sop for operation and calibration of dissolution Apparatus
1.0 OBJECTIVE
1.1 To describe a procedure for operation & calibration of Dissolution Apparatus.
2.0 SCOPE
2.1 This procedure is applicable for operation & calibration of Dissolution Apparatus to the Quality control department
3.0 RESPONSIBILITY
3.1 Officer/Executive – QC
4.0 ACCOUNTABILITY
4.1 Head – QC
5.0 PROCEDURE
5.1 Operating Procedure
5.1.1 Place the instrument on a table.
5.1.2 Place the bath and connect the tubes from the Inlet & Outlet of the temperature Controller.
5.1.3 Connect the Power Cord and Printer Cable.
5.1.4 Lift the stirrer Unit. Insert the shaft rod as per the method selected.
5.1.5 Place the Test Sample in the test vessel.
5.1.6 Press the DOWN Arrow key from the front panel to lower down the stirrer unit. The stirrer unit will adjust automatically
to required height depending on the test method selected.
5.1.7 Press the START key from the front panel to start the test. The instrument will now show the Run Time on screen.
5.1.8 The Run Time on screen will display the Protocol Running (Refer Idle Screen), The step No. (Refer option (c) in Parameter Setting),
The Elapsed Sampling Time for that particular Step
(Refer option (c) in Parameter setting), Actual RPM of stirrer (Refer option (a) in Parameter Setting), Actual Bath Temperature
and the Temperature being measured by the external probe (If Connected). The RUN LED on the Front Panel will glow indicating the instrument is in RUN Condition.
5.2 Procedure for Cleaning:
5.2.1 Switch OFF the power Supply of the instrument before removing water from the Bath.
5.2.2 Do not use aggressive solvents or material for cleaning the Bath and the Test Vessels.
5.2.3 Tube and Bath Cleaning: Fill half bath with liquid soap and circulate. Frequencyof cleaning once in a week.
This will prevent algae propagation in the bath.
5.2.4 Filters should be sonicated after every test and once in a week it should be flushed with methanol.
5.2.5 In case of any discrepancy or malfunction, attach status label and inform Maintenance department for cleaning.
5.3 Calibration
5.3.1 Physical Calibration
5.3.1.1 Calibration frequency: Quarterly or after major break down.
5.3.1.2 Calibration of bath temperature: Measured by the calibrated thermometer.
5.3.1.3 Calibration of vessel temperature using calibrated thermometer.
5.3.1.4 Permissible limit of temperature inside bath & vessel as per IP: 37.0°C + 0.5°C.
5.3.1.5 Calibration of RPM: Calibrate the RPM using calibrated tachometer.
5.3.1.6 Set one by one 50 & 100 RPM and check the actual RPM using calibrated tachometer.
5.3.1.7 Tolerance limit of RPM is + 4% of the reading.
5.3.1.8 Calibration of distance between blade and inside vessel as per IP: 25 mm + 2 mm
5.3.1.9 Checking of wobbling and distance parameters for the apparatus
5.3.1.10 Wobbling: Fix the paddle/basket apparatus to the instrument, lower the apparatus.
5.3.1.11 Fix the wobbling meter on the top of the bowl.
5.3.1.12 The needle of wobbling meter should touch to the surface of the apparatus.
5.3.1.13 Measure the wobbling on the meter.
5.3.1.14 Permissible limit of wobbling is NMT 1 mm on 360º revolution.
5.3.1.15 Distance from the apparatus: Fix the paddle/basket apparatus to the instrument, lower the apparatus.
5.3.1.16 Insert the Depth gauge between the bottom of the apparatus and the bottom of the bowl.
5.3.1.17 Measure the reading on the display.
5.3.1.18 Calibration of Timer: Set the time in instrument to 30 minutes. Start the equipment &stopwatch simultaneously. Note down the
time on the stopwatch when the set time is off displayed by equipment.
5.3.1.19 Jar Centering: Loosen the base plate of each jar, fix the centering device with spindle, lower down the spindle inside
the jar (Jar is automatically centered) and tighten the base plate of each jar. Fix the wobbling gauge in each jar,
rotate by 360° and record the reading.
