sop for Calibration and Maintenance of Laboratory Instruments and Equipment
1.0 OBJECTIVE:
To lay down a procedure for the Usage, Calibration and Maintenance of Laboratory Instruments and Equipments in Quality Control Laboratory
2.0 SCOPE:
This procedure is applicable for the Usage, Calibration and Maintenance of Laboratory Instruments and
Equipments which are installed and being used in the Quality Control Department.
3.0 RESPONSIBILITY:
Officer / Executive- QC
4.0 ACCOUNTABILITY:
Head of the QC Department.
To ensure the Compliance of this SOP, in all aspects
5.0 PROCEDURE:
5.1 Before using any instrument, the analyst shall ensure that the respective instrument is placed on a leveled, Sturdy and vibration free platform or table.
5.2 Analyst shall ensure that the respective instrument is neat and clean. In case the cleaning is not proper,
the said instrument shall be cleaned properly by using a soft lint free cloth and or a soft brush.
5.3 Analyst shall check and ensure that all the respective electrical and software connections are fixed properly.
5.4 Prior to usage, the analyst shall ensure the respective instrument is calibrated, and is suitable for the intended usage.
5.5 No instrument should be used immediately after switching “ON” the power supply, adequate time shall be given
(As per the requirement of the respective instrument) to get it stabilized.
5.6 The analyst shall ensure to make the relevant entries in the Usage log book of the respective instrument.
5.7 All the respective entries shall be done as per Annexure-I & Annexure-II (As Applicable).
5.8 After the usage, the analyst shall ensure the proper cleanliness of the respective instrument and record the same as per Annexure-I & Annexure-II (As Applicable).
5.9 For accurate and reliable results all the laboratory instruments should be calibrated as per their pre- defined frequency/schedule, procedure and parameters.
5.10 The instruments, for which calibration facility is not available at Site, shall get calibrated by authorized
approved external agencies, and a record of the same shall be maintained accordingly. The responsibility for the same lies with QA Department.
5.11 Master calibration schedule should be prepared for all the laboratory instruments, and the calibration record shall be maintained accordingly.
5.12 Master Calibration Schedules shall be prepared for the instruments which calibration facility is available within the laboratory/site,
and all instruments which need to be calibrated by an approved / authorized external agency.
5.13 The respective Master schedule(s) shall be prepared as per Annexure-III, on yearly basis.
5.14 The first (Initial) Master Calibration schedule shall have the Version No. as 01, and if due to any updation
(Addition / Deletion) of laboratory instruments list specified in the existing schedule, Change in Calibration frequency
of any Laboratory instrument, or due to any other unavoidable / unforeseen reason, the master calibration schedule needs to
be updated and revised, then the new schedule shall have the Version No. as 02, and next as 03 and so on. Reason for making
any change(s) in the existing master calibration schedule shall be mentioned in the revised schedule, against the Remarks option.
5.15 All the versions (Initial and the revised ones) of the Master Calibration schedule(s) should be maintained, recorded and documented properly.
5.16 After the calibration of the respective instrument is completed successfully, the GLP section in- charge shall update
the master calibration schedule by making the entry in the box mentioned against “Executed On”, and put his / her sign
with date against the option “Checked By/On”. This is applicable for the instruments for which facility for calibration is available at site QC Laboratory.
5.17 For all the instruments which are being calibrated by an approved / authorized external agency, the responsibility for
updating the Master Calibration schedule lies with Manager / Executive / Designee QA Department.
5.18 All the instruments should be calibrated as per their respective calibration frequency and within the
5.19 If due to any unforeseen / unavoidable circumstances, any instrument is not calibrated within the specified period,
a Deviation shall be raised for the same by Head –QC, with proper justification, and the respective instrument can only be used for routine analytical activities after the deviation is accepted and approved by the Head – QA.
5.20 Same shall also be informed to the Head Quality
5.21 Head QC shall ensure that the respective instrument be calibrated at the earliest and get the respective deviation closed.
5.22 In case the respective instrument is malfunctioning, is out of order and / or is not complying with the specified acceptance criterion
of calibration, then immediately inform the Head – QC, site engineering department, QA and the respective service provider – As applicable,
and the user should immediately put a tag having the information as shown below.
5.23 Same shall also be recorded in the usage log book of the respective instrument, and in the “Maintenance/Service History Card”
for the maintenance and service activities of the said instrument, as per Annexure-IV.
