Operation and Calibration of Sartorius Analytical Balance
1.0 Objective
To lay down a Procedure for Operation and calibration of Sartorius Analytical Balance.
2.0 Scope
This Standard Operating Procedure is applicable for Operation and calibration of Sartorius Analytical Balance in Quality Control Department of abc company.
3.0 Responsibility
3.1 Officer/Executive QC or his designee shall be responsible to followed the SOP.
3.2 Head QC or his designee shall be responsible to ensure the activity as per SOP.
3.3 Head QA or his designee shall be responsible for implementation and compliance with SOP.
4.0 Abbreviations and Definitions
SOP : Standard Operating Procedure
QC : Quality Control
QA : Quality Assurance
5.0 Procedure
5.1 Operation procedure
5.1.1 Check and ensure that the instrument is clean and free from dust.
5.1.2 The level of the balance is properly adjusted by adjusting the bubble at the center.
5.1.3 Connect the balance to the power supply.
5.1.4 Press the ‘I/Q’ key to “ON/OFF” the balance. The balance will perform self test and “0.000 g” is displayed.
5.1.5 If the balance shows a small circle at the right top of the display it means the balance is in “OFF” mode.
5.1.6 If the balance shows a small circle at the left bottom of the display it means the balance is in “STAND BY” mode.
5.1.7 If the balance shows “0.000g” on the display it means balance is “STABLE”.
5.1.8 In order to take a weight, Open the side window and place the object or butter paper on which
the weight has to be taken on the pan. After the weight, unit “g” or other selected unit is displayed as the stability symbol.
5.1.9 Place the specimen on the object or butter paper (tare the weight of the object or butter paper)
and use the balance for weighing. In case the butter paper is used for weighing carefully transfer the
specimen to the respective container and again place the same butter paper on the pan.
5.1.10 Clean the pan and the platform of the balance using a brush after the weighing.
5.1.11 Make entry of balance usage in instrument log.
5.2 Handling of Weight Box
5.2.1 Store the weight box in dust free area.
5.2.2 Do not touch the weight with bare hand, for handling the weights use a forceps.
5.2.3 Use certified weights only.
5.2.4 For cleaning of weights use a lint free cloth moistened with small amount of Di-ethyl ether and air dry.
5.2.5 Do not drop the weight on the pan to avoid any damage to the balance. Gently place the weight
on the center of the pan to eliminate corner weighing differences.
5.3 General Instruction
5.3.1 Clean the pan and Platform of the balance after weighing is over.
5.3.2 Do not over load the balance of their maximum capacity.
5.3.3 Keep doors of the balance always closed.
5.3.4 Before closing and opening, ensure that the pan is empty.
5.3.5 Keep the balance in “STAND BY” position when not in use, by pressing I/Q’ key of the balance.
5.4 Cleaning
5.4.1 Before cleaning, disconnect the balance from the power supply.
5.4.2 Do not use any aggressive cleaning agents or strong solvents.
5.4.3 Use a piece of cloth which has been wet with a mild detergent (soap) to clean the weighing pan and balance housing.
5.4.4 Make sure that no liquid enters the balance housing.
5.4.5 After cleaning, wipe down the balance with a soft, dry piece of cloth.
5.4.6 Carefully remove any sample residue / spilled powder by using a brush that is placed with the balance.
5.5 Daily verification of Analytical Balance
5.5.1 Internal calibration (Self calibration/ Auto calibration)
5.5.1.1 Press “CAL” key. Balance will go into automatic self calibration mode and “CAL” will be displayed,
the “CC” will be displayed after self calibration is over display will show 0.000 g, then balance is
ready for use. Record the observation in the given Annexure-1.
5.5.2 With Secondary standard weight
5.5.2.1 Calibrate the balance daily with the standard weights of 1g, 10g, 100 g & 200g individually. Record
the results as given in Annexure – 1.
