GLASSWARE CLEANING VALIDATION
Name of Product | GLASSWARE CLEANING VALIDATION
|
Parameter | GLASSWARE CLEANING VALIDATION
|
Method followed | BY UV |
Supersede No. | Nil |
Total No of pages | |
Effective Date |
Sr. No. | Title | Page |
— | Product Description | |
— | Table of contents | |
1.0 | Document Preparation and Protocol Approval | |
2.0 | Revision Control History | |
3.0 | Objective | |
4.0 | Scope | |
5.0 | Reason For Validation | |
6.0 | Responsibility | |
7.0 | Description of Method Belonging | |
8.0 | Training Details | |
9.0 | Glassware Cleaning Validation | |
10.0 | Cleaning glassware | |
11.0 | Preparation of Solution | |
12.0 | Procedure | |
13.0 | Erumpent/ Instrument Required during Validation | |
14.0 | Acceptance criteria | |
15.0 | References |
DOCUMENT PREPARATION AND PROTOCOL APPROVAL
Preparation and Approval of this Cleaning Validation document will be joint responsibility of the following functional area. Any modification in this document shall be documented and approved
Prepared By:
Name | Designation | Signature | Date |
Executive – QC |
Reviewed By:
Name | Designation | Signature | Date |
Executive – QC | |||
Manager – QC |
Approved By:
Name | Designation | Signature | Date |
Head – QA |
Revision Control History:
Effective Date | Version no. | Superseded no. | CCF no. | Reason for Review |
00 | Nil | NA | New documents |
OBJECTIVE:
To perform Glassware Cleaning Validation provides documentary evidence that the method used for Glassware Cleaning Validation of said product full fill the regulatory & compendia requirement. This protocol presents and discusses the result obtained in the Validation study.
4.0 SCOPE:
This document is applicable for the Glassware Cleaning Validation of UV Method which is applying at ABC Company
5.0 REASON FOR VALIDATION:
Regulatory Requirement
6.0 RESPONSIBILITY:
The following functions are involved in the Analytical method Validation performing and shall carry out the following responsibilities.
6.1 Quality Control:
6.1.1 Officer & Executive Quality Control: To perform the analysis in the Verification and evaluate the data as per protocol.
6.1.2 Sr. Officer: To prepare the protocol and review all data evaluated as per protocol.
6.1.3 Manager Quality Control: To provide technical support, review the protocol and ensure that all parameter performed as per protocol.
6.2 Regulatory:
6.2.1 To provide technical support, review and approved the protocol and ensure that all parameter performed as per protocol.
7.0 DESCRIPTION OF METHOD BELONGING:
7.1 Product Detail:
7.1.1 Name of Product: GLASSWERE CLEANING VALIDATION
7.1.2 Generic Name: GLASSWERE CLEANING VALIDATION
7.2 Method Details:
Methodology: UV
8.0 TRAINING DETAILS:
Training shall be imparted to all concern persons involved in Glassware cleaning validation activity w.r.t. cGMP
and concerned analytical technique required for analysis and details will be attached in Glassware cleaning
validation report’ activity, as per below mentioned format.
Purpose | To train all personnel involved in the execution of this protocol | ||
Topic | · Purpose
· Procedure · Method details · Acceptance criteria · Documentation etc. |
||
S. No. | Name of Participants | Trainees area of concern | Sign/Date |
DETAILS OF GLASSWERE CLEANING VALIDATION
10.0 Preparation of cleaning solutions
10.1 For the cleaning of glassware, those are used for chemical analysis special precautions will take place to avoid any chemical residue in glassware, which may later interfere with the results of chemical analysis.
Following cleaning solutions were used for cleaning of glassware.
10.2 2% Liquid soap solution
10.2.1 Measure 20 ml soap solution and add it to container containing 980 ml of purified water and stir it well.
10.2.2 Record the solution preparation details in glassware cleaning solution preparation record.
10.3 Cleaning glassware
10.3.1 Cleaning of glassware, which has contained hazardous materials, must be solely undertaken by experienced personnel.
