reason for Revision
00 New SOP NA

Operating Procedure of Centrifuge

sop for preparation and storage of reagent solution

sop for Calibration and Maintenance of Laboratory Instruments and Equipment

Disposal of Residual Sample or Left Over Material

sop for for Laboratory Incident

standard operating procedure temperature monitoring

sop for drying of silica gel

sop for operation of infrared moisture balance

sop for preparation of mobile phase

sop for treated RO water

sop for Preparation and Issuance of Analysis protocol standard

sop of placebo and impurity stock solutions

sop for disposal of residual sample  

sop for handling of pharmacopoeial changes

sop for procedure for operation of ultrasonic cleaner

difference between UPLC and HPLC

sop for for Emergency Eyewash and Shower

sop for operation and calibration of total organic carbon analyzers

sop for operation of cobb tester

sop for Operation and calibration of atomic absorption spectrophotometer

sop for Operation and calibration of gas liquid chromatograph

sop for operation of humidity oven

sop for operation and calibration of serological water bath

sop for monitoring of drain trap

sop for destruction of analytical samples after testing and control samples

sop for destruction of used chemicals

Sop for Operation of suction pump

sop for Operation and calibration uv cabinet

sop for Operation and calibration of bulk density apparatus

sop for operation and calibration of shore hardness tester

sop for operation of rub proofness tester

sop for monitoring of purified water

sop for Retesting of packaging materials

sop for Retesting and resampling of raw materials

sop for Control of issuance of record of analysis green sheets

sop for Control of computer passwords

sop for sampling of packaging materials PM

Operation and Calibration Procedure for pH Meter

Calibration of Melting Point Apparatus

Procedure For HPLC column washing procedure

Procedure For Retest Period of Raw Material

Procedure For Training Programme For Quality Control Personnel

Standard Operating Procedure is to provide guidelines for Preparation

Procedure For Analysis of Water rinsing

Procedure For Calibration of Glassware

Procedure For Procurement and storage of reference standards

Procedure For Preparation of buffer solution

 

Leave a Comment

error: Content is protected !!