5.5.2.2 Acceptance Criteria: ±0.1% of the Certified weight.
5.6 Monthly Calibration of Analytical Balance
5.6.1 Measurement of Uncertainty
5.6.1.1 Ensure that the standard weights used are within the validity period of its calibration.
5.6.1.2 The weights selected for uncertainty measurement are 1g, 5g & 10 g.
5.6.1.3 Measure the weights by keeping the standard weight at the positions (10 times, individually) as
given in the following diagram.
5.6.1.4 Calculate mean and standard deviation (S.D) for the weights taken above.
5.6.1.5 Calculate the uncertainty for all the selected weights by using following formula:
5.6.1.6 Acceptance criteria for uncertainty measurement: NMT 0.001.
5.6.1.7 Record the observations in annexure-2.
5.6.2 Calibration with Standard weights
5.6.2.1 Place 1 g certified weight in the centre of the pan. Close the door.
5.6.2.2 Wait until the weighing result is stable.
5.6.2.3 Now read the displayed weight.
5.6.2.4 Take out the 1 g weight and allow the balance to come to zero.
5.6.2.5 Repeat the above procedure for the certified weights of 2g, 5g, 10g, 20g, 50g, 100g and 200g.
5.6.2.6 Record the observations in annexure-2.
5.6.2.7 Acceptance Criteria: ±0.1% of the Certified weight.
6.0 Forms and Records
6.1 Daily verification record of Analytical Balance – Annexure-1
6.2 Monthly calibration of Analytical Balance – Annexure-2
7.0 Distribution
7.1 Master copy – Documentation Cell (Quality Assurance)
7.2 Control copy – Quality Control
8.0 History
Revision Number | Details For Change |
Reason for Revision |
00 | New SOP | NA |
Standard testing procedure of Iron Sucrose Injection
Standard testing procedure lactose
Standard testing procedure mefenamic acid
standard testing procedure domperidone
standard testing procedure flavour mixed fruit
standard testing procedure dicyclomine hydrochloride
Standard testing procedure honey pure
standard testing procedure dextromethorphan hydrobromide
standard procedure of levocarnitine injection
Analysis of Ivermectin Suspension
standard testing procedure artemether injection
standard testing procedure artemether injection
standard testing procedure Carbocisteine syrup
standard testing procedure Phytomenadione injection
standard testing procedure serratiopeptidase
standard testing procedure starch IP
standard testing procedure sucrose refined sugar
standard testing procedure titanium dioxide
standard testing procedure tramadol hydrochloride
standard testing procedure zinc sulphate
standard testing procedure croscarmellose sodium
standard testing procedure colour erythrosine supra
standard testing procedure magnesium hydroxide
standard testing procedure diclofenac sodium
standard testing procedure dibasic calcium phosphate
standard testing procedure cyanocobalamin
standard testing procedure cholecalciferol
standard testing procedure Calcium carbonate oyster shell powder
standard test procedure Calcium Citrate
standard testing procedure Bronopol
standard testing procedure Bromhexine Hydrochloride
Standard Testing Procedure diclofenac sodium injection
Standard Testing Procedure Drotaverine Hydrochloride injection
analysis of Procyclidine HCl 5 mg Tablet
analysis of silodosin 8 mg capsule
Method of Analysis Ceftriaxone and Sulbactam For Injection
Analysis of Glimepiride and Metformin Hydrochloride (SR) Tablets
General Test Procedure Melting Range and Freezing Point
General Test Procedure Weight per ML and Specific Gravity
General Test Procedure Distillation Range
General Test Procedure Non Volatile Matter and Residue on Evaporation
General Test Procedure Clarity of Solution
General Test Procedure Limit Test Chloride
General Test Procedure Limit Test Iron
General Test Procedure Limit Test Heavy Metals
General Test Procedure Water Soluble Substances
General Test Procedure Acid Soluble Substances
General Test Procedure 0.1M Ammonium Thiocyanate
General Test Procedure 0.1M Ethanolic Sodium Hydroxide
Calibration Of Disintegration Test Apparatus
Procedure for Glass Ware Cleaning
Operation and Calibration of Sartorius Analytical Balance