sop for Chromatographic practices and system suitability

 

sop for Chromatographic practices and system suitability

 

1.0. OBJECTIVE:
To lay down a procedure for Chromatographic practices and system suitability.

2.0. SCOPE:

This SOP is applicable for procedure of Chromatographic practices and system suitability in quality control department



3.0. RESPONSIBILITY:
Executive /Officer QC

4.0 ACCOUNTABILITY:
Head- Quality Control

5.0 DEFINITION:
Chromatography is a Laboratory technique for the identification and separation of drugs mixture. Where it is used on a large

scale to purify the intermediates and products in various synthesis in Pharmaceuticals industries etc.

PROCEDURE
Standard Practices:
Chromatographic conditions are set as per the method specified in respective standard testing

Procedure.

Analyst should not change any system parameters without prior information to Head-QC. If any change in system



parameters is performed then it should be acknowledged and properly documented.
Prepare mobile phase, according to STP or Pharmacopoeia and filter with 0.45 µ filter.
Mobile phase bottles and all solvents bottles in use should be labeled.
Mobile phase should be freshly prepared for analysis and solvents should not be used for more than 72 hours.
Check the solvent quantity in solvent bottles before start the instrument.; no any air bubbles should be observed in line. Solvent bottles should be covered.
Check no any blockage in wastage line and outlet of wastage line not touch with surface of waste bottles.
Waste bottles wastage should be discarded when it will be fill.
Hot water should be run once in 07 days for about 1 hour for all HPLC instrument. This removes any particulate matter that may cause blockages, Pump plungers, seals
Buffers should be prepared freshly on the day required. and solvents
must be filtered through 0.45 micron filter paper.

Freshly prepared mobile phase should be thoroughly degassed to remove all dissolved gases with sonicator.
Ensure that the buffer is soluble in the proposed wash or storage phase. If it is not, first flush the system with a solvent mix that is highly aqueous to remove the buffer from the system and column , then change to the proposed wash or storage solvent mix.
integration procedure for assay

Integrate all samples and standards both with chromatograms within a sequence using the same integration events.
Check the each chromatogram within a sequence to ensure that integration is correct.
integration procedure for related substances

Integrate the standard chromatograms within a sequence using the same integration events.
If required same integration method for Blank, Placebo, system suitability injections can be prepared & in test injection, auto integration can be performed, if required.
Integrate the sample chromatograms within a sequence using the integration events and disregard the blank and placebo peaks. Integrate the known and unknown peaks

in sample chromatograms with using following types of integrations.

Check each chromatogram to ensure correct integration obtained for all relevant peaks. After peak integration, save the processing methods for Blank, Placebo, Standard, System suitability and sample injections.
baseline to baseline integration

integration baseline is a continuation of the existing baseline. Begin integration when the signal first begins to rise above the baseline and end when the signal returns to the original baseline.
types of integration practices

Pictorial representation of peak area, peak height and Gaussian peak shape are as follows:





Base to Base: The peak should be symmetrically shaped and exhibits no indication of co-elution. The baseline is stable

and returns to the same level. Peaks of this nature are usually appropriately integrated baseline-to-baseline integration.
Below mentioned is an example of improperly integrated peak. The tailing facet of the peak has important area unit that

has been eliminated that shall be enclosed within the peak and another one is an example for improperly elevated baseline.

This clearly excludes an outsized space of peak: during this case a base-to-base integration is correct apply.

 

When peak eluted on falling of baseline, excessive baseline area need to be eliminated as below manner:

When excessive peak tailing is observed peak shall be integrated covering its complete area

base-to-base with dropline:
The drop methodology involves addition of a vertical line from the valley between the peaks to the horizontal baseline

that is drawn between the beginning and stop points of the peak group.

VALLEY TO VALLEY: This timed events sets the baseline from the peak starts and where it ends in a less resolved peak Integrate each peak separately.

