sop for BET (Bacterial Endotoxin) test in Microbiology
1.0 OBJECTIVE
1.1 To down the bacterial endotoxin in WFI & finished pharmaceutical product.
2.0 SCOPE
2.1 This procedure applicable for Bacterial Endotoxin Test in Microbiology Laboratory .
3.0 RESPONSIBILITY
3.1 Microbiologist QC
4.0 ACCOUNTABILITY
4.1 QC Manager
5.0 PROCEDURE
5.1 MATERIALS REQUIRED:
5.1.1 LAL Reagent Water, Control Standard Endotoxin, LAL Reagent, Depyrogenated Micro tips (100 μl &1000 μl), Glass test tube (depyrogenated), Sample to be tested.
5.2 EQUIPMENTS REQUIRED:
5.2.1 Micro pipettes (1000 μl &100 μl ), Vortex Mixer, Heating block
5.3 PROCEDURE FOR TESTING:
5.4 Preparation of Maximum valid dilution:
5.4.1 When the endotoxin limit in the substance or preparation being examined is specified in terms of
weight or units of active drug, the MVD may be calculated by the following formula:
Endotoxin Limit (in EU/ ml /mg) X Potency of product
MVD = ———————————————————————
Labeled sensitivity of lysate in EU /ml (l)
Where Potency =Concentration of the drug in mg / ml or unit /ml
5.4.2 When the endotoxin limit in the sample preparation is specified in terms of volume, the MVD may be
calculated by the following formula.
Endotoxin Limit (EU/ml)
MVD = —————————————————–
Labeled sensitivity of lysate in EU/ml (λ)
5.4.3 Dilute the sample MVD/2 times with LAL water when performing the test at MVD. If performing the test
at MVD/2, dilute the sample MVD/4time with LAL water.
5.5 Preparation of E-Coli Control Standard Endotoxin (CSE):
5.5.1 Reconstitute the CSE with appropriate volume (as stated on certificate) of LAL reagent water vortex
continuous 30 minutes at the interval of 10minutes (When new bottles reconstitute) & vortex vigorously
for five minutes before further dilution
5.6 Preparation of CSE dilution series:
5.6.1 Confirm the labeled sensitivity using at least one vial of each batch/lot of lysate. Prepare a series
of two-fold dilution of the CSE to give concentrations of 2λ, λ, 0.5λ and 0.25λ, where λ is the labeled
sensitivity of the lysate in EU per ml.
5.6.2 Perform the test as given as under Procedure on these four standard concentrations in duplicate or
greater in include negative control consisting of water BET.
5.6.3 Calculate the average of the logarithms of the lowest concentration of endotoxin in each series of
dilutions for which a positive result is found. The antilogarithm of this average gives the estimated
lysate sensitivity, which must be greater than or equal to 0.5λ and less than or equal to 2.0λ.
Confirm the labeled sensitivity of each new batch of lysate prior to use in the test. Test CSE results
are only valid when the positive water and specimen controls are positive at the 2 λ & λ endotoxin concentration.
5.7 Preparation of LAL Reagents:
5.7.1 Test LAL reagent for its sensitivity at the time of receiving. For reconstitution, Collect LAL powder into the bottom of the vial by tapping on a firm surface, unseal and release the vacuum by slowly lifting the
stopper, avoiding touch contamination. The small amount of LAL on the stopper is insignificant.
5.7.2 Rehydrate the reagent with appropriate volume of (as stated on label and certificate) LAL reagent water
by pipetting directly into the vial immediately before use. Cover the vial with an endotoxin free surface or
the inner side of perafilm when not in immediate use. Gently swirl until LAL dissolves into a colorless
solution.
5.8 Test procedure for gel clot:
5.8.1 Each assay should include dilutions of a sample or product, a negative water control, a positive
product control, and either a 2λ water control or a series of two-fold dilutions from an endotoxin
standard which bracket the labeled LAL sensitivity.
5.9 Interpretation of results:
5.9.1 Each tube in the gel-clot method is interpreted as either positive or negative. A positive test is defined as
formation of a firm gel capable of maintaining its integrity when the test tube is inverted 180°C. A negative
test is characterized by the absence of gel or by the formation of a viscous mass, which does not hold
when the tube is inverted. Calculate sample endotoxin concentration by the following formula.
Sample Endotoxin Conc. = Dilution Factor x Sensitivity of λ/ Conc. of Sample (Potency)
5.10 Re-tests:
5.10.1If the positive result is found for one of the test duplicates and a negative result for the other, the test
should be repeated as described above. The results of the retest should be interpreted as for the initial
test.
5.10.2The substance or preparation being examined meets the requirements of the test if the concentration of
endotoxin is less than the endotoxin limit stated in the monograph.
5.11 PRECAUTIONS:
5.11.1Vortex standard Endotoxin for not less than 30 minutes by taping vial to vortex mixture.
5.11.2Vortex all Endotoxin dilution for not less than 1 minute.
