sop for Operation and calibration of gas liquid chromatograph

 

sop for Operation and calibration of gas liquid chromatograph

 

1.0. OBJECTIVE:

The objective of this SOP is:
1.1 To lay down a procedure for the operation and calibration of Gas Chromatograph (GC).

2.0. RESPONSIBILITY:

Officer/Executive – QC

3.0. ACCOUNTABILITY:

Head – Quality Control

4.0. PROCEDURE:

4.1 Connection Of Analytical Column In Column Oven In GC:

4.1.1 For glass columns use PTFE ferrules with hexagonal nuts or O-ring seals with round nuts to be tightened by hand.
4.1.2 For metal columns use double brass ferrules and hexagonal 1/4” column fixing nuts.
4.1.3 Insert the ends of the column as required into injector and detector base body.
4.1.4 Do not use spanners to fit the glass column.
4.1.5 Check the fittings leak with soap solution.
4.1.6 Check the flow with the help of soap bubble flow meter on the detector base body.
4.1.7 For capillary columns, position the column inside the oven on the column support in such a way that the correct alignment of limbs with detector and injector us easily obtained.



4.2 Performing the Leak check of carrier gas.

4.2.1 Ensure all other gas lines are closed.
4.2.2 Ensure all exits from the Chromatograph are closed. Use blanking caps provided for the detector base body, HWD outlets and SSL valves as appropriate.
4.2.3 Adjust the carrier gas pressure to 100-200 kPa with help of regulator in instrument.
4.2.4 Turn off carrier gas supply at the cylinder or other convenient position.
4.2.5 See that the pressure of all gauges remains unchanged for at least 3 minutes. Then be sure that the system is leak tight.

4.3 Operation Of Temperature Control Unit

4.3.1 Before starting the instrument check the following carefully.
4.3.2 The correct column is properly installed and conditioned.
4.3.3 Ensure Oven temperature limit is correct see.
4.3.4 Ensure sufficient gas supply of N2 gas, H2 gas and Zero Air.
4.3.5 All external connections are plugged in.
4.3.6 Injector septum where appropriate, is in good order.
4.3.7 Switch ‘On’ the control unit powder supply.
4.3.8 Switch ‘On’ the control unit at the front and rear. Display will light after beep.
4.3.9 “PROG MEM” will appear after 5 sec. And will show the current parameter file e.g‘ PO’.
4.3.10 Keyboard of control panel can be operated by two modes 1. “Direct mode of data entry” 2. “Indirect Mode of data entry”.
4.3.11 Switch “ON” the GC from back & front on/off Switch.
4.3.12 Check the display on front panel
4.3.13 Check functioning of keyboard by entering your require program.
4.3.14 Enter OVEN Temperature in Set valve of SB ISO 1for initial temperature &* ISO 2for final temperature.
4.3.15 Enter Injector temperature in set value of ZONE 1.

4.3.16 Enter Detector Temperature in set valve of Zone 2 & Zone 3.
4.3.17 Check all temperature by pressing Actual key.
4.3.18 Adjust flow of Hydrogen 35 ml/min. and Air 350 ml/min.
4.3.19 Ignite the flame.
4.3.20 Wait for ready status of GC.
4.3.21 Start Chrom – Card Software & check Baseline in view.
4.3.22 Zero the Base line by pressing Auto zero key.
4.3.23 Open/ prepare require method in Main Menu.

4.4 Calibration:



4.4.1 Temperature Calibration:
4.4.1.1 Set the measure Oven temperature, Injector temperature & Detector temperature as per operation procedure.
4.4.1.2 Calibrate measurement temperature with help of temperature sensor and record the details in Annexure – 1.
4.4.1.3 Frequency: Six Month

4.4.2 Flow Calibration:

4.4.2.1 Connect the flow meter to the column port and operate the Gas Chromatograph as per SOP.
4.4.2.2 Check the leak test by help of soap solution.
4.4.2.3 Measure the flow by using the flow meter.
4.4.2.4 Details note in Annexure – 2
4.4.2.5 Frequency: Six month

4.4.3 Detector Calibration:

4.4.3.1 Calibrate the Detector by using of IPA (Isopropyl Alcohol).
4.4.3.2 Set the temperature of oven, injector and Detector as per Operation procedure.
4.4.3.3 Set Flow rate as per Operation procedure.
4.4.3.4 Ignite flame by help of ignite key.
4.4.3.5 Injected 0.1 µl of IPA in triplicate and note down in Annexure – 2
4.4.3.6 For above both Tolerance Limit of max. 5%
4.4.3.7 Frequency: Six Month

4.4.4 Affix a calibration status label on the instrument containing “Calibrated On”, “Due On” and “Signature”.
4.4.5 Report to Head – QC, if any discrepancy observed during calibration or operating the instrument and affix ‘Under Maintenance’ label on the instrument.

4.4.6 In case of calibration failure, follow the procedure given in SOP

5.0. REASON FOR REVISION:

The SOP is revised to specify the System suitability test and incorporating the bracketing of standard after 10 injections.

6.0 TRAINING:

Trainer — Head – Quality Control
Trainees — Quality Control Chemists and Assistants

7.0 DISTRIBUTION:
Certified Copy : Quality Control
Reference Copy : Display copy Near Gas Chromatograph
Reference Copy : File copy in Gas Chromatograph Calibration Record
Original Copy : QUALITY ASSURANCE.

