Site icon Pharma Dekho

standard operating procedure leak testing

 

standard operating procedure leak testing

 

1. PURPOSE & SCOPE
To lay down the procedure to check the leak test in blister and strip

2. RESPONSIBILITY
Production / QC supervisor

3. AUTHORITY
QA manager

4. PROCEDURE
1) INTRODUCTION
An important process control in the packaging of tablets and capsule in strips of laminated metal foil and other water impermeable

materials is the test for integrity of sealing of strips. Defects in strips which may due to faulty sealing or due to cracks and

pinholes in the laminate may be detected by a test which is based on the permeation of water through the leaks,

cracks and pinholes when the strips are immersed in water in a container under reduced pressure. This document gives the details of the test.
2) EQUIPMENT
The testing assembly consists of the following :
2.1 A vacuum desiccator half filled with water
2.2 Vacuum Pump fitted with a gauge



2.3 Stiff rubber tubing connecting the desiccator to the pump
2.4 A suitable non-corroding weighting device such as a stainless steel roll
2.5 A suitable housing for the entire assembly such as a wooden box to contain the glass in the event of an implosion of the desiccator under vacuum.
3) FREQUENCY OF TESTING
The frequency of testing is decided by the nature of the product being tested for moisture sensitive products the tests should

be performed a half hourly intervals during the packing run.

4) SAMPLING
The number of strips to be taken for each test will depend on the number of pockets in each track of die roles of the strip packing machine.

It should be ensured that atleast one tablet from each pocket of each track is included in every test. Thus, if we have a

strip of 10 tablets in two rows of five and the machine has 6 tracks with 15 pockets in each track of the die roll a total of 90 tablets in 9 strips should be sampled.

5) TESTING
5.1 The required number of strips should be collected from the machine and tied to the weighing device by means of rubber band.
5.2 The desiccator should be half filled with water. The bundle of strips should be placed inside the desiccator and it should be ensured that it is submerged in water.
5.3 The desiccator should be covered with the lid and the containing box should also be closed.
5.4 The rubber tubing should be attached to the lid of the desiccator and then connected to the vacuum pump. The pump should be switched on and the air inside the desiccator should be evaluated till a minimum vacuum of 381-mm (15”) of mercury is obtained. The stop cork of the desiccator should then be closed and the rubber tubing disconnected.
5.5 The vacuum should be maintained for atleast 30 seconds. Air should be gently passed into the desiccator till complete equilibrium is obtained.
5.6 The cover of the box should be opened and the desiccator lid gently slid. The bundle of strips should be removed and adhering water wiped dry with a clean dry duster.
5.7 The strips should be cut open and the contents examined carefully. The number of capsules or tablets, which have become sticky or moist should
be counted and calculated as a percentage of the total number tested.

 

