Operation, Cleaning and Calibration of water bath

 

 

Operation, Cleaning and Calibration of water bath

 

 

1.0 OBJECTIVE

1.1 To lay down a procedure for Operation, Cleaning and Calibration of water bath.

2.0 SCOPE

2.1 This procedure is applicable for Cleaning ,Operation and Calibration of water bath in Quality Control Department

3.0 RESPONSIBILITY
3.1 QC Officer or QC Executive

4.0 ACCOUNTABILITY
4.1 QC Manager

5.0 PROCEDURE (for operation and cleaning)


5.1 Check general cleanliness of the water bath and clean outer surface by lint free cloth .

5.2 Fill the purified water up to two third part of the water bath to properly dip the coil of heater.

5.3 Switch “ON” the main electric supply followed by switch “ON” the instrument, red indicator will glow on.

5.4 Adjust the temperature of the bath with temperature controller knob as per requirement.

5.5 When the required temperature of the bath reaches, the green light will be “OFF”

5.6 When a fix temperature is required for analysis then check the temperature with the help of calibrated
thermometer for verification.

5.7 Place the flask or beaker-containing sample on the bath as per the size & requirement for complete digestion .

5.8 When the operation completed switch “OFF” the instrument followed by switch “OFF” the main electric supply.

5.9 Remove the sample container from water bath carefully .

5.10 Maintained the record.


5.11 CALIBRATION

5.11.1 Operate the water bath according to procedure 5.1 to 5.6

5.11.2 Set temperature at 35°C and wait till temperature reaches, Take suitable calibrated thermometer and measure the temperature after, 1 hour , 2 hours and 3 hours.

5.11.3 Repeat the same procedure 5.11.2 for temperature 90°C.

5.11.4 Note down the observed temperature in Annex I.

5.12 Frequency

5.12.1 Three months.

5.13 Acceptance Criteria

5.13.1 For temperature 35°C = ± 2°C.

5.13.2 For temperature 90°C = ± 2°C.

5.14 Precautions

5.14.1 Ensure that there is sufficient water in water bath.

5.14.2 Instrument should be clean & free from dust.


6.0 ABBREVIATION

Abbreviation used

Extended form

SOP

Standard Operating Procedure

No.

Number

°C

Degree Centigrade

QC

Quality Control

QA

Quality Assurance

Annex

Annexure




Annexure-1

Calibration record of water Bath


Instrument ID: Date:

CALIBRATION DATE

Thermometer
ID. No. of Thermometer :
Thermometer Calibrated On :
Thermometer Calibration Due :

CALIBRATION DATA

Abbreviation used

Extended form

SOP

Standard Operating Procedure

No.

Number

°C

Degree Centigrade

QC

Quality Control

QA

Quality Assurance

Annex

Annexure

Calibration Status: The Instrument is calibrated as per Schedule/After Maintenance/Other (Specify) _________________ and

complies/does not comply with the predefined acceptance criteria and is suitable/not suitable for the routine usage.

Frequency: Quarterly (± 7 Days of Schedule date)

