sop for Operation and calibration of active air sampler

 

sop for Operation and calibration of active air sampler

 

1.0 : OBJECTIVE :

To lay down a procedure for operation and calibration of active-air sampler.

2.0 : RESPONSIBILITY

Microbiologist / Executive

3.0 : ACCOUNTABILITY

Q.C Manager

4.0 : PROCEDURE

: PRE –CHECKS

4.1.1 : Before transferring the active-air sampler to the clean-area check the power-supply
and proper working of the active-air sampler.

4.2 : OPERATION:



4.2.1 : Connect the main power cord of active-air sampler with power board.

4.2.2 : Open the sterile mesh of the active-air sampler and place sterile plate inside the active-
air sampler and open the lid of the plate.

4.2.3 : Switch on the power button of the active-air sampler, green light will glow at the
power indicator.

4.2.4 : Press the ON button on the active-air sampler & set the timer to 200 seconds(to
sample 1000liters of air).

4.2.5 : By pressing the ON button ,green light will glow at the motor on indicator and
seconds clock starts .

4.2.6 : Come out of the area within ten seconds.

4.2.7 : The active-air sampler will sample the air of the location upto 200 seconds and
automatically stops after completion of 200 seconds.

4.2.8 : Cover the plate with lid and take out the plate from the sampler.

4.2.9 : Label the area and location number under the plate .

5.0 Calibration:
Fan-speed : 2400 RPM



5.1 : Connect the calibrated tachometer with the upper edge of the fan of active air sampler
and note the readings.

5.2 : Repeat the process for at least three times and note the readings in the format

5.3 : Limit :  percent limit

5.4 : Affix calibration tag, if calibration is satisfactory and under maintenance if found
Unsatisfactory.

6.0 : ABBREVIATION:

Nil

7.0 : REFERENCES:
USP

 

Leave a Comment

error: Content is protected !!