5.11.3Do not allow dilution of Endotoxin to stand for more than 10 minutes without re-vortexing
5.11.4Reconstitute Lysate very gently and mix carefully so as to avoid the formation of bubbles.
5.11.5Incubate tube in heating block kept on a stable surface, which do not get vibration, shake or bump.
5.11.6Do not transfer or disturb the tube at any stage of incubation till reading results.
5.11.7Observe result by inverting the tube very gently with constant speed.
5.11.8Observe result only once. Gel is fragile & may break when handled.
5.11.9Start the heating block before around 30 minutes before starting the test to get the temperature stabilize
at the time of incubation.
6.0 ABBREVIATIONS
Abbreviation used Extended Form
SOP : Standard Operating Procedure
QA : Quality Assurance
MB : Microbiology Lab
QC : Quality Control
NPC : Negative Product Control
PPC : Product Positive Control
PC : Positive Control
λ : Sensitivity Of LAL Reagent
EU : Endotoxin unit
MVD : Maximum Valid Dilution
LRW : LAL Reagent Water
LAL : Limulus Amoebocyte Lysate
CSE : Control Standard Endotoxin
sop for calibration and validation of micro autoclave
sop for Sterility failure investigation
cleaning and operation of discard autoclave
sop for operation of fogger machine
sop for Biological assay of lactic acid bacillus
sop for preparation of culture inoculum
STP for sterility testing of sterile gloves
sop for Operation and calibration of active air sampler
sop for transfer of material for testing and sampling in sterile area
entry & exit procedure in microbiology laboratory
Growth Promotion Test In Microbiology Laboratory
Operation of B.O.D in Microbiology Laboratory
Operation of Horizontal Laminar Air Flow in the microbiology laboratory
Operation and cleaning of Pass Box.
Operation and cleaning of air sampler
Cleaning and Sterilization of Glassware
Analysis of water for microbial load in microbiology lab
Operation and temperature monitoring of Refrigerator
Fumigation of Microbiology Laboratory.
Entry & Exit procedure In Sterility Area
SOP for Microbial analysis of Raw Material Finished Products
SOP for Operation & Calibration of pH meter in Micro Department
SOP for Operation & Calibration of pH meter in Micro Department
SOP Operation and calibration of Hot Air Oven In Microbiology
SOP for operation cleaning & calibration of Digital colony counter
SOP for Operation And Cleaning of Microscope
sop for Media Preparation and Consumption
sop for Receipt Storage and Usage of Culture Media
sop for Cleaning Sanitization And Disinfection In Microbiology
sop for Environmental monitoring of all the Classified area
sop for Handling and Sub culturing of Microbial cultures
sop for Media Growth Promotion Test and various Microbiological test
sop for BOD incubator operation and cleaning
sop sampling of water for microbiological analysis
sop for Disinfectant Efficacy Test
sop for for cleaning and operation of vortex mixture
sop for Temperature & Relative Humidity Monitoring
sop for Operation and Calibration of Heating Block
sop for Sterility Testing of Microbiology
sop for Disposal of Culture Media
sop for Drain point of Microbiology
sop for entry & exit procedure In Microbial limit test and Biosafety
sop for Gram Staining of Bacteria in Microbiology Laboratory
sop for Monitoring of Compressed Air/gases for microbiological analysis
sop for BET (Bacterial Endotoxin) test in Microbiology
sop for receipt storage and Determining the population of Biological indicators
sop for qualification of analyst microbiologist
sop for Bioburden test of Packing materials in Microbiology Laboratory
sop for microbiological assay of erythromycin antibiotic
sop for liquid particle counter
sop for operation and calibration of digital zone reader
sop for monitoring of ultraviolet efficiency LAF and pass box
microbiological assay of cyanocobalamin or vitamin B12
gowning procedure for microbiological testing area
swab testing of various surfaces for bioburden determination
sop for endotoxin challenge test
Hold time study protocol for sterilized media
sop for personnel Qualification protocol for aseptic area
sop for sampling and testing of drain water
Sop for Operation of Airborne Particle Counter
sop for Validation protocol of steam sterilizer autoclave
sop for pathogen detection from drain point
Sop for Analysis of Raw water Purified water water for injection and pure steam water
sop for preservatives efficacy test
sop for collection and preservation of in house isolated microorganisms
sop for Operation Calibration and Maintenance of Micropipette
sop for microbiological testing of water
sop for depyrogenation of apparatus
sop for fertility test growth promotion test of media
sop for Operation and cleaning of moist heat sterilizer
sop for monitoring by active air sampler
sop for swab sampling and testing for clean rooms in production area
sop for monitoring in microbiology laboratory
sop for Fumigation of aseptic area and microbiology lab
sop for monitoring of personnel in aseptic area
sop for maintenance of cultures
sop for Operation and cleaning of laminar bench
sop for monitoring of pure steam
sop for entry and exit procedure to m.l.t and b.e.t room
sop for storage of and use of media
sop for disposal of microbiological media and cleaning of microbiological glassware