Site icon Pharma Dekho

General Test Procedure Specific Optical Rotation

General Test Procedure Specific Optical Rotation

 

  1. Purpose
    1. This procedure is applicable to all incoming materials.
  2. Apparatus
    1. Polarimeter
    2. Polarimeter Tube
    3. Calibrated Balance
    4. Thermometer
  3. Reagents
    1. None
  4. Procedure
    1. The optical rotation, a, unless otherwise specified, is measured at the wavelength of the sodium D-line

(589.3 nm) at a temperature of 25° in a layer 1 dm thick.

Dextrorotation is designated (+) and laevorotation is designated (-).

    1. The specific optical rotation of a liquid is determined by measuring the angle of rotation at the

wavelength of the sodium D-line at a temperature of 25°, unless otherwise specified, calculating the

optical rotation with reference to a layer 1 dm thick and dividing by the relative density at 25°.

    1. The specific optical rotation of a solid substance is determined, using the solution specified in the

monograph, by measuring the angle of rotation at the wavelength of the sodium D-line at a

temperature of 25°, unless otherwise specified, and calculating the result with reference to a layer 1

dm thick of a solution containing 1 g of the substance per ml. The specific optical rotation of a solid is always expressed with reference to a given solvent and concentration.

  1. Method

Connect power plug to proper power supply through voltage stabilizer.

Switch on main power supply & on the instrument by main ON / OFF Button; allow 15 minutes for the Sodium lamp glow up.

Weigh accurately a suitable quantity of the substance in to a volumetric flask. & Prepare the solution in water, or other solvent, if specified in the monograph, reserving a portion of the solvent for the blank determination.

Transfer the prepared solution/or liquid sample to the Polarimeter tube of required length (1dcm or 2dcm) within 30 minutes from the time of solution prepared & maintain the solution / liquid sample at a temperature of 25ºC.

Polarimeter tube should be filled in such a way as to avoid air bubbles interfere with the passage source of light.

Determine the Zero point of the Polarimeter and then take at least five readings at 25ºC with the test solution. Take an equal number of readings in the same tube with the blank solution. The blank reading is subtracted from the average observed reading of test solution.

Calculate the specific optical rotation of a liquid substance, or of a solid in solution on dried basis.

Lens of Polarimeter tube should be thoroughly cleaned before taking the reading.

Polarimeter tube should be clean properly with distilled water.

Lastly off the switch of voltage stabilizer and main power supply.

Calculate the specific optical rotation using the expression =

  1. Reference
    1. IP
  2. Revision History

Revision No.

Details of Changes

Reason for change

00

New GTP

NA

End of Document

Standard testing procedure Phytomenadione injection

standard testing procedure serratiopeptidase



standard testing procedure starch IP

standard testing procedure sucrose refined sugar

standard testing procedure titanium dioxide

standard testing procedure tramadol hydrochloride

standard testing procedure zinc sulphate

standard testing procedure croscarmellose sodium

standard testing procedure colour erythrosine supra

standard testing procedure magnesium hydroxide

standard testing procedure diclofenac sodium

standard testing procedure dibasic calcium phosphate

standard testing procedure cyanocobalamin

standard testing procedure cholecalciferol

standard testing procedure Calcium carbonate oyster shell powder

standard test procedure Calcium Citrate

standard testing procedure Bronopol

standard testing procedure Bromhexine Hydrochloride

Standard Testing Procedure diclofenac sodium injection

Standard Testing Procedure Drotaverine Hydrochloride injection

analysis of Procyclidine HCl 5 mg Tablet

analysis of silodosin 8 mg capsule

Method of Analysis Ceftriaxone and Sulbactam For Injection

Analysis of Glimepiride and Metformin Hydrochloride (SR) Tablets

General Test Procedure Identification by IR

General Test Procedure Identification by Solubility

General Test Procedure Specific Optical Rotation

 

 

 

Exit mobile version