handling of reference standard and preparation of working standard
1.0 OBJECTIVE
1.1 To describe a procedure for handling of reference standard and Preparation of Working standard.
2.0 SCOPE
2.1 This procedure applies for handling of reference standard and Preparation of Working standard to the Quality Control Department
3.0 RESPONSIBILITY
3.1 Officer /Executive – QC
4.0 ACCOUNTABILITY
4.1 Head – QC
5.0 PROCEDURE
5.1 Procedure reference standards from sources as mentioned in Format for preparation and standardization of the WS and enter
the details in register as mentioned in Format
5.2 Store the reference standards in their original Stoppard containers away from heat and protected from light. If any special
storage conditions are specified for a particular reference standard, the same shall be followed.
5.3 Select currently approved raw material and collect the required amount of sample for qualified, working standard qualify against reference standard.
5.4 Procure reference standard from as per own suitable resource. Record the same in Format Prepare the working standard using
reference standard Record the same in Format
5.5 After the preparation of the WS, allotment a unique number as below. E.g. AA/XX/YY/ZZZ.
Where AA Stands for Department, XX Stands Working Standard, YY Stands for year and ZZZ stands for serial number.
For Example QC/WS/18/001.
5.6 The WS shall never be used after the validity. If any change in description is observed during storage the same shall be discarded.
5.7 A thorough investigation shall be made to find out the reason for such change and documented.
5.8 Refrigerated stored vials are to be brought to room temperature before use.
5.9 For Pharmacopoeia API’s :
5.9.1 The material to be used for the standardization of working standard must comply with Pharmacopoeia specifications.
5.9.2 The following tests shall be performed for standardizing working standards.
5.9.3 Description: Note down description of the material and physical appearance.
5.9.4 Identification: Wherever applicable, compare the IR spectrum of the sample under test with that of the Pharmacopoeia
reference standard. If IR is not applicable, test such as melting point or UV absorption maxima or Retention time in chromatogram
of test sample can be used for comparison with the reference standard.
5.9.5 Note: Use relevant Pharmacopoeia standard (IPRS / USPRS / BPCRS / EPCRS) of current lot.
5.9.6 Loss On Drying / Water Content by KF: Determine loss on drying by hot air/vacum oven or water content by
KF as mentioned in the individual standard testing procedure in Triplicate.
5.9.7 Acceptance criteria for Loss On Drying or water content: Each individual test result shall comply with the specification limit.
5.9.8 The overall RSD of six test result shall comply to the following requirements
Specification Limit | % RSD |
Below 1.0 % | 50.0 |
Between 1.1 % and 5.0 % | 25.0 |
Above 5.0 % | 10.0 |
5.9.9 Related Substances: Perform Related Substances as mentioned in the individual standard testing procedure
Pharmacopeia / in – house use relevant Pharmacopoeia reference standard (IPRS / USPRS / BPCRS / EPCRS) of current lot for standard preparation.
Acceptance criteria for Related Substances: Each Related Substances shall be complying within limit as per respective specification.
5.9.10 Assay: Perform assay in Triplicate as mentioned in the individual standard testing procedure.
In case of assay by HPLC / UV, use relevant Pharmacopoeia reference standard (IPRS / USPRS / BPCRS / EPCRS) of current lot
for standard preparation. For standardization HPLC method shall be given preference.
5.9.11 Acceptance criteria for assay: Calculate assay on dried basis or anhydrous basis.
Each individual assay result shall comply with the specification limit. The overall RSD of six assay values shall comply to the following requirements.
B (%) | % RSD |
£ 2.0 | NMT 1.0 |
From 2.1 to 2.5 | NMT 1.25 |
From 2.6 to 3.0 | NMT 1.5 |
B = Upper limit given in the definition of the individual monograph minus 100%, assuming that the upper limit
is set according to the reproducibility of the method.
5.9.12 Purity of working standards: Calculate % purity (i.e., assay on as is basis) of the working standard using the following formula:
% Purity = A (100-B)/100
Where,
A = Average assay of four test results calculated on dried basis or anhydrous basis.
B= Average LOD or water content by KF of four determinations.
5.9.13 Wherever the calculated % purity (i.e. Assay on as is basis) of the working standard exceeds 100%, the same shall be reported as 100% only.
5.9.14 Record the analysis details in data sheet.
5.10 Standardization of new lot: Standardization of new lot of working standards should be always against Pharmacopoeia reference standards of current lot.
5.11 For Non- Pharmacopoeia API:
5.11.1 For non-Pharmacopoeia materials (API) working / primary standards shall be procured along with certificate of analysis
from API manufacturer and used as test standard to qualify the WS against them.
5.11.2 Alternatively working standard of required quantity shall also be procured from the API manufacturer for routine use.