.System Suitability :

During HPLC analysis, Analyst injects the blank and standard preparation as per the test procedure and checks the chromatogram

for peak shape and system suitability parameters like Theoretical plates, Tailing factor and resolution, etc. if applicable.
System Suitability is established as per the procedure and shall be determined by the first chromatogram of replicate injections of standard solution except resolution.
Following system suitability criteria shall be consider if not mentioned in pharmacopoeia for Theoretical

plates NLT 1500, tailing factor NMT 2 and RSD of three replicate injection NMT 2.0%.
During analysis of Bulk and In-process sample 3 standards shall be injected.

Sequence:

Sequence of blank, standard and sample injections are followed as – Blank, replicate standard solution, duplicate

injection of sample solution for assay and one Bracketing standard solution or as specified in respective standard test procedure.

After injections of sample Solution (i.e. in assay) inject bracketing standard solution or as specified in the respective standard test procedure.
In case of content uniformity (CU), after 5 injections inject bracketing standard solution or as specified in the respective standard test procedure.
RSD shall be calculated with Bracketing solution and standard solution or specified in respective standard test procedure.
Assay calculation should be done using the mean area of replicate injections of standard solution.
An acceptance criteria of relative standard deviation (RSD) with bracketing standard is NMT 2.0%.
If the acceptance criteria for RSD not met for any bracketing standard the analysis in between of those run shall be

canceled with proper justification and authorization of Head-QC.
In case of Content uniformity and related substances single injection of sample solution shall be injected, rest

of the injection sequence shall remains the same as specified under assay.
At the end of sequence analyst shall add the column washing method.

Procedure for Column washing

Column washing should be performed with appropriate solvent.

Column should be stored in organic solvent.

After each batch table column washing should be added.

Column washing should be performed for at least 30 minutes.

Columns washing should be done by using ratio of Solvent and water in 30:70.

general:

Section In charge shall prepare the instrument method.

Before run the sequence, Instrument method shall be checked as per STP by Section In charge / designee.
documentation

System suitability chromatogram, sequence table, and chromatographic method (Instrument method) print shall be attach with sample report.
Every chromatogram and page which is compile with report shall be duly signed and checked.



7.0 ABBREVIATION:

SOP    Standard Operating Procedure

QC    Quality Control

STP   Standard Testing Procedure

CU    Content Uniformity

RSD   Relative Standard Deviation

8.0.
ANNEXURES :
NA

9.0
DISTRIBUTION:
Controlled Copy         Quality Control Department
Master Copy             Quality Assurance Department

10.0
REFERENCES:
In-House

 

 

