sop for endotoxin challenge test
1.0 OBJECTIVE :
To lay down a procedure for Endotoxin Challenge Test.
2.0 SCOPE :
This SOP is applicable for Endotoxin Challenge Test in Microbiology Lab. Of Quality Control
Department
3.0 RESPONSIBILITY :
3.1 Officer / Executive – QC (Microbiology)
4.0 ACCOUNTABILITY :
4.1 Head – QC
5.0 DEFINITION :
The Endotoxin Challenge Vial (ECV) is used in the validation of dry heat depyrogenation cycles. The ability of a particular oven / Tunnel cycle to destroy / inactivate endotoxin is measured by comparing the endotoxin level in baked ECVs vs unbaked control ECVs.
6.0 PROCEDURE:
6.1 material & Instruments :
6.1.1 Limulus Amebocyte Lysate Reagent
6.1.2 Endotoxin Indicator Vial (100000 EU/ml)
6.1.3 LAL Reagent Water
6.1.4 Ampoules or Vial according to requirement
6.1.5 Depyrogenated Dilution Tubes (13×100 mm)
6.1.6 Depyrogenated Assay Tubes (10×75 mm)
6.1.7 Micropipette with Pyrogen free tip (20-200 µl)
6.1.8 Micropipette with Pyrogen free tip (100-1000 µl)
6.1.9 Vortex Mixer
6.1.10 Heating Block
6.2 Preparation Of Challenge Vials :
6.2.1 Reconstitute the Challenge Vial of Endotoxin (100000 EU/ml) with 10 ml LRW to yield 10,000 EU/ml) and vortex according to manufacturer”s instructions.
6.2.2 Transfer 0.1 ml aliquot in to ampoules or vials used for Challenge Test (10000 EU/Vial).
6.2.3 Dry the ampoules or vials in Laminar Air Flow for overnight . Each ampoule or vial now contains 10000 EU/Vial. Mark the above prepared Ampoules/Vials as 1 to 10 numbers.
6.2.4 Keep at least 1 Ampoule/Vial as positive control (do not expose through Oven / Tunnel).
6.2.5 Mark the remaining ampoules/vials as NPC.
6.2.6 Expose these Ampoules/Vial to appropriate location in DHS/Tunnel as per its typical Depyrogenation Cycle.
6.3 Dilution Of Positive Control Ampoule / Vial :
6.3.1 Reconstitute the Ampoules/Vials with 1 ml LRW and vortes vigorously for 15 minutes and each subsequent dilution for 2-4 minutes.
6.3.2 Now the concentration of Endotoxin in the PPC will be 10000 EU/ml.
6.3.3 Prepare 1:20 dilution of the 10000 EU/ml to obtain 500 EU/ml.
6.3.4 Prepare 1:20 dilution of the 500 EU/ml to obtain 25 EU/ml.
6.3.5 Prepare 1:25 dilution of the 25 EU/ml to obtain 01 EU/ml.
6.3.6 From the above 1 EU/ml preparation, prepare a two fold dilution series up to 2λ, λ, λ/2, λ/4, where λ = Labelled Lysate Sensitivity, λ = 0.125 EU/ml.
DILUTION TABLE
Sr. No. | ENDOTOXIN | LRW | ENDOTOXIN CONCENTRATION (EU/ML) |
1. | 10000 EU/Vial | 1 ml | 10000 EU/ml |
2. | 0.1 ml of 10000 EU/ml | 1.9 ml | 500 EU/ml |
3. | 0.1 ml of 500 EU/ml | 1.9 ml | 25 EU/ml |
4. | 0.1 ml of 25 EU/ml | 2.4 ml | 01 EU/ml (8 λ) |
5. | 1 ml of 01 EU/ml | 1 ml | 0.5 EU ml (4 λ) |
6. | 1 ml of 0.5 EU/ml | 1 ml | 0.25 EU ml (2 λ) |
7. | 1 ml of 0.25 EU/ml | 1 ml | 0.125 EU ml (λ) |
8. | 1 ml of 0.125 EU/ml | 1 ml | 0..0625 EU ml (λ/2) |
9. | 1 ml of 0.0625 EU/ml | 1 ml | 0.0325 EU ml (λ/4) |
6.4 dilution And Test Of Npc Ampoules / Vial :
6.4.1 It is assumed that three log reduction is achieved after exposure of ampoules / vials in oven /Tunnel, the Endotoxin concentration in the vial is 10 EU/Vial now.