SOP for testing of ETP water

sop for Calibration and Maintenance of Laboratory Instruments and Equipment

Disposal of Residual Sample or Left Over Material

sop for for Laboratory Incident

standard operating procedure temperature monitoring

sop for drying of silica gel

sop for operation of infrared moisture balance

sop for preparation of mobile phase

sop for treated RO water

sop for Preparation and Issuance of Analysis protocol standard

sop of placebo and impurity stock solutions

sop for disposal of residual sample  

sop for handling of pharmacopoeial changes

sop for procedure for operation of ultrasonic cleaner

difference between UPLC and HPLC

sop for for Emergency Eyewash and Shower

sop for operation and calibration of total organic carbon analyzers

sop for operation of cobb tester

sop for Operation and calibration of atomic absorption spectrophotometer

sop for Operation and calibration of gas liquid chromatograph

sop for operation of humidity oven

sop for operation and calibration of serological water bath

sop for monitoring of drain trap

sop for destruction of analytical samples after testing and control samples

sop for destruction of used chemicals

Sop for Operation of suction pump

sop for Operation and calibration uv cabinet

sop for Operation and calibration of bulk density apparatus

sop for operation and calibration of shore hardness tester

sop for operation of rub proofness tester

sop for monitoring of purified water

sop for Retesting of packaging materials

sop for Retesting and resampling of raw materials

sop for Control of issuance of record of analysis green sheets

sop for Control of computer passwords

sop for sampling of packaging materials PM

sop for sampling of sterile raw material

sop for sampling of intermediates and finished products

sop for operation and calibration of friability test apparatus

sop for approval and rejection of packaging materials

sop for non conformance of RM PM and finished product

sop for collection storage and disposal of control samples

sop for trend analysis of finished products

sop for Chromatographic practices and system suitability

SOP For Good Laboratory Practices

sop for cleaning and operation of sieve shaker

general specification of packing material cartons

sop for Password for Analytical Instrument and LIMS software

sop for Rounding off numerical analytical results

sop for sampling of bulk and finished product

sop for cleaning of spillage material

sop for Handling of Reference Standard

sop for hplc column maintenance and washing

procedure for sampling and handling of bulk sample

STP for borewell generation point (raw water storage tank)

sop for preparation and standardization of 0.1M Zinc Sulphate

Operation &calibration of analytical balance (dhona)

Operation and Calibration Procedure for Disintegration Test apparatus

sop for preparation and standardization of 1 M Hydrochloric Acid

Preparation and standardization of 0.1 M ceric ammonium sulphate solution

sop for preparation and standardization of 0.05 m iodine solution

validation of volumetric solution 0.1m ammonium thiocyanate

handling of reference standard and preparation of working standard

sop for water sampling and analysis

sop for operation for validation of excel worksheets

sop for laboratory cleaning

sop for stability of volumetric solutions

sop for preparation of raw material in process finish product packing material data sheets

sop for handling of hazardous chemicals

sop for handling of glassware and allocation of identification number

sop for operation cleaning and calibration of bursting strength tester

sop for rounding off the analytical test results

procedure for Analyst Qualification

sop for operation and calibration of dissolution Apparatus

procedure for maintenance of desiccators

sop for for hplc column receipt checking id no and regeneration

safety data sheet for laboratory chemicals

procedure for handling of poisonous chemicals

sop for cleaning of sampling devices

sop for calibration procedure of instruments

sop for specification and standard testing raw material packing material and finished product

procedure for operation and calibration of potentiometric titrator

procedure for operating and calibration of digital hardness tester

procedure for disposal of expired chemicals, reagents and solvents

sop for behavior in quality control department

sop for preparation and standardization 0.1M sodium thiosulphate

sop for preparation and standardization 0.1M Disodium Edetate

preparation and standardization 0.1M Sodium Hydroxide Solution

Preparation and standardization of 0.1M Perchloric acid solution

sop for preparation 0.05M edetate disodium

sop for preparation 0.1M silver nitrate

sop for Operation and Calibration of High Performance Liquid Chromatography

sop for UV & Visible Spectrophotometer

procedure for Cleaning of laboratory glassware

Cleaning of Instrument, Instrument bench and surrounding area of Quality

Safety Precaution in Quality Control Department

Operation & Calibration of Analytical Balance

Calibration of Glassware in Quality Control Department

handling of samples received in Quality Control

Cleaning and Operation of Refrigerator

Operation, Cleaning and Calibration of water bath

Operation & Calibration of Refractometer

Operation and Cleaning of Centrifuge Apparatus

cleaning, operation & calibration of Vernier caliper

Calibration of Fourier Transform Infrared Spectrophotometer (FTIR)

Cleaning and operation of Moisture Analyzer

Cleaning & Operation of Vacuum pump in Quality Control Department

Operation and Calibration of Polarimeter

Cleaning and operation of Magnetic Stirrer

Cleaning Operation and Calibration of Melting Point

Operation Cleaning and Calibration of Muffle Furnance in Quality Control Department

procedure of operation and Cleaning of Sonicator

Operation Cleaning & Calibration of pH meter in Quality Control Department

Entry and Exit in Quality Control Department

Leave a Comment

error: Content is protected !!