sop for treated RO water
STORAGE REQUIREMENTS
Store at room temperature in a well closed stainless steel or glass container.
a) Analysis samples drawn for chemical analysis within 24 hours after sampling.
b) Analysis samples drawn for microbiological analysis within 8 hours after sampling. Refrigerate the samples (between 2°C to 8°C) until testing is commenced.
SAMPLING:
A) For Chemical Analysis:
Collect the sample in clean, glass stopper bottle of 800 ml capacity from DM Water plant / usage point with utmost care to ensure that no contamination occurs at the time of collection or prior examination. Remove the stopper of the bottle and handle carefully to avoid contamination (preferably hold in hand). While collecting samples allow the sample to run to Water for sufficient time as per validated drain time to permit clearing of service line. The glass bottles must be thoroughly rinsed three times with the same Water before sampling. Restrict the flow from the tap so that the bottle can be filled without splashing. During sampling, hold the bottle near the base, do not touch the stopper and the neck of the bottle. Replace the stopper immediately.
For conductivity test sample as above but by collecting the sample in a tight container with minimum head space and test in a timely manner to minimize the impact of organic contamination on the closure and container.
B) For Microbiological Analysis :
Sampling bottles : Collect the sample in non –reactive borosilicate screw capped glass bottles of 200ml capacity that have been cleaned and rinsed carefully ,i.e. final rinse with RO Water and sterilized by using a steam sterilized by using a steam sterilizer at 121°C for validated time period.
SAMPLING:
Sampling Procedure: Collect the sample from DM Water plant / usage point with utmost care to ensure that no contamination occurs at the time of collection or prior to analysis. Remove the cap of the bottle and handle carefully to avoid contamination (preferably hold in hand). While collecting samples allow the sample to run to waste for sufficient time as per validated drain time to permit clearing of service line. Restrict the flow from tap so that the bottle can be filled without splashing. During sampling, hold the bottle near the base and fill without rinsing. Do not touch the cap and the neck of the bottle. Replace the cap immediately.
Note : Do not open the sampling bottle except at the time of sampling.
QUANTITY OF SAMPLE FOR ANALYSIS
For Chemical analysis : About 800ml
For Microbiological analysis : About 200ml
QUANTITY OF RESERVE SAMPLE:
Not applicable
CATEGORY
Pharmaceutical aid (Solvent)
.
Description: A Clear, colorless, tasteless &, odorless liquid.
pH: Not less than 5.0 and not more than 7.0
Conductivity: Not more than 5.1 µS/cm at 25°C
Chlorides: Shall be free from chlorides.
Total Dissolved Solids: Not more than 0.001% (1mg).
MICROBIOLOGICAL CONTAMINATION:
Total viable count:
Not more than 500 cfu/ml
Test for Pathogens:
Escherichia coli : Should be Absent / 100ml
Salmonella : Should be Absent / 100ml
Pseudomonas aeruginosa : Should be Absent / 100ml
Staphylococcus aureus : Should be Absent / 100ml
Description: Examine the sample for clarity, color and odour.
Reporting: Report as Complies / does not comply
pH: Take 100ml of the sample and determine the pH with previously calibrated pH meter.
Reporting: Report as Complies / does not comply
Limit: Between 5.0 and 7.0.
Conductivity: The Water conductivity can be determined by the following procedure:
1. Transfer a sufficient amount of water (100ml or more to a suitable container, and stir the test specimen. Adjust the temperature, if necessary, and while maintaining it at 25°C (+/-) 1°C, begin vigorously agitating the test specimen while periodically observing the conductivity. When the change in conductivity (due to uptake of atmospheric carbon dioxide) is less than a net of 0.1µS/cm per 5 minutes, note the conductivity.
2. If the conductivity is not greater than 5.1µS/cm , the Water meets the requirements of the test for conductivity.
3. Perform this test within approximately 5 minutes of the conductivity determination, while maintaining the sample temperature at 25°C (+/-) 1°C.
If either the measured conductivity is greater than this value or the pH is outside the range of 5.0 to 7.0, the Water
does not meet the requirement of the test for conductivity.
Reporting: Report as a value in µS/cm (micro Siemen per centimeter) along with the respective pH value.
Chlorides: Take 10ml sample add 1ml of 2M Nitric acid and 0.2ml of 0.1M Silver Nitrate solution, the appereance of solution does not change for at least 15 mibnutes.
Reporting: Report as Complies / does not comply
Total Dissolved Solids: Evaporate to dryness ………. (250.0 ml) of the water sample and dry the residue at 105ºC till to constant weight.