6.4.2 Reconstitute each ampoules/Vials with 1 ml LRW and vortes vigorously for 15 minutes and the Endotoxin concentration in the each ampoule or vial is 10 EU/ml.
6.4.3 Take 0.1 ml from above each vials or ampoules of 10 EU/ml in assay tubes as duplicate and add 0.1 ml Lysate which have the potency 0.125 EU/ml.
Dilution Table
Sr. No. | Endotoxin | Lysate | LRW | Endotoxin Concentration (Eu/ml) |
1. | 10 EU/Vial or Ampoule Assumed | – | 1 ml | 10 EU/ml |
2. | 0.1 ml of 10 EU/Vial | 0.1 ml | – | 01 EU/ml |
6.5 LAL Test Procedure :
6.5.1 Test the two fold dilution series prepared from the positive controls ampoules / vial in duplicate.
6.5.2 Test the 10 EU/ml dilutions prepared from each of exposed ampoules / Vial.
6.5.3 Test should be carried out in Clean depyrogenated 10×75 mm assay tubes only.
6.5.4 Protocol for Positive Control :
Sr. No. | Dilutions | CSE Dilution Used | LRW | Lysate in µl | No. of Replicates |
1. | 2λ | 100 µl of 2λ | – | 100 µl | 2 |
2. | λ | 100 µl of λ | – | 100 µl | 2 |
3. | λ/2 | 100 µl of λ/2 | – | 100 µl | 2 |
4. | λ/4 | 100 µl of λ/4 | – | 100 µl | 2 |
5. | Negative Water Control (NWC) | – | 100 µl | 100 µl | 2 |
6.5.5 Protocol for Negative Product Control (Challenged Ampoules/Vial):
Ampoule No. | Dilutions | Endotoxin Indicator Dilution (01 EU/ml) | Lysate in µl | No. of Replicates |
1. | NPC | 100 µl | 100 µl | 2 |
2. | NPC | 100 µl | 100 µl | 2 |
3. | NPC | 100 µl | 100 µl | 2 |
4. | NPC | 100 µl | 100 µl | 2 |
5. | NPC | 100 µl | 100 µl | 2 |
6. | NPC | 100 µl | 100 µl | 2 |
7. | NPC | 100 µl | 100 µl | 2 |
8. | NPC | 100 µl | 100 µl | 2 |
9. | NPC | 100 µl | 100 µl | 2 |
10. | NPC | 100 µl | 100 µl | 2 |
11. | NPC | 100 µl | 100 µl | 2 |
12. | NPC | 100 µl | 100 µl | 2 |
6.6 Calculation :
6.6.1 Formula :
Endotoxin Concentration = Lysate sensitivity × Reconstitued Volume × Dilution
For Unbaked (Positive Control) Vials = 0.125 EU/ml × 1 ml/Vial × 10,000 = 1250 EU/ml
For Baked (Heat Treated) Vials = 0.125 EU/ml × 1 ml/Vial × 1 = 0.125 EU/ml
Calculate the minimum log reduction as follows :
Minimum Log Reduction = Log Endotoxin Concentration of the Unbaked Control – Log Endotoxin Concentration of the Baked Vials.
OR
Minimum Log Reduction = Log value of recovered Endotoxin from positive control – Log value of recovered sample from heat treated sample.
6.7 Interpretation Of Results / Acceptance Criteria :
6.7.1 Test results are valid of recovery of Endotoxin in non exposed vials is with in a two fold dilution of the labeled claim.
6.7.2 The depyrogenation cycle is considered as successfully validated if there is more than 3 log reduction is achieved in challenge Endotoxin vials exposed in to Oven/ Tunnel at specified place. Heat Exposed Vials.
6.7.3 For a valid Depyrogenation cycle, the PPC must be positive and NPC’s must be negative indicating a greater than 3 log reduction of pyroburden.
