operation cleaning and maintenance of safety shower and eye washer

 

operation cleaning and maintenance of safety shower and eye washer

 

1.0 Purpose
1.1 To describe a procedure for Operation, Cleaning, and Maintenance of Safety shower and Eye washer.

2.0 Scope
2.1 This procedure is applicable for Water Safety shower and Eye washer used in Quality Control

department at A B C Pvt. Ltd. India.

3.0 Responsibility
Officer QC/ Executive QC Preparation of SOP and follow the procedure as per SOP
Assistant Manager Reviewing and Training of SOP
Manager QC Implementation of SOP
4.0 Accountability
4.1 Manager Quality Control is accountable for implementation and compliance with SOP.

5.0 Procedure
5.1 Instrument Description
5.1.1 Purpose: – It is designed so as to wash eyes and body in case of any chemical spillage due to Chemicals or Acids.

5.2 Procedure
5.2.1 In case of any chemical spillage into the Eye, run to the eye washer in the wet lab and push the foot lever to operate.
5.2.2 Hold your eye on top of the Atomizer (Dual water tab) with your finger opening the eye lash.
5.2.3 Let the water flow into the eye and flush the eye for minimum 15 minutes.
5.2.4 After this if the eye is still irritating, consult to the doctor.
5.2.5 In case of chemical spillage on analyst body, move under the head shower unit and pull the front chain to operate.
5.2.6 Let the water flow down the analyst body (from head to toe) for at least 15 min.

and after that pull the backside chain to stop the water.
5.2.7 If required, in case of major spillage immediately consult the doctor after emergency shower.
5.2.8 After completion of washing pull the chain from back side to stop the water.

5.3 Cleaning
5.3.1 Clean the surface of the eye washer and shower using a damp cloth. If necessary, a mild detergent may be used.

5.4 Maintenance
5.4.1 In case of any problems and maintenance affix “UNDER MAINTENANCE” label with

date and time and inform maintenance person for further action.
5.4.2 After completion of maintenance work remove the “UNDER MAINTENANCE” label.

6.0 Reference
In house specification

7.0 Attachments
Not Applicable

8.0 Distribution
Distribution of the controlled copy to concerned departments.
Additional copies may be issued for specific purpose.

9.0 Abbreviations
QA : Quality Assurance
QC : Quality Control
SOP : Standard Operating Procedure

10.0 Revision History

Revision No. Supersedes Effective Date Details of reason
01 Nil New SOP

 

Operating Procedure of Centrifuge

sop for preparation and storage of reagent solution

sop for Calibration and Maintenance of Laboratory Instruments and Equipment

Disposal of Residual Sample or Left Over Material

sop for for Laboratory Incident

standard operating procedure temperature monitoring

sop for drying of silica gel

sop for operation of infrared moisture balance

sop for preparation of mobile phase

sop for treated RO water

sop for Preparation and Issuance of Analysis protocol standard

sop of placebo and impurity stock solutions

sop for disposal of residual sample  

sop for handling of pharmacopoeial changes

sop for procedure for operation of ultrasonic cleaner

difference between UPLC and HPLC

sop for for Emergency Eyewash and Shower

sop for operation and calibration of total organic carbon analyzers

sop for operation of cobb tester

sop for Operation and calibration of atomic absorption spectrophotometer

sop for Operation and calibration of gas liquid chromatograph

sop for operation of humidity oven

sop for operation and calibration of serological water bath

sop for monitoring of drain trap

sop for destruction of analytical samples after testing and control samples

sop for destruction of used chemicals

Sop for Operation of suction pump

sop for Operation and calibration uv cabinet

sop for Operation and calibration of bulk density apparatus

sop for operation and calibration of shore hardness tester

sop for operation of rub proofness tester

sop for monitoring of purified water

sop for Retesting of packaging materials

sop for Retesting and resampling of raw materials

sop for Control of issuance of record of analysis green sheets

sop for Control of computer passwords

sop for sampling of packaging materials PM

Operation and Calibration Procedure for pH Meter

operation and calibration of muffle furnace

Operation of magnetic stirrer

Operation and cleaning of centrifuge

Operation and cleaning of sonicator

Operation and calibration of dissolution test apparatus

Operation and cleaning of hot air oven

Operation and calibration of bursting strength

Operating procedure of refrigerator

Operating procedure of stop watch

sop for retesting for raw materials

Operation cleaning and maintenance of water purification system

operation cleaning and maintenance of safety shower and eye washer

Leave a Comment

error: Content is protected !!