DOCUMENT CHANGE HISTORY
Revision No. | Changes |
00 | First Time Prepare. |
END OF DOCUMENT
sop for Calibration and Maintenance of Laboratory Instruments and Equipment
Disposal of Residual Sample or Left Over Material
sop for for Laboratory Incident
standard operating procedure temperature monitoring
sop for operation of infrared moisture balance
sop for preparation of mobile phase
sop for Preparation and Issuance of Analysis protocol standard
sop of placebo and impurity stock solutions
sop for disposal of residual sample
sop for handling of pharmacopoeial changes
sop for procedure for operation of ultrasonic cleaner
difference between UPLC and HPLC
sop for for Emergency Eyewash and Shower
sop for operation and calibration of total organic carbon analyzers
sop for operation of cobb tester
sop for Operation and calibration of atomic absorption spectrophotometer
sop for Operation and calibration of gas liquid chromatograph
sop for operation of humidity oven
sop for operation and calibration of serological water bath
sop for monitoring of drain trap
sop for destruction of analytical samples after testing and control samples
sop for destruction of used chemicals
Sop for Operation of suction pump
sop for Operation and calibration uv cabinet
sop for Operation and calibration of bulk density apparatus
sop for operation and calibration of shore hardness tester
sop for operation of rub proofness tester
sop for monitoring of purified water
sop for Retesting of packaging materials
sop for Retesting and resampling of raw materials
sop for Control of issuance of record of analysis green sheets
sop for Control of computer passwords
sop for sampling of packaging materials PM
sop for sampling of sterile raw material
sop for sampling of intermediates and finished products
sop for operation and calibration of friability test apparatus
sop for approval and rejection of packaging materials
sop for non conformance of RM PM and finished product
sop for collection storage and disposal of control samples
sop for trend analysis of finished products
sop for Chromatographic practices and system suitability
SOP For Good Laboratory Practices
sop for cleaning and operation of sieve shaker
general specification of packing material cartons
sop for Password for Analytical Instrument and LIMS software
sop for Rounding off numerical analytical results
sop for sampling of bulk and finished product
sop for cleaning of spillage material
sop for Handling of Reference Standard
sop for hplc column maintenance and washing
procedure for sampling and handling of bulk sample
STP for borewell generation point (raw water storage tank)
sop for preparation and standardization of 0.1M Zinc Sulphate
Operation &calibration of analytical balance (dhona)
Operation and Calibration Procedure for Disintegration Test apparatus
sop for preparation and standardization of 1 M Hydrochloric Acid
Preparation and standardization of 0.1 M ceric ammonium sulphate solution
sop for preparation and standardization of 0.05 m iodine solution
validation of volumetric solution 0.1m ammonium thiocyanate
handling of reference standard and preparation of working standard
sop for water sampling and analysis
sop for operation for validation of excel worksheets
sop for stability of volumetric solutions
sop for preparation of raw material in process finish product packing material data sheets
sop for handling of hazardous chemicals
sop for handling of glassware and allocation of identification number
sop for operation cleaning and calibration of bursting strength tester
sop for rounding off the analytical test results
procedure for Analyst Qualification
sop for operation and calibration of dissolution Apparatus
procedure for maintenance of desiccators
sop for for hplc column receipt checking id no and regeneration
safety data sheet for laboratory chemicals
procedure for handling of poisonous chemicals
sop for cleaning of sampling devices
sop for calibration procedure of instruments
sop for specification and standard testing raw material packing material and finished product
procedure for operation and calibration of potentiometric titrator
procedure for operating and calibration of digital hardness tester
procedure for disposal of expired chemicals, reagents and solvents
sop for behavior in quality control department
sop for preparation and standardization 0.1M sodium thiosulphate
sop for preparation and standardization 0.1M Disodium Edetate
preparation and standardization 0.1M Sodium Hydroxide Solution
Preparation and standardization of 0.1M Perchloric acid solution
sop for preparation 0.05M edetate disodium
sop for preparation 0.1M silver nitrate
sop for Operation and Calibration of High Performance Liquid Chromatography
sop for UV & Visible Spectrophotometer
procedure for Cleaning of laboratory glassware
Cleaning of Instrument, Instrument bench and surrounding area of Quality
Safety Precaution in Quality Control Department
Operation & Calibration of Analytical Balance
Calibration of Glassware in Quality Control Department
handling of samples received in Quality Control
Cleaning and Operation of Refrigerator
Operation, Cleaning and Calibration of water bath
Operation & Calibration of Refractometer
Operation and Cleaning of Centrifuge Apparatus
cleaning, operation & calibration of Vernier caliper
Calibration of Fourier Transform Infrared Spectrophotometer (FTIR)
Cleaning and operation of Moisture Analyzer
Cleaning & Operation of Vacuum pump in Quality Control Department
Operation and Calibration of Polarimeter
Cleaning and operation of Magnetic Stirrer
Cleaning Operation and Calibration of Melting Point
Operation Cleaning and Calibration of Muffle Furnance in Quality Control Department
procedure of operation and Cleaning of Sonicator
Operation Cleaning & Calibration of pH meter in Quality Control Department