5.24 No instrument should be used for routine analytical activities if its calibration fails to comply with the pre-defined acceptance criterion.
5.25 For non-functioning instruments give detail of follow up in progress register for their fault rectification and arrangement of spare
part for the same instruments as per Annexure-V of this SOP.
5.26 After the error / fault is rectified, the respective instrument should be calibrated and shall be used for routine analytical
activities only if it complies with the pre-defined acceptance criterion. impact analysis for the various Materials / Products which were analyzed
using the instrument which failed to comply with the acceptance criterion.
5.28 The impact analysis shall be carried out for the products / materials which were analyzed using the respective instrument,
from the date when the respective instrument fails to comply with the acceptance criterion to the Last successful calibration date of the said instrument.
5.29 For impact analysis, all the materials / products which were analyzed using the said instrument in the last 24 hours should
be reanalyzed (Immediately) as per the respective Specifications and MOA.
5.30 The impact analysis shall be extended to rest of the materials / products which were analyzed using the said instrument
by selecting randomly the samples analyzed as per the criteria described below:
5.31 Any material / product analyzed one week before the date when the instrument fails to qualify the calibration acceptance criterion.
5.32 Any material / product analyzed 15 days before the date when the instrument fails to qualify the calibration acceptance criterion.
5.33 Any material / product analyzed one Month before (And thereafter on monthly basis) the date when the instrument fails to qualify
the calibration acceptance criterion, till the date of the last successful calibration.
5.34 If all the reanalyzed materials / products meet the respective acceptance criterion (Specifications), the results of the initial analysis remain valid.
5.35 If any of the reanalyzed materials / products fails to meets the respective acceptance criterion, proceed as per the SOP on “Handling of Out of Specification” .
5.36 If the findings of OOS investigation complies with the respective acceptance criterion, then the
result(s) of the initial analysis stands valid, and if after the OOS investigation the respective
material / product fails to comply with the predefined acceptance criteria, In such cases Head QC shall immediately inform to Head QA
for further action and appropriate decision.
5.37 Following Tolerance Window(s), shall be followed, As mentioned in Table-I, while calibrating the respective instruments:
5.38 The respective instruments shall be calibrated as per the Frequency mentioned in Respective SOP.
S.No. | CALIBRATION | TOLERANCE |
1.0 | Daily | Same day |
2.0 | Weekly | ± 1 Day |
3.0 | Fortnightly | ± 2 Days |
4.0 | Monthly | ± 3 Days |
5.0 | Quarterly | ± 7 Days |
6.0 | Half Yearly | ± 10 Days |
7.0 | Yearly | ± 15 Days |
5.4 Note: The tolerance window shall be applicable for working days only if any holiday falls in between tolerance window date then it shall extend accordingly.
5.39 Initial (First Time) Calibration date should be denoted with a ‘Standard Date’. Instrument having Calibration Frequency as Quarterly,
and 15/04/20 being its “Standard Date”, shall have calibration schedule as 15/04/20 (Initial), followed by 15/07/20; 15/10/20 and 15/01/20. However
the respective (Followed after initial calibration) calibrations can be performed within the Tolerance Window of ±7 days. Hence all the calibrations,
Followed by the initial calibration, can be performed anywhere between 8th and 22nd of the respective month, When the calibration is due.
5.40 In case the respective instrument is required to be calibrated, in between, before the next due date, For Example: After performing any
maintenance activity, which may be required due to the malfunctioning of the instrument, Change of any part of the said instrument,
or Relocating the respective instrument, etc., then the next due date of the calibration of the respective remains “Unaffected”.
6.0 ABBREVIATIONS:
SOP Standard Operating Procedure
No. Number
QA Quality Assurance
QC Quality Control
MOA Method of Analysis
GLP Good Laboratory Practice
S.No Serial Number.