5.11.3 Standardization of new lot: New lot of working standard shall be standardized against the working / primary standard
procured from the manufacturer or the required quantity of working standard shall be taken from the manufacturer for routine use.
5.12 Selection of API’s for Working Standards:
5.12.1 i) API’s of Raw material shall be use for working standard qualification of current/latest manufacturing date.
ii) API’s of Raw material shall be use for working standard qualification of highly purity.
iii) API’s of Raw material shall be use for working standard qualification of approved vendor.
5.13 Assignment of expiry period for working standards:
5.13.1 i) Highly stable API’s shall be given a validity period of two year (Non-hygroscopic and having an expiry period of three years or above).
ii) Antibiotics or Hygroscopic API’s shall be given a validity period of one year (having an expiry period of two to four years).
iii) Vitamins or highly unstable API’s shall be given a validity period of six month ( having an expiry period of two to four years)
5.13.2 Before assigning expiry period for working standard QC personnel shall fill the assessment form (QCA/016/F02/R-00) along with
available supporting data and submit the same to QC In-charge.
5.13.3 The QC In-charge shall review the data and shall approve the proposed expiry period.
5.14 Storage of working standards: Working standards shall be stored in amber glass vial, place these vials in desiccator containing
self indicating Silica gel and keep the desiccator below 25ºC.
5.14.1 Working standard Certificate: Prepare working standard certificate as mentioned in format
5.15.1 Usage of working standards:
5.15.2 Distribute as per required quantity of working standard from the standardized quantity into 12/12/06 respectively in amber
glass vials and one/one vial for breakage and stock. Label affix of the working standard vials as per format
5.15.3 Note: The working standard shall be labeled for drying or moisture determination in accordance to the Pharmacopeia
standard or as recommended by the API manufacturer, wherever applicable based on the lab data generated.
5.15.4 The working standard vials should be used within 2 Month for two year validity and within 1 Month for one
year / six month validity respectively from the date of opening.
5.15.5 Note: If the API is highly hygroscopic single time usage is preferred. In such cases the vials shall be
filled with about 100 mg of working standard and label such vials as “for single use only”.
5.15.6 Enter the inventory details of each working standard uses in relevant format
5.15.7 Handling of working standard :
5.15.8 At the time of use, analyst is to ensure that the standard has not crossed the validity date. The analyst will keep
the standard out of the storage for a shorter period possible to prevent light and moisture from affecting the standard. When the
standards are stored in cool place, equilibrate to room temperature before use.
5.15.9 Do not transfer back into the vial, any excess portion of working standard used during weighing.
5.15.10 After the usage by the analyst the standards it is to be returned to the appropriate storage location immediately.
5.15.11 *NOTE: Working standards received with certificate of Analysis from Raw Material Manufacturer or from any other location can be used.
5.15.12 When self-indicating silica gel colour turns blue to red, dry the silica at 105°C till it regains its original colour. If it does not
regain its original colour change the silica.
5.15.13 All WS should be listed as mentioned in Format
5.15.14 In case we shall be qualify working standard from PTL (NABL Approved Lab), Perform as per site approved
respective sop’s and related document should be comply.