8.0 ANNEXURES:

Annexure – 1 : Format For Calibration Record Of Temperature
Annexure – 2 : Format For Calibration Record Of Flow Rate and Detector

9.0 REFERENCES:

In house.

 

 

SOP for testing of ETP water

sop for Calibration and Maintenance of Laboratory Instruments and Equipment

Disposal of Residual Sample or Left Over Material

sop for for Laboratory Incident

standard operating procedure temperature monitoring

sop for drying of silica gel

sop for operation of infrared moisture balance

sop for preparation of mobile phase

sop for treated RO water

sop for Preparation and Issuance of Analysis protocol standard

sop of placebo and impurity stock solutions

sop for disposal of residual sample  

sop for handling of pharmacopoeial changes

sop for procedure for operation of ultrasonic cleaner

difference between UPLC and HPLC

sop for for Emergency Eyewash and Shower

sop for operation and calibration of total organic carbon analyzers

sop for operation of cobb tester

sop for Operation and calibration of atomic absorption spectrophotometer

sop for Operation and calibration of gas liquid chromatograph

sop for operation of humidity oven

sop for operation and calibration of serological water bath

sop for monitoring of drain trap

sop for destruction of analytical samples after testing and control samples

sop for destruction of used chemicals

Sop for Operation of suction pump

sop for Operation and calibration uv cabinet

sop for Operation and calibration of bulk density apparatus

sop for operation and calibration of shore hardness tester

sop for operation of rub proofness tester

sop for monitoring of purified water

sop for Retesting of packaging materials

sop for Retesting and resampling of raw materials

sop for Control of issuance of record of analysis green sheets

sop for Control of computer passwords

sop for sampling of packaging materials PM

sop for sampling of sterile raw material

sop for sampling of intermediates and finished products

sop for operation and calibration of friability test apparatus

sop for approval and rejection of packaging materials

sop for non conformance of RM PM and finished product

sop for collection storage and disposal of control samples

sop for trend analysis of finished products

sop for Chromatographic practices and system suitability

SOP For Good Laboratory Practices

sop for cleaning and operation of sieve shaker

general specification of packing material cartons

sop for Password for Analytical Instrument and LIMS software

sop for Rounding off numerical analytical results

sop for sampling of bulk and finished product

sop for cleaning of spillage material

sop for Handling of Reference Standard

sop for hplc column maintenance and washing

procedure for sampling and handling of bulk sample

STP for borewell generation point (raw water storage tank)

sop for preparation and standardization of 0.1M Zinc Sulphate

Operation &calibration of analytical balance (dhona)

Operation and Calibration Procedure for Disintegration Test apparatus

sop for preparation and standardization of 1 M Hydrochloric Acid

Preparation and standardization of 0.1 M ceric ammonium sulphate solution

sop for preparation and standardization of 0.05 m iodine solution

validation of volumetric solution 0.1m ammonium thiocyanate

handling of reference standard and preparation of working standard

sop for water sampling and analysis

sop for operation for validation of excel worksheets

sop for laboratory cleaning

sop for stability of volumetric solutions

sop for preparation of raw material in process finish product packing material data sheets

sop for handling of hazardous chemicals

sop for handling of glassware and allocation of identification number

sop for operation cleaning and calibration of bursting strength tester

sop for rounding off the analytical test results

procedure for Analyst Qualification

sop for operation and calibration of dissolution Apparatus

procedure for maintenance of desiccators

sop for for hplc column receipt checking id no and regeneration

safety data sheet for laboratory chemicals

procedure for handling of poisonous chemicals

sop for cleaning of sampling devices

sop for calibration procedure of instruments

sop for specification and standard testing raw material packing material and finished product

procedure for operation and calibration of potentiometric titrator

procedure for operating and calibration of digital hardness tester

procedure for disposal of expired chemicals, reagents and solvents

sop for behavior in quality control department

sop for preparation and standardization 0.1M sodium thiosulphate

sop for preparation and standardization 0.1M Disodium Edetate

preparation and standardization 0.1M Sodium Hydroxide Solution

Preparation and standardization of 0.1M Perchloric acid solution

sop for preparation 0.05M edetate disodium

sop for preparation 0.1M silver nitrate

sop for Operation and Calibration of High Performance Liquid Chromatography

sop for UV & Visible Spectrophotometer

procedure for Cleaning of laboratory glassware

Cleaning of Instrument, Instrument bench and surrounding area of Quality

Safety Precaution in Quality Control Department

Operation & Calibration of Analytical Balance

Calibration of Glassware in Quality Control Department

handling of samples received in Quality Control

Cleaning and Operation of Refrigerator

Operation, Cleaning and Calibration of water bath

Operation & Calibration of Refractometer

Operation and Cleaning of Centrifuge Apparatus

cleaning, operation & calibration of Vernier caliper

Calibration of Fourier Transform Infrared Spectrophotometer (FTIR)

Cleaning and operation of Moisture Analyzer

Cleaning & Operation of Vacuum pump in Quality Control Department

Operation and Calibration of Polarimeter

Cleaning and operation of Magnetic Stirrer

Cleaning Operation and Calibration of Melting Point

Operation Cleaning and Calibration of Muffle Furnance in Quality Control Department

procedure of operation and Cleaning of Sonicator

Operation Cleaning & Calibration of pH meter in Quality Control Department

Entry and Exit in Quality Control Department

Leave a Comment

error: Content is protected !!