E and M Engineering & maintenance
sop for Calibration and Maintenance of Laboratory Instruments and Equipment
Disposal of Residual Sample or Left Over Material
sop for for Laboratory Incident
standard operating procedure temperature monitoring
sop for operation of infrared moisture balance
sop for preparation of mobile phase
sop for Preparation and Issuance of Analysis protocol standard
sop of placebo and impurity stock solutions
sop for disposal of residual sample
sop for handling of pharmacopoeial changes
sop for procedure for operation of ultrasonic cleaner
difference between UPLC and HPLC
sop for for Emergency Eyewash and Shower
sop for operation and calibration of total organic carbon analyzers
sop for operation of cobb tester
sop for Operation and calibration of atomic absorption spectrophotometer
sop for Operation and calibration of gas liquid chromatograph
sop for operation of humidity oven
sop for operation and calibration of serological water bath
sop for monitoring of drain trap
sop for destruction of analytical samples after testing and control samples
sop for destruction of used chemicals
Sop for Operation of suction pump
sop for Operation and calibration uv cabinet
sop for Operation and calibration of bulk density apparatus
sop for operation and calibration of shore hardness tester
sop for operation of rub proofness tester
sop for monitoring of purified water
sop for Retesting of packaging materials
sop for Retesting and resampling of raw materials
sop for Control of issuance of record of analysis green sheets
sop for Control of computer passwords
sop for sampling of packaging materials PM
sop for sampling of sterile raw material
sop for sampling of intermediates and finished products
sop for operation and calibration of friability test apparatus
sop for approval and rejection of packaging materials
sop for non conformance of RM PM and finished product
sop for collection storage and disposal of control samples
sop for trend analysis of finished products
sop for Chromatographic practices and system suitability
SOP For Good Laboratory Practices
sop for cleaning and operation of sieve shaker
general specification of packing material cartons
sop for Password for Analytical Instrument and LIMS software
sop for Rounding off numerical analytical results
sop for sampling of bulk and finished product
sop for cleaning of spillage material
sop for Handling of Reference Standard
sop for hplc column maintenance and washing
procedure for sampling and handling of bulk sample
STP for borewell generation point (raw water storage tank)
sop for preparation and standardization of 0.1M Zinc Sulphate
Operation &calibration of analytical balance (dhona)
Operation and Calibration Procedure for Disintegration Test apparatus
sop for preparation and standardization of 1 M Hydrochloric Acid
Preparation and standardization of 0.1 M ceric ammonium sulphate solution
sop for preparation and standardization of 0.05 m iodine solution
validation of volumetric solution 0.1m ammonium thiocyanate
handling of reference standard and preparation of working standard
sop for water sampling and analysis
sop for operation for validation of excel worksheets
sop for stability of volumetric solutions
sop for preparation of raw material in process finish product packing material data sheets
sop for handling of hazardous chemicals
sop for handling of glassware and allocation of identification number
sop for operation cleaning and calibration of bursting strength tester
sop for rounding off the analytical test results
procedure for Analyst Qualification
sop for operation and calibration of dissolution Apparatus
procedure for maintenance of desiccators
sop for for hplc column receipt checking id no and regeneration
safety data sheet for laboratory chemicals
procedure for handling of poisonous chemicals
sop for cleaning of sampling devices
sop for calibration procedure of instruments
sop for specification and standard testing raw material packing material and finished product
procedure for operation and calibration of potentiometric titrator
procedure for operating and calibration of digital hardness tester
procedure for disposal of expired chemicals, reagents and solvents
sop for behavior in quality control department
sop for preparation and standardization 0.1M sodium thiosulphate
sop for preparation and standardization 0.1M Disodium Edetate
preparation and standardization 0.1M Sodium Hydroxide Solution
Preparation and standardization of 0.1M Perchloric acid solution
sop for preparation 0.05M edetate disodium
sop for preparation 0.1M silver nitrate
sop for Operation and Calibration of High Performance Liquid Chromatography
sop for UV & Visible Spectrophotometer
procedure for Cleaning of laboratory glassware
Cleaning of Instrument, Instrument bench and surrounding area of Quality
Safety Precaution in Quality Control Department
Operation & Calibration of Analytical Balance
Calibration of Glassware in Quality Control Department
handling of samples received in Quality Control
Cleaning and Operation of Refrigerator
Operation, Cleaning and Calibration of water bath
Operation & Calibration of Refractometer
Operation and Cleaning of Centrifuge Apparatus
cleaning, operation & calibration of Vernier caliper
Calibration of Fourier Transform Infrared Spectrophotometer (FTIR)
Cleaning and operation of Moisture Analyzer
Cleaning & Operation of Vacuum pump in Quality Control Department
Operation cleaning and calibration of Karl Fischer Apparatus
Operation and Calibration of Polarimeter
Cleaning and operation of Magnetic Stirrer
Cleaning Operation and Calibration of Melting Point
Operation Cleaning and Calibration of Muffle Furnance in Quality Control Department
procedure of operation and Cleaning of Sonicator
Operation Cleaning & Calibration of pH meter in Quality Control Department
Entry and Exit in Quality Control Department