Site icon Pharma Dekho

sop for Operation of Double Door Autoclave

 

Operation of Double Door Autoclave

 

1.0 Objective
To lay down a procedure for the Operation of Double Door Autoclave.
2.0 Scope
This Standard Operating Procedure is applicable for Operation of Double door Autoclave in Microbiology

Laboratory Quality control Department ABC Company.
3.0 Responsibility
3.1 Officer / Executive Quality Control are responsible for operation of double door autoclave.
3.2 Head Quality control is responsible for compliance of this SOP.
4.0 Abbreviations and Definitions
SOP : Standard Operating Procedure
PLC : Programmed logic Controller.
LCD : Liquid crystal Display
NST : Non Sterile
5.0 Procedure
5.1 Generation operating Instructions
5.1.1 Ensure the compressed air supply is available at appropriate utility location.
5.1.2 Ensure air pressure of at least 6.00 kg/cm2.
5.1.3 Ensure gasket lubrication is proper and the gasket pressure 3.00 kg/cm2.
5.1.4 Switch ON the main provided on panel board.
5.1.5 Opening the door of the Non Sterile end of the double door autoclave; Press door open button present on

Non sterile side of the panel board.
5.1.6 Now Load the material to be autoclave from NST side.
5.1.7 In HPHV cycle load 120 petri plates, 20 Bottles, 02 dresses & 10 hand gloves or In Media cycle

kept 2.5 liter SCDA & 1.5 liter SCDM media.
5.1.8 After loading the material Press the close button present on the NST Side and Door

pre conduction indicator will glow.
5.2 Start up Parameter
5.2.1 When all faults are clear & both doors are closed then follow the start up parameters

with the help of operating panel present on the panel board.
5.2.2 The Start up parameters of the operating panel along with the LCD display are mentioned as below.

Parameter Mode of Operation
LCD SHOWS MAIN MENU
Press F1 Key SET/VIEW PARAMETER
Press F2 Key LOGIN
Press F 3 MAIN MENU-2

5.2.3 Press F3 Key for entering main menu-2 and than LCD will display are mention as below

Parameter Mode of Operation
LCD SHOWS MAIN MENU- 2
Press F3 Key START UP SCREEN
Press F1 Key BATCH DATA
Press F 2 Key MONITOR SCREEN
Press F 4 Key MAIN MENU

5.2.4 Press F3 Key for start up screen and than LCD will display are Mention as below.

Parameter Mode of Operation
LCD SHOWS [PLEASE SELECT PRO.]
Press F1 Key START UP PROCESS
Press F2 Key PRINT PARAMETER
Press F 3 Key PRO.SELECT
Press F 4 Key NEXT

5.2.5 Select the process by press F3 key and use the arrow F2 key for print out the parameter.
5.2.6 For start up process use UP and down arrow key to select F1and pressF1 key for start up cycle.
5.2.7 Similarly switch ON the Dot matrix printer for print the data.
5.2.8 For charter recorder press Func Key than screen will display print data, after that press CHUP Key twice

for print the charte. Now chart printing will start and press the Record key for recording the chart.
5.2.9 When process will complete than LCD screen display End process.
5.2.10 Now press the Open button for Open the Door from sterile side and unload the Material from autoclave.
5.2.11 After unload the material, push the door close button and press Ask button when sterile side door close

indication will glow.
5.2.12 Now check the general parameter for next cycle operation and fallow 5.2.1 to 5.2.11 points.
5.2.13 Start the Dot matrix Printer attached with the PLC so that it will record the complete operating

cycle Parameters and the temperature throughout the cycle. Preserve the Temperature chart in an

authorized record for each and every cycle operation.
6.0 Forms and Records
6.1 Autoclave log Book – Annexure-1

7.0 Distribution
7.1 Master copy – Documentation Cell (Quality Assurance)
7.2 Control copy – Quality Control
8.0 History

Revision Number Details For Change
Reason for Revision
00 New SOP NA

 

sop for calibration and validation of micro autoclave

sop for Sterility failure investigation

cleaning and operation of discard autoclave

sop for operation of fogger machine

sop for Biological assay of lactic acid bacillus

sop for preparation of culture inoculum

STP for sterility testing of sterile gloves

sop for Operation and calibration of active air sampler

sop for transfer of material for testing and sampling in sterile area

entry & exit procedure in microbiology laboratory

Growth Promotion Test In Microbiology Laboratory

Operation of B.O.D in Microbiology Laboratory




Operation of Horizontal Laminar Air Flow in the microbiology laboratory

Operation and cleaning of Pass Box.