6.7.4 Record the “ Endotoxin Challenge Test Record in Annexure –I.
7.0 ABBREVIATION :
SOP Standard Operating Procedure
No. Number
QA Quality Assurance
QC Quality Control
QM QC-Microbiology
Ltd. Limited
GPT Growth Promotion Test
LAF Laminar Air Flow
°C Degree Celsius
ml Milliliter
NLT Not Less Than
PPC Positive Product Control
NPC Negative Product Control
LRW LAL Reagent Water
EU Endotoxin Unit
ml Mililitre
µl Micron litre
λ Sensitivity of the Lysate
8.0 ANNEXUREURE :
Annexure No. | Title of Annexure | Format No. |
Annexure – I | Endotoxin Challenge Test Report |
9.0 DISTRIBUTION :
- Master Copy Quality Assurance Department
- Controlled Copy No. 01 Quality Assurance
- Controlled Copy No. 02 Quality Control (Microbiology)
sop for calibration and validation of micro autoclave
sop for Sterility failure investigation
cleaning and operation of discard autoclave
sop for operation of fogger machine
sop for Biological assay of lactic acid bacillus
sop for preparation of culture inoculum
STP for sterility testing of sterile gloves
sop for Operation and calibration of active air sampler
sop for transfer of material for testing and sampling in sterile area
entry & exit procedure in microbiology laboratory
Growth Promotion Test In Microbiology Laboratory
Operation of B.O.D in Microbiology Laboratory
Operation of Horizontal Laminar Air Flow in the microbiology laboratory
Operation and cleaning of Pass Box.
Operation and cleaning of air sampler
Cleaning and Sterilization of Glassware
Analysis of water for microbial load in microbiology lab
Operation and temperature monitoring of Refrigerator
Fumigation of Microbiology Laboratory.
Entry & Exit procedure In Sterility Area
SOP for Microbial analysis of Raw Material Finished Products
SOP for Operation & Calibration of pH meter in Micro Department
SOP for Operation & Calibration of pH meter in Micro Department
SOP Operation and calibration of Hot Air Oven In Microbiology
SOP for operation cleaning & calibration of Digital colony counter
SOP for Operation And Cleaning of Microscope
sop for Media Preparation and Consumption
sop for Receipt Storage and Usage of Culture Media
sop for Cleaning Sanitization And Disinfection In Microbiology
sop for Environmental monitoring of all the Classified area
sop for Handling and Sub culturing of Microbial cultures
sop for Media Growth Promotion Test and various Microbiological test
sop for BOD incubator operation and cleaning
sop sampling of water for microbiological analysis
sop for Disinfectant Efficacy Test
sop for for cleaning and operation of vortex mixture
sop for Temperature & Relative Humidity Monitoring
sop for Operation and Calibration of Heating Block
sop for Sterility Testing of Microbiology
sop for Disposal of Culture Media
sop for Drain point of Microbiology
sop for entry & exit procedure In Microbial limit test and Biosafety
sop for Gram Staining of Bacteria in Microbiology Laboratory
sop for Monitoring of Compressed Air/gases for microbiological analysis
sop for BET (Bacterial Endotoxin) test in Microbiology
sop for receipt storage and Determining the population of Biological indicators
sop for qualification of analyst microbiologist
sop for Bioburden test of Packing materials in Microbiology Laboratory
sop for microbiological assay of erythromycin antibiotic
sop for liquid particle counter
sop for operation and calibration of digital zone reader
sop for monitoring of ultraviolet efficiency LAF and pass box
microbiological assay of cyanocobalamin or vitamin B12
gowning procedure for microbiological testing area
swab testing of various surfaces for bioburden determination
sop for endotoxin challenge test
Hold time study protocol for sterilized media
sop for personnel Qualification protocol for aseptic area
sop for sampling and testing of drain water
Sop for Operation of Airborne Particle Counter
sop for Validation protocol of steam sterilizer autoclave
sop for pathogen detection from drain point
Sop for Analysis of Raw water Purified water water for injection and pure steam water
sop for preservatives efficacy test
sop for collection and preservation of in house isolated microorganisms
sop for Operation Calibration and Maintenance of Micropipette
sop for microbiological testing of water
sop for depyrogenation of apparatus
sop for fertility test growth promotion test of media
sop for Operation and cleaning of moist heat sterilizer
sop for monitoring by active air sampler
sop for swab sampling and testing for clean rooms in production area
sop for monitoring in microbiology laboratory
sop for Fumigation of aseptic area and microbiology lab
sop for monitoring of personnel in aseptic area
sop for maintenance of cultures
sop for Operation and cleaning of laminar bench
sop for monitoring of pure steam
sop for entry and exit procedure to m.l.t and b.e.t room
sop for storage of and use of media
sop for disposal of microbiological media and cleaning of microbiological glassware