Procedure For Storage and maintenance of raw material control sample
1.0 Objective
To define a procedure for Storage and maintenance of raw material control sample.
2.0 Scope
Raw Material Control Sample retained in Quality Control Laboratory at abc Pvt. Ltd.
3.0 Responsibility
3.1 Executive – Quality Control.
3.2 In charge/Head-Quality Control.
4.0 Abbreviations and Definitions
SOP : Standard Operating Procedure
QC : Quality Control
CC NO : Change Control Number
cGMP : Current Good Manufacturing Practice
G : General
5.0 Procedure
5.1 Control Sample of every raw material received and collected at the time of sampling
in a Quantity sufficient to carry out all tests twice, except sterility / Bacterial End toxin
shall be kept in appropriate amber coloured glass bottles securely closed.
5.2 Sample of corrosive substance like strong acids, ammonia will not be kept as control samples.
5.3 Every container will be labeled with name G.R.N. No., Batch No., Mfg. Date and Exp. Date,
wherever available, Laboratory reference No. date of sampling and the signature of the sampler.
5.4 The control sample should be stored in the corrugated box month wise in the control sample
room and the entry should be made in the register meant for raw material retain sample.
5.5 The control sample should be store for one year after the expiry of the sample.
5.6 After one year from the expiry of sample, the control samples will be destroyed.
6.0 Forms and Records
6.1 None.
7.0 Distribution
7.1 Master copy – Documentation Cell (Quality Assurance)
7.2 Control copy – Quality Assurance, Quality Control
8.0 History
Revision Number | Details For Change |
Reason for Revision |
00 | New SOP | NA |
Operating Procedure of Centrifuge
sop for preparation and storage of reagent solution
sop for Calibration and Maintenance of Laboratory Instruments and Equipment
Disposal of Residual Sample or Left Over Material
sop for for Laboratory Incident
standard operating procedure temperature monitoring
sop for operation of infrared moisture balance
sop for preparation of mobile phase
sop for Preparation and Issuance of Analysis protocol standard
sop of placebo and impurity stock solutions
sop for disposal of residual sample
sop for handling of pharmacopoeial changes
sop for procedure for operation of ultrasonic cleaner
difference between UPLC and HPLC
sop for for Emergency Eyewash and Shower
sop for operation and calibration of total organic carbon analyzers
sop for operation of cobb tester
sop for Operation and calibration of atomic absorption spectrophotometer
sop for Operation and calibration of gas liquid chromatograph
sop for operation of humidity oven
sop for operation and calibration of serological water bath
sop for monitoring of drain trap
sop for destruction of analytical samples after testing and control samples
sop for destruction of used chemicals
Sop for Operation of suction pump
sop for Operation and calibration uv cabinet
sop for Operation and calibration of bulk density apparatus
sop for operation and calibration of shore hardness tester
sop for operation of rub proofness tester
sop for monitoring of purified water
sop for Retesting of packaging materials
sop for Retesting and resampling of raw materials
sop for Control of issuance of record of analysis green sheets
sop for Control of computer passwords
sop for sampling of packaging materials PM
Operation and Calibration Procedure for pH Meter
Calibration of Melting Point Apparatus
Procedure For HPLC column washing procedure
Procedure For Storage and maintenance of raw material control sample