5.3.1.20 Permissible limit of Jar Centering is NMT 2 mm on 360º revolution.
5.3.2
Performance Verification Test
5.3.2.1 Calibration frequency: Half yearly or after major maintenance job.
5.3.2.2 Performance Verification Test using IP Prednisone tablets.
5.3.2.3 Use disintegrating type Prednisone tablet having 10 mg claim.
5.3.2.4 Performance Verification Test applied for apparatus 1st and apparatus 2nd both.
5.3.2.5 Dissolution System :
Apparatus: Use specified type.
Duration: 30 minutes Speed: 50 rpm Temperature: 37.0°C + 0.5°C
Dissolution medium: 500 ml of purified water
5.3.2.6 Preparation of dissolution medium: Heat 500 ml of water for each vessel at 45°C and filtered with help of vacuum for
additional 5 minute continuous stirring. The temperature of
dissolution medium should be maintained at 37.0°C + 0.5°C prior to the initiation of test.
5.3.2.7 Preparation of standard solution in duplicate A & B: Weight accurately 0.030 gm of USP Prednisone working standard and
transfer into 200 ml volumetric flask. Added 10 ml ethanol and sonicate to dissolve if required. Dilute to volume with dissolution medium and mix.
Further dilute 5 ml of the resulting solution to 50 ml with dissolution medium and mix.
5.3.2.8 System suitability requirements: The test is not valid unless, the similarity factor between standard solutions
A & B (Separately prepared) is between 0.98 to 1.02. Calculate the similarity factor using following formula :
Absorbance of Prednisone Standard A obtain x weight of IP Prednisone WS with standard B
Absorbance of Prednisone Standard B obtain weight of IP Prednisone WS with standard A
(Note: If the similarity factor does not fall within 0.98 to 1.02, prepare fresh standard solution in duplicate and calculate similarity factor again as above.)
5.3.2.9 Preparation of sample solution: carry out the dissolution using six tablets. Transfer one tablet each separately into individual jar
containing 500 ml of dissolution medium. Ensure the temperature of all jars at the starting and the end of the set time using a calibrated
thermometer or temperature sensor. Set the stated parameters and carry out dissolution. At the end of the set time, immediately
withdraw 10 ml of the aliquot (the zone midway between the surface of the dissolution medium and the top of the rotating paddle/basket,
not less than 1 cm from the jar wall) and filter (0.45 µm syringe filter).
5.3.2.10 Procedure: Determine the absorbance of the standard solution and the sample solution at 240 nm using dissolution
medium as a blank. Calculate the percentage of Prednisone released per tablet by using the following formula.
Sample absorbance x weight of standard (mg) x 5 x 500 x 100 x purity of std. Absorbance of standard A x 200 x 50 x 1 x Label claim of tablet in mg x 100
5.3.2.11 Record the percentage dissolve obtained at the end of set time specified for the apparatus. Transform the percentage dissolve results to the natural log scale, and then determine the mean (M1) and variance (V1).
5.3.2.12 Convert result obtained in step 2 to a geometric mean (GM1) and percent coefficient of variation (%CV1). Compare the result obtained with the 1st stage of two-stage acceptance criteria.
5.3.2.13 Acceptance criteria: The GM must not fall outside the limit and % CV must not be greater than the limit and mentioned in the
certificate of IP Prednisone Tablets Reference Standard.
5.3.2.14 If the result of point 5.3.2.12 satisfy both acceptance criteria, the assembly has passed the Performance Verification test.
If the result obtained with 1st stage of two stages are not within acceptance criteria then follow the bellow procedure.
5.3.2.15 Carry out the dissolution with an additional set of tablets and repeat the point 5.3.2.9 and 5.3.2.10 again.
Transform the percentage dissolved results to the data obtained at this step.
