Operation cleaning and maintenance of water purification system
1.0 Purpose
1.1 To describe a procedure for Operation, Cleaning, and Maintenance of Water Purification System.
2.0 Scope
2.1 This procedure is applicable for Water Purification System used in Quality Control department at ABC Pvt. Ltd.
3.0 Responsibility
Officer QC/ Executive QC Preparation of SOP and follow the procedure as per SOP
Assistant Manager Reviewing and Training of SOP
Manager QC Implementation of SOP
4.0 Accountability
4.1 Manager Quality Control is accountable for implementation and compliance with SOP.
5.0 Procedure
5.1 Instrument Description
5.1.1 Purpose: To get highly pure water for HPLC and Reagent Preparation.
5.1.2 Type of Operation: Semi Automatic and Fully Automatic Continuously.
5.1.3 Production Flow Rate: 10 Liter/hour
5.2 Precaution
5.2.1 Do not turn off the power to the Milli-Q System when it is not used for a long time.
5.2.2 Do not remove the covers of the Milli-Q System at any time.
5.2.3 Water pressure should be always more than 2Kg/cm2.
5.2.4 Do not split the water on the keyboard.
5.3 Operation
5.3.1 Switch on the instrument from main power supply.
5.3.2 Self test will start and the screen will show message “Integrality checking please wait”.
5.3.3 Wait for 2 minutes for self testing.
5.3.4 After self testing main screen will appear.
5.3.5 It shows the current date, time and instrument status (either it is ready or standby) mode as shown below
5.3.6 Select the Ready mode by pressing, the instrument will start automatically.
5.3.7 Dispensing Water:
5.3.8 Make sure the Milli-Q System is in READY mode for the dispensing of water.
5.3.9 Select the Recirculation and press key.
5.3.10 Wait until the product water quality is optimized. Than press key.
5.3.11 Press down on the POD Unit plunger while In ready mode.
5.4 Volumetric Dispensing
5.4.1 Make sure the Milli – Q System is in ready mode.
5.4.2 Select the Volume and press key.
5.4.3 Select the desired volume of water to be delivered. Than press √ .
5.4.4 When the volumetric dispensing is finished, the system recirculates water for 3 minutes.
5.4.5 The system stops re-circulating water.
5.5 Viewing Water Quality (Milli-Q Water)
5.5.1 Make sure the Milli-Q system is in ready mode.
5.5.2 Select Menu and press .
5.5.3 Select the Milli-Q water quality and press .
5.5.4 Press three times to go the main screen.
5.5.5 Acceptance Criteria for Milli-Q water as per as table.
| S. No. | Test | Specification |
| 1. | Description | A clear ,colourless liquid |
| 2. | pH | 5.0 to 7.0 |
| 3. | Conductivity | Not more than 3.0 µS/cm |
5.6 Cleaning Procedure
5.6.1 Disconnect the Instrument from main power supply.
5.6.2 Clean the instrument externally with clean dry cloth daily.
5.6.3 Cleaning of water tank shall be done after 15 days and clean with wet cloth soaked in hot water.
5.6.4 Enter the details of the Cleaning of water tank cleaning record as per Annexure I.
5.7 Maintenance
5.7.1 Report any discrepancy observed during operation of the instrument to Manager-QC or his
authorized representative, and notify the defect to Engineering Department/ Service Engineer to
rectify the defect. Affix ‘Under Maintenance’ label on the instrument.
5.7.2 After the problem has been rectified, remove the “UNDER MAINTENANCE” label.
6.0 Reference
In house specification
7.0 Attachments
Annexure I : Elix Water Storage Tank Cleaning Record
Annexure II: pH and Conductivity Record of Milli Q Water
8.0 Distribution
Distribution of the controlled copy to concerned departments.
Additional copies may be issued for specific purpose.
9.0 Abbreviations
QA : Quality Assurance
QC : Quality Control
SOP : Standard Operating Procedure
10.0 Revision History
| Revision No. | Supersedes | Effective Date | Details of Revision |
| 01 | New SOP |


Operating Procedure of Centrifuge
sop for preparation and storage of reagent solution
sop for Calibration and Maintenance of Laboratory Instruments and Equipment
Disposal of Residual Sample or Left Over Material
sop for for Laboratory Incident
standard operating procedure temperature monitoring
sop for operation of infrared moisture balance
sop for preparation of mobile phase
sop for Preparation and Issuance of Analysis protocol standard
sop of placebo and impurity stock solutions
sop for disposal of residual sample
sop for handling of pharmacopoeial changes
sop for procedure for operation of ultrasonic cleaner
difference between UPLC and HPLC
sop for for Emergency Eyewash and Shower
sop for operation and calibration of total organic carbon analyzers
sop for operation of cobb tester
sop for Operation and calibration of atomic absorption spectrophotometer
sop for Operation and calibration of gas liquid chromatograph
sop for operation of humidity oven
sop for operation and calibration of serological water bath
sop for monitoring of drain trap
sop for destruction of analytical samples after testing and control samples
sop for destruction of used chemicals
Sop for Operation of suction pump
sop for Operation and calibration uv cabinet
sop for Operation and calibration of bulk density apparatus
sop for operation and calibration of shore hardness tester
sop for operation of rub proofness tester
sop for monitoring of purified water
sop for Retesting of packaging materials
sop for Retesting and resampling of raw materials
sop for Control of issuance of record of analysis green sheets
sop for Control of computer passwords
sop for sampling of packaging materials PM
Operation and Calibration Procedure for pH Meter
operation and calibration of muffle furnace
Operation and cleaning of centrifuge
Operation and cleaning of sonicator
Operation and calibration of dissolution test apparatus
Operation and cleaning of hot air oven
Operation and calibration of bursting strength
Operating procedure of refrigerator
Operating procedure of stop watch
sop for retesting for raw materials
Operation cleaning and maintenance of water purification system