Procedure for Glass Ware Cleaning

 

Procedure for Glass Ware Cleaning

 

1.0 Objective
To lay down a procedure for Glass ware cleaning.
2.0 Scope
This SOP applies to procedure for Glass ware cleaning is applicable at Quality Control Department in ABC Company.
3.0 Responsibility
3.1 Concerned Officer/Executive shall be responsible to follow this SOP.
3.2 Department Head or his/her designee shall be responsible to implement and ensuring to compliance of this SOP.
4.0 Abbreviations and Definitions
SOP : Standard Operating Procedure
QC : Quality Control
QA : Quality Assurance
5.0 Procedure
5.1 Cleaning agents
5.1.1 Preparation of 2% Teepol: Measure and transfer 10 ml of Teepol liquid in to 500 ml volumetric flask,

make up to the mark with purified water with intermediate shakings. Then transfer into a suitable volume can.
5.1.2 10% Nitric acid: Measure and transfer 10ml of Nitric acid in to 100ml volumetric flask, make up to

the mark with purified water with intermediate shakings. Then transfer into a suitable volume can.
5.1.3 Chromic acid solution: Dissolve 200g of sodium dichromate in 100 ml of purified water in a two liter

hard borosilicate glass beaker. Slowly and cautiously add sulfuric acid 1500ml by stirring. Cool to room

temperature and transfer to a glass Stoppard bottle.
5.2 Glass ware Cleaning for routine
5.2.1 Keep the used or contaminated glass apparatus in a plastic tray and send them to washing area.
5.2.2 Rinse each glass ware with potable water, and transfer them to sufficient capacity plastic area.
5.2.3 Measure and transfer potable water in to plastic tray that is sufficient able to dip all glass ware.
5.2.4 Measure and transfer Teepol in to the tray to produce 2 % in potable water.
5.2.5 Note: Prepare required quantity of cleaning solution in the same ratio given above.
5.2.6 Dip the glass apparatus in cleaning solution.
5.2.7 Soak all glass apparatus for about 1 hour in cleaning solution and take them out. Put back cleaning

solution retained in the glassware.
5.2.8 Volumetric flasks, conical flasks, measuring cylinders and test tubes: Take potable water upto half

of the capacity, shake more than 5 times and discard.
5.2.9 Pipettes & burettes: Take potable water upto half of the capacity and rinse them by inverting and

shaking more than 5 times. Rinse with potable water for 4 more times.
5.2.10 Other glass apparatus shall be cleaned with nylon brush. Rinse for 5 times with potable water by

taking potable water upto half of the capacity each time.
5.2.11 Finally rinse all the glass apparatus twice with small amount of purified water.
5.2.12 After final rinse with purified water, put them in a clean and dry plastic/SS tray.
5.2.13 Dry the glassware in Hot Air Oven at 60oC for 1 hour. After the Drying period is over allow the glassware

to cool and use it.
5.3 Glass ware Cleaning for un-removable residues.
5.3.1 Continue the point number 5.3.8 for the glass ware which ever having un-removable residues inside.
5.3.2 Fill the glass apparatus with chromic acid solution, and keep it idle for 20 minutes, if required keep

it in sonicator till residues are dissolved.
5.3.3 Further follow the point number 5.3.10 to 5.3.14.
5.4 Precautions:
5.4.1 Wear safety goggles and gloves at the time of preparation and use of chromic acid cleaning mixture.
5.4.2 Chromic acid cleaning mixture is extremely corrosive and hygroscopic and should be stored in glass stopper

bottles in a safe place.
5.4.3 When the mixture acquires a green color, discard the mixture by diluting several times with water.
6.0 Forms and Records
6.1 None
7.0 Distribution
7.1 Master copy – Documentation Cell (Quality Assurance)
7.2 Control copy – Quality Control
8.0 History

Revision Number Details For Change
Reason for Revision
00 New SOP NA

 

Standard testing procedure of Iron Sucrose Injection

Standard testing procedure lactose

Standard testing procedure mefenamic acid

standard testing procedure domperidone

standard testing procedure flavour mixed fruit

standard testing procedure dicyclomine hydrochloride

Standard testing procedure honey pure

standard testing procedure dextromethorphan hydrobromide

standard procedure of levocarnitine injection

Analysis of Ivermectin Suspension

standard testing procedure artemether injection

standard testing procedure artemether injection

standard testing procedure Carbocisteine syrup

standard testing procedure Phytomenadione injection

standard testing procedure serratiopeptidase



standard testing procedure starch IP

standard testing procedure sucrose refined sugar

standard testing procedure titanium dioxide

standard testing procedure tramadol hydrochloride

standard testing procedure zinc sulphate

standard testing procedure croscarmellose sodium

standard testing procedure colour erythrosine supra

standard testing procedure magnesium hydroxide

standard testing procedure diclofenac sodium

standard testing procedure dibasic calcium phosphate

standard testing procedure cyanocobalamin

standard testing procedure cholecalciferol

standard testing procedure Calcium carbonate oyster shell powder

standard test procedure Calcium Citrate

standard testing procedure Bronopol

standard testing procedure Bromhexine Hydrochloride

Standard Testing Procedure diclofenac sodium injection

Standard Testing Procedure Drotaverine Hydrochloride injection

analysis of Procyclidine HCl 5 mg Tablet

analysis of silodosin 8 mg capsule

Method of Analysis Ceftriaxone and Sulbactam For Injection

Analysis of Glimepiride and Metformin Hydrochloride (SR) Tablets

General Test Procedure Melting Range and Freezing Point

General Test Procedure Weight per ML and Specific Gravity

General Test Procedure Distillation Range

General Test Procedure Non Volatile Matter and Residue on Evaporation

General Test Procedure Clarity of Solution

General Test Procedure Limit Test Chloride

General Test Procedure Limit Test Iron

General Test Procedure Limit Test Heavy Metals

General Test Procedure Water Soluble Substances

General Test Procedure Acid Soluble Substances

General Test Procedure 0.1M Ammonium Thiocyanate

General Test Procedure 0.1M Ethanolic Sodium Hydroxide

Calibration Of Disintegration Test Apparatus

Procedure for Glass Ware Cleaning

 

Leave a Comment

error: Content is protected !!