5.3.2.16 Considering the average of the two mean (i.e. M1 & M2) and variance (i.e. V1 & V2) obtained in
point 5.3.2.11 and point 5.3.2.15 convert the result to geometric mean (GM) and percentage coefficient of variation (% CV).
5.3.2.17 Acceptance criteria: Compare the results obtained in point 5.3.2.16 with the 2nd stage of Two-Stage acceptance
criteria mentioned in the certificate of IP Prednisone tablets reference standard. The GM must not fall outside the limit and %CV must not be greater than the limit.
If both meet the acceptance criteria, the assembly has passed the performance verification test (PVT).
For calculation of geometric mean (GM) and % coefficient of variation (%CV) use the following formula and calculate the result.
For example :
Run 1 : X1,X2,……………Xn in natural log scale : LnX1,LnX2………..LnXn
Run 2 : Xn+1,Xn+2,……………X2n in natural log scale : LnXn+1,LnXn+2………..LnX2n
1st Stage of Two-Stage : For n = 6,7,8
GM1 = exp(average(Ln X1 : LnXn ) )
%CV1 = 100*sqrt(exp(var(Ln X1 : LnXn ) ) -1)
2nd Stage of Two-Stage :
GM = exp(avg(avg(Ln X1 : LnXn ) ), (avg( LnXn+1 : LnX2n ))))
= exp(avg(Ln X1 : LnX2n ))
%CV = 100*sqrt(exp(avg((var(Ln X1 : LnXn )),(var( LnXn+1 : LnX2n ))))-1)
Where,
exp : exponential
var : variance
sqrt : square root
100 : conversion factor to percentage
Ln : Natural log
avg : average
5.3.3 Calibration check by the external party :
5.3.3.1 Calibration frequency: yearly
5.3.3.2 Calibration of bath & vessel temperature: temperature controller with sensor reading.
5.3.3.3 Permissible limit of temperature inside bath & vessel as per IP: 37.0°C + 0.5°C
5.3.3.4 Calibration of RPM: Calibrate RPM using tachometer.
5.3.3.5 Set one by one 25, 50, 75,100,150 & 200 RPM and check the actual RPM using calibrated tachometer.
5.3.3.6 Tolerance limit of RPM is + 4% of the reading.
6.0 REFERENCE
6.1 In house
7.0 ENCLOSURES
Dissolution Apparatus Calibration
Prednisone tablet receiving & consumption record
8.0 FORMATS
8.1 QC : Quality Control
8.2 QA : Quality Assurance
8.3 QCA : Quality Control Analytical
8.4 SOP : Standard Operating Procedure
8.5 mg : Miligram
8.6 °C: Degree Centigrate
8.7 % : Percentage
8.8 IP: Indian Pharmacopoeia
8.9 ml: Milliliter
8.10 µm: Micrometer
8.11 µ: Micron
8.12 NMT: Not more than
8.13 PVT: Performance Verification Test
8.14 RPM: Round Per Minutes
8.15 mm: Micrometer
8.16 GM : Geometric mean
8.17 CV : Coefficient of variation
8.18 cm : Centimeter
8.19 gm : Gram
sop for Calibration and Maintenance of Laboratory Instruments and Equipment
Disposal of Residual Sample or Left Over Material
sop for for Laboratory Incident
standard operating procedure temperature monitoring
sop for operation of infrared moisture balance
sop for preparation of mobile phase
sop for Preparation and Issuance of Analysis protocol standard
sop of placebo and impurity stock solutions
sop for disposal of residual sample
sop for handling of pharmacopoeial changes
sop for procedure for operation of ultrasonic cleaner
difference between UPLC and HPLC
sop for for Emergency Eyewash and Shower
sop for operation and calibration of total organic carbon analyzers
sop for operation of cobb tester
sop for Operation and calibration of atomic absorption spectrophotometer
sop for Operation and calibration of gas liquid chromatograph
sop for operation of humidity oven
sop for operation and calibration of serological water bath
sop for monitoring of drain trap
sop for destruction of analytical samples after testing and control samples
sop for destruction of used chemicals
Sop for Operation of suction pump
sop for Operation and calibration uv cabinet
sop for Operation and calibration of bulk density apparatus
sop for operation and calibration of shore hardness tester
sop for operation of rub proofness tester
sop for monitoring of purified water
sop for Retesting of packaging materials
sop for Retesting and resampling of raw materials
sop for Control of issuance of record of analysis green sheets
sop for Control of computer passwords
sop for sampling of packaging materials PM
sop for sampling of sterile raw material