5.15.15 Destruction of the working standards
5.15.16 Deface the vial label.
5.15.17 Transfer the content of the vial in a container.
5.15.18 Rinse the vial with water and send it for disposal.
6.0 RELATED DOCUMENT
6.1 Nil
7.0 REFERENCE
7.1 In House
8.0 FORMATS
8.1 List of Reference Standard
8.2 Working Standard Assessment Form
8.3 Working Standard Certificate / Report
8.4 Working Standard Analytical Data Sheet
8.5 Label for Working Standard
8.6 List of Working Standard
8.7 Working Standard Consumption Record
8.8 Reference Standard Receipt, Issuance and Consumption Record
8.9 Working Standard Destruction Record
9.0 ABBREVIATIONS
9.1 No. : Number
9.2 QA : Quality Assurance
9.3 QC : Quality Control
9.4 QCA : Quality Control Analytical
9.5 SOP : Standard Operating Procedure
9.6 RSD : Relative Standard Deviation
9.7 LOD : Loss on drying
9.8 Ref. No : Reference Number
9.9 % : Percentage
9.10 °C : Degree Celsius
9.11 A. C. : Air conditioned
9.12 COA : Certificate of Analysis
9.13 WS : Working Standard
9.14 Dept. : Department
9.15 PTL : Public Testing Laboratory
sop for Calibration and Maintenance of Laboratory Instruments and Equipment
Disposal of Residual Sample or Left Over Material
sop for for Laboratory Incident
standard operating procedure temperature monitoring
sop for operation of infrared moisture balance
sop for preparation of mobile phase
sop for Preparation and Issuance of Analysis protocol standard
sop of placebo and impurity stock solutions
sop for disposal of residual sample
sop for handling of pharmacopoeial changes
sop for procedure for operation of ultrasonic cleaner
difference between UPLC and HPLC
sop for for Emergency Eyewash and Shower
sop for operation and calibration of total organic carbon analyzers
sop for operation of cobb tester
sop for Operation and calibration of atomic absorption spectrophotometer
sop for Operation and calibration of gas liquid chromatograph
sop for operation of humidity oven
sop for operation and calibration of serological water bath
sop for monitoring of drain trap
sop for destruction of analytical samples after testing and control samples
sop for destruction of used chemicals
Sop for Operation of suction pump
sop for Operation and calibration uv cabinet
sop for Operation and calibration of bulk density apparatus
sop for operation and calibration of shore hardness tester
sop for operation of rub proofness tester
sop for monitoring of purified water
sop for Retesting of packaging materials
sop for Retesting and resampling of raw materials
sop for Control of issuance of record of analysis green sheets
sop for Control of computer passwords
sop for sampling of packaging materials PM
sop for sampling of sterile raw material
sop for sampling of intermediates and finished products
sop for operation and calibration of friability test apparatus
sop for approval and rejection of packaging materials
sop for non conformance of RM PM and finished product
sop for collection storage and disposal of control samples
sop for trend analysis of finished products
sop for Chromatographic practices and system suitability
SOP For Good Laboratory Practices
sop for cleaning and operation of sieve shaker
general specification of packing material cartons
sop for Password for Analytical Instrument and LIMS software
sop for Rounding off numerical analytical results
sop for sampling of bulk and finished product
sop for cleaning of spillage material
sop for Handling of Reference Standard
sop for hplc column maintenance and washing
procedure for sampling and handling of bulk sample
STP for borewell generation point (raw water storage tank)
sop for preparation and standardization of 0.1M Zinc Sulphate
Operation &calibration of analytical balance (dhona)
Operation and Calibration Procedure for Disintegration Test apparatus
sop for preparation and standardization of 1 M Hydrochloric Acid
Preparation and standardization of 0.1 M ceric ammonium sulphate solution
sop for preparation and standardization of 0.05 m iodine solution
validation of volumetric solution 0.1m ammonium thiocyanate
handling of reference standard and preparation of working standard
sop for water sampling and analysis
sop for operation for validation of excel worksheets
sop for stability of volumetric solutions
sop for preparation of raw material in process finish product packing material data sheets
sop for handling of hazardous chemicals
sop for handling of glassware and allocation of identification number
sop for operation cleaning and calibration of bursting strength tester
sop for rounding off the analytical test results
procedure for Analyst Qualification
sop for operation and calibration of dissolution Apparatus
procedure for maintenance of desiccators
sop for for hplc column receipt checking id no and regeneration
safety data sheet for laboratory chemicals
procedure for handling of poisonous chemicals
sop for cleaning of sampling devices
sop for calibration procedure of instruments
sop for specification and standard testing raw material packing material and finished product
procedure for operation and calibration of potentiometric titrator
procedure for operating and calibration of digital hardness tester
procedure for disposal of expired chemicals, reagents and solvents
sop for behavior in quality control department
sop for preparation and standardization 0.1M sodium thiosulphate
sop for preparation and standardization 0.1M Disodium Edetate
preparation and standardization 0.1M Sodium Hydroxide Solution
Preparation and standardization of 0.1M Perchloric acid solution
sop for preparation 0.05M edetate disodium
sop for preparation 0.1M silver nitrate
sop for Operation and Calibration of High Performance Liquid Chromatography
sop for UV & Visible Spectrophotometer
procedure for Cleaning of laboratory glassware
Cleaning of Instrument, Instrument bench and surrounding area of Quality
Safety Precaution in Quality Control Department
Operation & Calibration of Analytical Balance
Calibration of Glassware in Quality Control Department
handling of samples received in Quality Control
Cleaning and Operation of Refrigerator
Operation, Cleaning and Calibration of water bath
Operation & Calibration of Refractometer
Operation and Cleaning of Centrifuge Apparatus
cleaning, operation & calibration of Vernier caliper
Calibration of Fourier Transform Infrared Spectrophotometer (FTIR)
Cleaning and operation of Moisture Analyzer
Cleaning & Operation of Vacuum pump in Quality Control Department
Operation and Calibration of Polarimeter
Cleaning and operation of Magnetic Stirrer
Cleaning Operation and Calibration of Melting Point
Operation Cleaning and Calibration of Muffle Furnance in Quality Control Department
procedure of operation and Cleaning of Sonicator
Operation Cleaning & Calibration of pH meter in Quality Control Department
Entry and Exit in Quality Control Department