Operation and calibration of fuming chamber
1.0 Purpose
1.1 To describe a procedure for operation and cleaning of fuming chamber.
2.0 Scope
2.1 This standard operating procedure is applicable for operation and cleaning of Fuming Chamber located
at Chemical Analysis Room in Quality Control laboratory at ABC Pvt. Ltd.
3.0 Responsibility
Officer QC/ Executive QC Preparation of SOP and follow the procedure as per SOP
Assistant Manager Reviewing and Training of SOP
Manager QC Implementation of SOP
4.0 Accountability
4.1 Manager Quality Control is accountable for implementation and compliance with SOP.
5.0 Procedure
5.1 Operation
5.1.1 Ensure that fuming chamber is clean and no other activity is under operation in chamber.
5.1.2 Open the glass door for the work by sliding up the door.
5.1.3 Push the green switch on the exhaust of the fuming chamber.
5.1.4 Connect the instrument (Muffle furnace etc.) with main power supply and others during the analysis.
5.1.5 Carry out the test as per standard test procedure in the chamber.
5.1.6 Fumigating solvent e.g. hydrochloric acid sulphuric acid and nitric acid on sand tray inside the fuming chamber.
5.1.7 Remove the test sample from chamber after analysis and close the front glass door.
5.1.8 After completion of work clean the chamber carefully.
5.1.9 Push the red switch off the exhaust of the fuming chamber.
5.2 Cleaning of Fuming Chamber
5.2.1 Proper functioning of fuming chamber is essential to maintain laboratory, pollution free operation.
5.2.2 Clean the exterior and top of the fume chamber, first by sweeping the top and dusting with cloth.
5.2.3 Shut off power supply from mains.
5.2.4 Clean the exhaust ducting with brush.
5.2.5 Clean the tube light and surface of fuming chamber with a piece of cloth.
5.2.6 Check the performance of these following point
5.2.6.1 Starter of exhaust fan
5.2.6.2 Performance of exhaust fan
5.2.6.3 Performance of tube light change if required
5.3 If the paint on interior surface has become dull or having stains, arrange for re-painting.
5.4 Calibration once in year
5.4.1 Fuming chamber calibrated by outside external calibration agencies.
5.5 Precaution
5.5.1 Ensure that there are no inflammable chemicals present in the chamber.
5.5.2 Clean the fume chamber after disconnecting all electrical connection.
5.5.3 Use strong Acid carefully in chamber.
5.5.4 Air exhaust should be on at the working time in fume chamber.
6.0 Reference
In House Specification
7.0 Attachments
Nil
8.0 Distribution
Distribution of the controlled copy to concerned departments.
Additional copies may be issued for specific purpose.
9.0 Abbreviations
QA : Quality Assurance
QC : Quality Control
SOP : Standard Operating Procedure
oC : Degree Centigrade
10.0 Revision History
| Revision No. | Supersedes | Effective Date | Details of Revision |
| 01 | Nil | New SOP |
Operating Procedure of Centrifuge
sop for preparation and storage of reagent solution
sop for Calibration and Maintenance of Laboratory Instruments and Equipment
Disposal of Residual Sample or Left Over Material
sop for for Laboratory Incident
standard operating procedure temperature monitoring
sop for operation of infrared moisture balance
sop for preparation of mobile phase
sop for Preparation and Issuance of Analysis protocol standard
sop of placebo and impurity stock solutions
sop for disposal of residual sample
sop for handling of pharmacopoeial changes
sop for procedure for operation of ultrasonic cleaner
difference between UPLC and HPLC
sop for for Emergency Eyewash and Shower
sop for operation and calibration of total organic carbon analyzers
sop for operation of cobb tester
sop for Operation and calibration of atomic absorption spectrophotometer
sop for Operation and calibration of gas liquid chromatograph
sop for operation of humidity oven
sop for operation and calibration of serological water bath
sop for monitoring of drain trap
sop for destruction of analytical samples after testing and control samples
sop for destruction of used chemicals
Sop for Operation of suction pump
sop for Operation and calibration uv cabinet
sop for Operation and calibration of bulk density apparatus
sop for operation and calibration of shore hardness tester
sop for operation of rub proofness tester
sop for monitoring of purified water
sop for Retesting of packaging materials
sop for Retesting and resampling of raw materials
sop for Control of issuance of record of analysis green sheets
sop for Control of computer passwords
sop for sampling of packaging materials PM
Operation and Calibration Procedure for pH Meter
Calibration of Melting Point Apparatus
Procedure For HPLC column washing procedure
Procedure For Retest Period of Raw Material
Procedure For Training Programme For Quality Control Personnel
Standard Operating Procedure is to provide guidelines for Preparation
Procedure For Analysis of Water rinsing
Procedure For Calibration of Glassware
Procedure For Procurement and storage of reference standards
Procedure For Preparation of buffer solution
safety guidelines for laboratory
responsibilities of quality control and microbiological department
responsibilities of quality control and microbiological department
handling of corrosive hazardous and flammable chemical
calibration of volumetric glassware
operation and cleaning of water bath
testing and approval rejection of bulk finish and packed drug products
procedure for preparation of general test procedures and standard test procedures
disposal of samples chemicals glassware columns
hplc column performance evaluation storage and destruction
operation and calibration of uv cabinet
operation and calibration of refractometer
operation and calibration of digital micrometer
handling of analytical weight box