SOP for testing of ETP water

sop for Calibration and Maintenance of Laboratory Instruments and Equipment

Disposal of Residual Sample or Left Over Material

sop for for Laboratory Incident

standard operating procedure temperature monitoring

sop for drying of silica gel

sop for operation of infrared moisture balance

sop for preparation of mobile phase

sop for treated RO water

sop for Preparation and Issuance of Analysis protocol standard

sop of placebo and impurity stock solutions

sop for disposal of residual sample  

sop for handling of pharmacopoeial changes

sop for procedure for operation of ultrasonic cleaner

difference between UPLC and HPLC

sop for for Emergency Eyewash and Shower

sop for operation and calibration of total organic carbon analyzers

sop for operation of cobb tester

sop for Operation and calibration of atomic absorption spectrophotometer

sop for Operation and calibration of gas liquid chromatograph

sop for operation of humidity oven

sop for operation and calibration of serological water bath

sop for monitoring of drain trap

sop for destruction of analytical samples after testing and control samples

sop for destruction of used chemicals

Sop for Operation of suction pump

sop for Operation and calibration uv cabinet

sop for Operation and calibration of bulk density apparatus

sop for operation and calibration of shore hardness tester

sop for operation of rub proofness tester

sop for monitoring of purified water

sop for Retesting of packaging materials

sop for Retesting and resampling of raw materials

sop for Control of issuance of record of analysis green sheets

sop for Control of computer passwords

sop for sampling of packaging materials PM

sop for sampling of sterile raw material

sop for sampling of intermediates and finished products

sop for operation and calibration of friability test apparatus

sop for approval and rejection of packaging materials

sop for non conformance of RM PM and finished product

sop for collection storage and disposal of control samples

sop for trend analysis of finished products

sop for Chromatographic practices and system suitability

SOP For Good Laboratory Practices

sop for cleaning and operation of sieve shaker

general specification of packing material cartons

sop for Password for Analytical Instrument and LIMS software

sop for Rounding off numerical analytical results

sop for sampling of bulk and finished product

sop for cleaning of spillage material

sop for Handling of Reference Standard

sop for hplc column maintenance and washing

procedure for sampling and handling of bulk sample

STP for borewell generation point (raw water storage tank)

sop for preparation and standardization of 0.1M Zinc Sulphate

Operation &calibration of analytical balance (dhona)

Operation and Calibration Procedure for Disintegration Test apparatus

sop for preparation and standardization of 1 M Hydrochloric Acid

Preparation and standardization of 0.1 M ceric ammonium sulphate solution

sop for preparation and standardization of 0.05 m iodine solution

validation of volumetric solution 0.1m ammonium thiocyanate

handling of reference standard and preparation of working standard

sop for water sampling and analysis

sop for operation for validation of excel worksheets

sop for laboratory cleaning

sop for stability of volumetric solutions

sop for preparation of raw material in process finish product packing material data sheets

sop for handling of hazardous chemicals

sop for handling of glassware and allocation of identification number

sop for operation cleaning and calibration of bursting strength tester

sop for rounding off the analytical test results

procedure for Analyst Qualification

sop for operation and calibration of dissolution Apparatus

procedure for maintenance of desiccators

sop for for hplc column receipt checking id no and regeneration

safety data sheet for laboratory chemicals

procedure for handling of poisonous chemicals

sop for cleaning of sampling devices

sop for calibration procedure of instruments

sop for specification and standard testing raw material packing material and finished product

procedure for operation and calibration of potentiometric titrator

procedure for operating and calibration of digital hardness tester

procedure for disposal of expired chemicals, reagents and solvents

sop for behavior in quality control department

sop for preparation and standardization 0.1M sodium thiosulphate

sop for preparation and standardization 0.1M Disodium Edetate

preparation and standardization 0.1M Sodium Hydroxide Solution

Preparation and standardization of 0.1M Perchloric acid solution

sop for preparation 0.05M edetate disodium

sop for preparation 0.1M silver nitrate

sop for Operation and Calibration of High Performance Liquid Chromatography

sop for UV & Visible Spectrophotometer

procedure for Cleaning of laboratory glassware

Cleaning of Instrument, Instrument bench and surrounding area of Quality

Safety Precaution in Quality Control Department

Operation & Calibration of Analytical Balance

Calibration of Glassware in Quality Control Department

handling of samples received in Quality Control

Cleaning and Operation of Refrigerator

Operation, Cleaning and Calibration of water bath

Operation & Calibration of Refractometer

Operation and Cleaning of Centrifuge Apparatus

cleaning, operation & calibration of Vernier caliper

Calibration of Fourier Transform Infrared Spectrophotometer (FTIR)

Cleaning and operation of Moisture Analyzer

Cleaning & Operation of Vacuum pump in Quality Control Department

Operation and Calibration of Polarimeter

Cleaning and operation of Magnetic Stirrer

Cleaning Operation and Calibration of Melting Point

Operation Cleaning and Calibration of Muffle Furnance in Quality Control Department

procedure of operation and Cleaning of Sonicator

Operation Cleaning & Calibration of pH meter in Quality Control Department

Entry and Exit in Quality Control Department

 

 

 

 

 

 

 

 

 

 

 

 

 

Leave a Comment

error: Content is protected !!