SOP for testing of ETP water

sop for Calibration and Maintenance of Laboratory Instruments and Equipment

Disposal of Residual Sample or Left Over Material

sop for for Laboratory Incident

standard operating procedure temperature monitoring

sop for drying of silica gel

sop for operation of infrared moisture balance

sop for preparation of mobile phase

sop for treated RO water

sop for Preparation and Issuance of Analysis protocol standard

sop of placebo and impurity stock solutions

sop for disposal of residual sample  

sop for handling of pharmacopoeial changes

sop for procedure for operation of ultrasonic cleaner

difference between UPLC and HPLC

sop for for Emergency Eyewash and Shower

sop for operation and calibration of total organic carbon analyzers

sop for operation of cobb tester

sop for Operation and calibration of atomic absorption spectrophotometer

sop for Operation and calibration of gas liquid chromatograph

sop for operation of humidity oven

sop for operation and calibration of serological water bath

sop for monitoring of drain trap

sop for destruction of analytical samples after testing and control samples

sop for destruction of used chemicals

Sop for Operation of suction pump

sop for Operation and calibration uv cabinet

sop for Operation and calibration of bulk density apparatus

sop for operation and calibration of shore hardness tester

sop for operation of rub proofness tester

sop for monitoring of purified water

sop for Retesting of packaging materials

sop for Retesting and resampling of raw materials

sop for Control of issuance of record of analysis green sheets

sop for Control of computer passwords

sop for sampling of packaging materials PM

sop for sampling of sterile raw material

sop for sampling of intermediates and finished products

sop for operation and calibration of friability test apparatus

sop for approval and rejection of packaging materials

sop for non conformance of RM PM and finished product

sop for collection storage and disposal of control samples

sop for trend analysis of finished products

sop for Chromatographic practices and system suitability

SOP For Good Laboratory Practices

sop for cleaning and operation of sieve shaker

general specification of packing material cartons

sop for Password for Analytical Instrument and LIMS software

sop for Rounding off numerical analytical results

sop for sampling of bulk and finished product

sop for cleaning of spillage material

sop for Handling of Reference Standard

sop for hplc column maintenance and washing

procedure for sampling and handling of bulk sample

STP for borewell generation point (raw water storage tank)

sop for preparation and standardization of 0.1M Zinc Sulphate

Operation &calibration of analytical balance (dhona)

Operation and Calibration Procedure for Disintegration Test apparatus

sop for preparation and standardization of 1 M Hydrochloric Acid

Preparation and standardization of 0.1 M ceric ammonium sulphate solution

sop for preparation and standardization of 0.05 m iodine solution

validation of volumetric solution 0.1m ammonium thiocyanate

handling of reference standard and preparation of working standard

sop for water sampling and analysis

sop for operation for validation of excel worksheets

sop for laboratory cleaning

sop for stability of volumetric solutions

sop for preparation of raw material in process finish product packing material data sheets

sop for handling of hazardous chemicals

sop for handling of glassware and allocation of identification number

sop for operation cleaning and calibration of bursting strength tester

sop for rounding off the analytical test results

procedure for Analyst Qualification

sop for operation and calibration of dissolution Apparatus

procedure for maintenance of desiccators

sop for for hplc column receipt checking id no and regeneration

safety data sheet for laboratory chemicals

procedure for handling of poisonous chemicals

sop for cleaning of sampling devices

sop for calibration procedure of instruments

sop for specification and standard testing raw material packing material and finished product

procedure for operation and calibration of potentiometric titrator

procedure for operating and calibration of digital hardness tester

procedure for disposal of expired chemicals, reagents and solvents

sop for behavior in quality control department

sop for preparation and standardization 0.1M sodium thiosulphate

sop for preparation and standardization 0.1M Disodium Edetate

preparation and standardization 0.1M Sodium Hydroxide Solution

Preparation and standardization of 0.1M Perchloric acid solution

sop for preparation 0.05M edetate disodium

sop for preparation 0.1M silver nitrate

sop for Operation and Calibration of High Performance Liquid Chromatography

sop for UV & Visible Spectrophotometer

procedure for Cleaning of laboratory glassware

Cleaning of Instrument, Instrument bench and surrounding area of Quality

Safety Precaution in Quality Control Department

Operation & Calibration of Analytical Balance

Calibration of Glassware in Quality Control Department

handling of samples received in Quality Control

Cleaning and Operation of Refrigerator

Operation, Cleaning and Calibration of water bath

Operation & Calibration of Refractometer

Operation and Cleaning of Centrifuge Apparatus

cleaning, operation & calibration of Vernier caliper

Calibration of Fourier Transform Infrared Spectrophotometer (FTIR)

Cleaning and operation of Moisture Analyzer

Cleaning & Operation of Vacuum pump in Quality Control Department

Operation and Calibration of Polarimeter

Cleaning and operation of Magnetic Stirrer

Cleaning Operation and Calibration of Melting Point

Operation Cleaning and Calibration of Muffle Furnance in Quality Control Department

procedure of operation and Cleaning of Sonicator

Operation Cleaning & Calibration of pH meter in Quality Control Department

Entry and Exit in Quality Control Department

Exit mobile version