Operation and cleaning of air sampler

Cleaning and Sterilization of Glassware

Analysis of water for microbial load in microbiology lab

Operation and temperature monitoring of Refrigerator

Fumigation of Microbiology Laboratory.

Entry & Exit procedure In Sterility Area

SOP for Microbial analysis of Raw Material Finished Products

SOP for Operation & Calibration of pH meter in Micro Department

SOP for Operation & Calibration of pH meter in Micro Department

SOP Operation and calibration of Hot Air Oven In Microbiology

SOP for operation cleaning & calibration of Digital colony counter

SOP for Operation And Cleaning of Microscope

sop for Media Preparation and Consumption

sop for Receipt Storage and Usage of Culture Media

sop for Cleaning Sanitization And Disinfection In Microbiology

sop for Environmental monitoring of all the Classified area

sop for Handling and Sub culturing of Microbial cultures

sop for Media Growth Promotion Test and various Microbiological test

sop for BOD incubator operation and cleaning

sop sampling of water for microbiological analysis

sop for Disinfectant Efficacy Test

sop for for cleaning and operation of vortex mixture

sop for Temperature & Relative Humidity Monitoring

sop for Operation and Calibration of Heating Block

sop for Sterility Testing of Microbiology

sop for Disposal of Culture Media

sop for Drain point of Microbiology

sop for entry & exit procedure In Microbial limit test and Biosafety

sop for Gram Staining of Bacteria in Microbiology Laboratory

sop for Monitoring of Compressed Air/gases for microbiological analysis

sop for BET (Bacterial Endotoxin) test in Microbiology

sop for receipt storage and Determining the population of Biological indicators

sop for qualification of analyst microbiologist

sop for Bioburden test of Packing materials in Microbiology Laboratory

sop for microbiological assay of erythromycin antibiotic

sop for liquid particle counter

sop for operation and calibration of digital zone reader

sop for monitoring of ultraviolet efficiency LAF and pass box

microbiological assay of cyanocobalamin or vitamin B12

gowning procedure for microbiological testing area

swab testing of various surfaces for bioburden determination

sop for endotoxin challenge test

Hold time study protocol for sterilized media

sop for personnel Qualification protocol for aseptic area

sop for sampling and testing of drain water

Sop for Operation of Airborne Particle Counter

sop for Inoculum Preparation

sop for Validation protocol of steam sterilizer autoclave

sop for pathogen detection from drain point

Sop for Analysis of Raw water Purified water water for injection and pure steam water

sop for preservatives efficacy test

sop for collection and preservation of in house isolated microorganisms

sop for Operation Calibration and Maintenance of Micropipette

Sop for UV Efficacy Test

sop for gram staining

Sop for swab testing

sop for microbiological testing of water

PROCEDURE FOR FUMIGATION

sop for depyrogenation of apparatus

sop for media preparation

sop for fertility test growth promotion test of media

sop for Operation and cleaning of moist heat sterilizer

sop for monitoring by active air sampler

sop for swab sampling and testing for clean rooms in production area

sop for monitoring in microbiology laboratory

sop for Fumigation of aseptic area and microbiology lab

sop for monitoring of personnel in aseptic area

sop for maintenance of cultures

sop for Operation and cleaning of laminar bench

preparation of settle plates

sop for monitoring of pure steam

sop for entry and exit procedure to m.l.t and b.e.t room

sop for storage of and use of media

sop for disposal of microbiological media and cleaning of microbiological glassware

Preparation Filtration Usage and Destruction of Disinfectant Solution

Entry Exit and gowning Procedure in Microbiology Laboratory

Safety measures to be followed in microbiology laboratory

Growth Promotion Test of Media

Preservation and Maintenance of Microbial Strain

Entry and Exit in Microbiology Sterile Area

Entry and Exit in Microbiology Limit testing area

Water sampling of raw water purified water WFI and PSG in microbiology Lab

sop of Microbiological assay

Microbial monitoring of Drain points

sop for Operation of Double Door Autoclave

 

 

Exit mobile version