10.3.2 Most new glassware is slightly alkaline in reaction. For precision chemical tests, new glassware should be soaked several hours in acid water (1% solution hydrochloric acid or nitric acid) before washing.
10.3.3 Wash glassware as quickly as possible after use if it is not possible then the articles should be allowed to soak in water.
10.3.4 For cleaning of glassware such as bottles, flasks, beakers, test tubes, etc use 2% liquid soap solution. For better results use hot water.
10.3.5 During the washing all parts of the article should be thoroughly scrubbed with a brush selected for the shape and size of the glassware. Brushes should always be in good condition to avoid any abrasion of the glassware.
10.3.6 Special types of precipitate material may require removal with nitric acid, aqua regia or fuming sulphuric acid. These are very corrosive substances and should be used only when required.
10.3.7 Before cleaning of glassware remove the labeling of marker pen with the help of IPA or acetone. For plastic ware do not use acetone for removing marker pen labeling.
10.3.8 It is imperative that all soap detergent and other cleaning fluids be removed from glassware before use. This is especially important with detergents, slight traces of which will interfere with serological and culture reactions.
10.3.9 After cleaning, thoroughly rinse with tap water ensuring that containers are partly filled with water, shaken and emptied several times. Finally rinse with purified water.
10.3.10 After cleaning dry the glassware in Hot air oven at 600 temperature ± 50.
10.3.11 Always protect clean glassware from dust by use of temporary closures or by placing in a dust free cabinet
11.0 Type of Analysis: BY UV
11.1 Preparation of Solutions:
11.2 Blank Solution Preparation:
11.3 Sample Solution Preparation:
11.5 For Blank Samples:
The Purified water as a blank scans for 200nm to 400 nm riding to UV Spectrophotometer.
11.6 For Glassware Rinse Samples:
After dry the Glassware cleaning for purified water (rinse sample) scan for 200 nm to 400 nm riding to UV Spectrophotometer.
The glassware cleaning sample absorbance not more than blank sample absorbance.
11.7 Procedure:
Switch on the main power supply
After getting the initialization of instrument About 15 minutes, instrument Triple beep sound.
Than Press enter and again sound of tree beep, Open the software on system UV-Visible analyst display on the system screen.
Display system status, close click on file menu, click new and select fixed point measurement for single wavelength and multiple wavelengths.
Click on file menu, click new and select Wavelength scan measurement for scanning.
For scan click wavelength scan measurement fill up the blank medium in booth cuvette and click z for both cuvette matching.
Then fill sample in sample cuvette and click red arrow.
11.8 Equipment/Instrument required during validation
Sr. No. | Name of Equipment | Equipment No. | Calibration Status | Make/Model | |
Calibration on | Calibration Due | ||||
1. | UV Spectrophotometer | ||||
11.9 Reagents and Chemicals
Sr. No. | Name of Reagent | Grade | Make | Lot No. |
1. | WATER | |||
2. | ph paper strips |
12.0 ACCEPTANCE CRITERIA
Sr. No. | Title | Acceptance Criteria | Observations/Conclusions |
1. | ABSORBANCE |
The glassware cleaning sample absorbance not more than blank sample absorbance. |
process validation protocol for methylcobalamin niacinamide and pyridoxine injection
validation protocol of sterility test
sop for Analytical Method Transfer
Protocol for hold time study of sterile garments
sop for swab sampling for validation of clean surfaces
cleaning validation maco and noel calculation formula
sop for performance qualification for analyst
Sop for Validation report for disinfectant efficacy
Sop for Method validation report for bacterial endotoxin test
Sop for method validation microbiology sterility testing
sop for validation report for preservative efficacy test
sop for Protocol cum report for efficacy qualification of uv light
sop for validation protocol for uv light efficacy of dpb & laf
sop for cleaning validation protocol tablet manufacturing equipment
sop for cleaning validation protocol ointment manufacturing equipment
concurrent process validation for rabeprazole ec and domperidone sr capsules
sop for Validation for cleaning procedure liquid injection
sop for Validation for cleaning procedure dry powder injection