calibration of volumetric glassware
1.0 Purpose
1.1 To describe a procedure for Calibration of Volumetric Glassware.
2.0 Scope
2.1 This standard operating procedure is applicable for calibration of volumetric glassware in
Quality Control laboratory at ABC Pvt. Ltd., Roorkee (UK), India.
3.0 Responsibility
Officer QC/ Executive QC Preparation of SOP
Assistant Manager Reviewing and Training of SOP
Manager QC Implementation of SOP
4.0 Accountability
4.1 Manager Quality Control is accountable for implementation and compliance with SOP.
5.0 Procedure
5.1 All the volumetric glassware shall be procured from standard glassware manufacturers.
5.2 Wash the Glass Apparatus as per respective SOP.
5.3 Finally wash and Rinse the Glassware as per respective SOP.
5.4 Take the Random sample of Glassware ‘B’ class for the Calibration.
Note: For Class ‘A’ type Glassware, No calibration is required.
5.5 Calibration of Volumetric Flask:
5.5.1 Weigh the Clean and Dry Volumetric Flask / Measuring Cylinder. Let the weight be ‘X’
5.5.2 Fill the Flask with Purified Water up to the mark considering the Lower Meniscus. Weigh the Flask. Let the weight be ‘Y’.
5.5.3 Find out the weight of Purified Water by the following Formula
Weight of Purified water = Y-X = A.
5.5.4 Calculate Capacity of the Volumetric Flask by using following Formula.
– A
Capacity of Volumetric Flask (ml) =—–
-D
Where;
A = Weight of Purified Water
D = Density of Water at 25oC
5.5.5 If the Capacity of the Flask is within the Limits specified in Annexure I then accept the respective Glassware.
If the capacity of the Flask is not within the Limits specified in respective Annexure I then reject the respective Glassware.
5.6 Calibration of One Mark Volumetric Pipettes and Graduated Pipettes
5.6.1 Weigh the Clean and Dry Beaker. Let the Tare Weight be ‘X’
5.6.2 Fill the Pipette to the Lower Meniscus with Purified Water. Empty out the Purified Water in the Beaker.
Weigh the Beaker. Let the weight be ‘Y’.
5.6.3 Find out the weight of the Purified Water by the following Formula.
Weight of the Purified Water = Y-X =A
5.6.4 Calculate Capacity of the Pipette using following Formula
—-A
Capacity of pipette (ml) = —–
—D
Where;
A = Weight of Purified Water
D = Density of Water at 25oC
5.6.5 If the Capacity of the Pipette is within the limits specified in Annexure II then accept the respective
Glassware. If the Capacity of the Pipette is not within the limits specified in respective
Annexure II then reject the respective Glassware.
5.7 Calibration of Burette
5.7.1 Weigh a Clean and Dry Beaker. Let the Tare Weight be ‘X’.
5.7.2 Fill the Burette up to the Lower Meniscus of zero mark with Purified Water. Empty out
the Purified Water in the Beaker and weigh. Let the Weight be ‘Y’.
5.7.3 Find out the weight of the Purified Water by the following Formula
Weight of the Purified Water = Y-X = A
5.7.4 Calculate Capacity of the Burette using following Formula
–A
Capacity of Burette(ml) =—-
–D
Where;
A = Weight of Purified Water
D = Density of Water at 25oC
5.7.5 If the Capacity of the Burette is within the limits specified in Annexure III then accept the respective Glassware.
If the Capacity of the Burette is not within the limits specified in respective Annexure III then reject the respective Glassware
6.0 Reference
IP/ In House Specification
7.0 Attachments
Annexure: I Calibration Data Sheet for Volumetric Flasks
Annexure: II Calibration Data Sheet for Pipette.
Annexure: III Calibration Data Sheet for Burette.
8.0 Distribution
Distribution of the controlled copy to concerned departments.
Additional copies may be issued for specific purpose.
9.0 Abbreviations
QA : Quality Assurance
QC : Quality Control
SOP : Standard Operating Procedure
10.0 Revision History
| Revision No. | Supersedes | Effective Date | Details of reason |
| 01 | Nil | New SOP |
Annexure: I Calibration Data Sheet for Volumetric Flasks

Annexure: II Calibration Data Sheet for Pipette.

Annexure: III Calibration Data Sheet for Burette.

Operating Procedure of Centrifuge
sop for preparation and storage of reagent solution
sop for Calibration and Maintenance of Laboratory Instruments and Equipment
Disposal of Residual Sample or Left Over Material
sop for for Laboratory Incident
standard operating procedure temperature monitoring
sop for operation of infrared moisture balance
sop for preparation of mobile phase
sop for Preparation and Issuance of Analysis protocol standard
sop of placebo and impurity stock solutions
sop for disposal of residual sample
sop for handling of pharmacopoeial changes
sop for procedure for operation of ultrasonic cleaner
difference between UPLC and HPLC
sop for for Emergency Eyewash and Shower
sop for operation and calibration of total organic carbon analyzers
sop for operation of cobb tester
sop for Operation and calibration of atomic absorption spectrophotometer
sop for Operation and calibration of gas liquid chromatograph
sop for operation of humidity oven
sop for operation and calibration of serological water bath
sop for monitoring of drain trap
sop for destruction of analytical samples after testing and control samples
sop for destruction of used chemicals
Sop for Operation of suction pump
sop for Operation and calibration uv cabinet
sop for Operation and calibration of bulk density apparatus
sop for operation and calibration of shore hardness tester
sop for operation of rub proofness tester
sop for monitoring of purified water
sop for Retesting of packaging materials
sop for Retesting and resampling of raw materials
sop for Control of issuance of record of analysis green sheets
sop for Control of computer passwords
sop for sampling of packaging materials PM
Operation and Calibration Procedure for pH Meter
Calibration of Melting Point Apparatus
Procedure For HPLC column washing procedure
Procedure For Retest Period of Raw Material
Procedure For Training Programme For Quality Control Personnel
Standard Operating Procedure is to provide guidelines for Preparation
Procedure For Analysis of Water rinsing
Procedure For Calibration of Glassware
Procedure For Procurement and storage of reference standards
Procedure For Preparation of buffer solution
safety guidelines for laboratory
responsibilities of quality control and microbiological department
responsibilities of quality control and microbiological department