sop for sampling of intermediates and finished products
sop for operation and calibration of friability test apparatus
sop for approval and rejection of packaging materials
sop for non conformance of RM PM and finished product
sop for collection storage and disposal of control samples
sop for trend analysis of finished products
sop for Chromatographic practices and system suitability
SOP For Good Laboratory Practices
sop for cleaning and operation of sieve shaker
general specification of packing material cartons
sop for Password for Analytical Instrument and LIMS software
sop for Rounding off numerical analytical results
sop for sampling of bulk and finished product
sop for cleaning of spillage material
sop for Handling of Reference Standard
sop for hplc column maintenance and washing
procedure for sampling and handling of bulk sample
STP for borewell generation point (raw water storage tank)
sop for preparation and standardization of 0.1M Zinc Sulphate
Operation &calibration of analytical balance (dhona)
Operation and Calibration Procedure for Disintegration Test apparatus
sop for preparation and standardization of 1 M Hydrochloric Acid
Preparation and standardization of 0.1 M ceric ammonium sulphate solution
sop for preparation and standardization of 0.05 m iodine solution
validation of volumetric solution 0.1m ammonium thiocyanate
handling of reference standard and preparation of working standard
sop for water sampling and analysis
sop for operation for validation of excel worksheets
sop for stability of volumetric solutions
sop for preparation of raw material in process finish product packing material data sheets
sop for handling of hazardous chemicals
sop for handling of glassware and allocation of identification number
sop for operation cleaning and calibration of bursting strength tester
sop for rounding off the analytical test results
procedure for Analyst Qualification
sop for operation and calibration of dissolution Apparatus
procedure for maintenance of desiccators
sop for for hplc column receipt checking id no and regeneration
safety data sheet for laboratory chemicals
procedure for handling of poisonous chemicals
sop for cleaning of sampling devices
sop for calibration procedure of instruments
sop for specification and standard testing raw material packing material and finished product
procedure for operation and calibration of potentiometric titrator
procedure for operating and calibration of digital hardness tester
procedure for disposal of expired chemicals, reagents and solvents
sop for behavior in quality control department
sop for preparation and standardization 0.1M sodium thiosulphate
sop for preparation and standardization 0.1M Disodium Edetate
preparation and standardization 0.1M Sodium Hydroxide Solution
Preparation and standardization of 0.1M Perchloric acid solution
sop for preparation 0.05M edetate disodium
sop for preparation 0.1M silver nitrate
sop for Operation and Calibration of High Performance Liquid Chromatography
sop for UV & Visible Spectrophotometer
procedure for Cleaning of laboratory glassware
Cleaning of Instrument, Instrument bench and surrounding area of Quality
Safety Precaution in Quality Control Department
Operation & Calibration of Analytical Balance
Calibration of Glassware in Quality Control Department
handling of samples received in Quality Control
Cleaning and Operation of Refrigerator
Operation, Cleaning and Calibration of water bath
Operation & Calibration of Refractometer
Operation and Cleaning of Centrifuge Apparatus
cleaning, operation & calibration of Vernier caliper
Calibration of Fourier Transform Infrared Spectrophotometer (FTIR)
Cleaning and operation of Moisture Analyzer
Cleaning & Operation of Vacuum pump in Quality Control Department
Operation and Calibration of Polarimeter
Cleaning and operation of Magnetic Stirrer
Cleaning Operation and Calibration of Melting Point
Operation Cleaning and Calibration of Muffle Furnance in Quality Control Department
procedure of operation and Cleaning of Sonicator
Operation Cleaning & Calibration of pH meter in Quality Control Department