Water sampling of raw water purified water WFI and PSG in microbiology Lab

 

Water sampling of raw water, purified water WFI and PSG in microbiology Lab

 

1.0 Objective
To lay down a procedure for water sampling of raw water, purified water, WFI and PSG in microbiology Lab
2.0 Scope
This Standard Operating Procedure is applicable for the water sampling of raw water, purified water, WFI & PSG in microbiology Lab to be followed at formulation plant of abc company.
3.0 Responsibility
3.1 Officer / Executive Quality Control Department are responsible for water sampling of raw water, purified water, WFI and PSG.
3.2 Head Quality Control is responsible for the compliance of this SOP.
4.0 Abbreviations and Definitions
SOP :   Standard Operating Procedure
QA :   Quality Assurance
QC :    Quality Control
No. :    Number
P.W. :   Purified water
R.W. :   Raw water
WFI :   Water for Injection
PSG :   Pure Steam Generation
S.S. :   Stainless Steel
Sec. :   Seconds
Hrs :   Hours
ºC :     Degree Centigrade
IPA :   Isopropyl Alcohol
5.0 Procedure
5.1 Water sampling for R.W. and P.W.
5.1.1 For sampling of water R.W. & P.W. glassware should be cleaned as per SOP No. for Cleaning

and Autoclave as per SOP No.
5.1.2 Open the sampling valve in such a way that the water should neither drip nor it should come with full force.
5.1.3 Use autoclave glass bottle for sampling of P.W. and R.W and drain the water from bottle and open

the sampling valve for 30 sec. before sampling.
5.1.4 Holding the base of the bottle fill the water sample from sampling point in such a way that the

bottle mouth should not touch the sampling port during the water sampling and water should not over flow from the bottle.
5.1.5 Replace the cap immediately after sampling and place the bottle in the same SS container/ plastic

bucket and close the lid of the SS container/ bucket.
5.1.6 Similarly sample the water from all respective sampling locations and then transfer the

SS container/bucket to the Microbiology laboratory for microbial analysis.
5.1.7 Analyze the sample as quickly as possible on arrival at the laboratory. Begin the analysis

within 3 Hours of sampling. If immediate analysis is not possible store the sample under refrigeration

at 2ºC to 8ºC for 6 hours.
5.2 Water sampling for WFI and PSG
5.2.1 For water sampling of WFI & PSG, glassware should be cleaned as per SOP No. and Depyrogenate as per SOP No.
5.2.2 Open the sampling valve in such a way that the water should neither drip nor it should come with full force.
5.2.3 Use depyrogenated glass bottle/tubes/flask for sampling of W.F.I and P.S.G and drain the water

sample for 30 sec before sampling.
5.2.4 Holding the bottle/tubes/flask with the help of holder/ tray of water sample from sampling point

in such a way that the bottle mouth should not touch the sampling port during the water sampling and

water should not over flow from the bottle/tubes/flask.
5.2.5 Replace the cap/aluminum foil immediately after sampling and place the bottle in the same

SS container/ plastic bucket and close the lid of the SS container/ bucket.
5.2.6 Similarly sample the water from all respective sampling locations and then transfer the

SS container/bucket to the Microbiology laboratory for microbial analysis.
5.2.7 Analyze the sample as quickly as possible on arrival at the laboratory. Begin the analysis

within 3 hrs of sampling. If immediate analysis is not possible store the sample under refrigeration

at 2ºC to 8ºC for 6 hours.
5.3 Precaution
5.3.1 Follow the gowning procedure at the respective area at the time of water sampling.
5.3.2 Wear the hand gloves, Nose mask at the time of water sampling.
5.3.3 Disinfectant the hand gloves with 70% I.P.A and wait till the evaporation of the I.P.A before

the water sampling.
6.0 Forms and Records
6.1 None
7.0 Distribution
7.1 Master copy – Documentation Cell (Quality Assurance)
7.2 Control copy – Quality Control

8.0 History

Revision Number Details For Change
Reason for Revision
00 New SOP NA

 

sop for calibration and validation of micro autoclave

sop for Sterility failure investigation

cleaning and operation of discard autoclave

sop for operation of fogger machine

sop for Biological assay of lactic acid bacillus

sop for preparation of culture inoculum

STP for sterility testing of sterile gloves

sop for Operation and calibration of active air sampler

sop for transfer of material for testing and sampling in sterile area

entry & exit procedure in microbiology laboratory

Growth Promotion Test In Microbiology Laboratory

Operation of B.O.D in Microbiology Laboratory




Operation of Horizontal Laminar Air Flow in the microbiology laboratory

Operation and cleaning of Pass Box.

Operation and cleaning of air sampler

Cleaning and Sterilization of Glassware

Analysis of water for microbial load in microbiology lab

Operation and temperature monitoring of Refrigerator

Fumigation of Microbiology Laboratory.

Entry & Exit procedure In Sterility Area

SOP for Microbial analysis of Raw Material Finished Products

SOP for Operation & Calibration of pH meter in Micro Department

SOP for Operation & Calibration of pH meter in Micro Department

SOP Operation and calibration of Hot Air Oven In Microbiology

SOP for operation cleaning & calibration of Digital colony counter

SOP for Operation And Cleaning of Microscope

sop for Media Preparation and Consumption

sop for Receipt Storage and Usage of Culture Media

sop for Cleaning Sanitization And Disinfection In Microbiology

sop for Environmental monitoring of all the Classified area

sop for Handling and Sub culturing of Microbial cultures

sop for Media Growth Promotion Test and various Microbiological test

sop for BOD incubator operation and cleaning

sop sampling of water for microbiological analysis

sop for Disinfectant Efficacy Test

sop for for cleaning and operation of vortex mixture

sop for Temperature & Relative Humidity Monitoring

sop for Operation and Calibration of Heating Block

sop for Sterility Testing of Microbiology

sop for Disposal of Culture Media

sop for Drain point of Microbiology

sop for entry & exit procedure In Microbial limit test and Biosafety

sop for Gram Staining of Bacteria in Microbiology Laboratory

sop for Monitoring of Compressed Air/gases for microbiological analysis

sop for BET (Bacterial Endotoxin) test in Microbiology

sop for receipt storage and Determining the population of Biological indicators

sop for qualification of analyst microbiologist

sop for Bioburden test of Packing materials in Microbiology Laboratory

sop for microbiological assay of erythromycin antibiotic

sop for liquid particle counter

sop for operation and calibration of digital zone reader

sop for monitoring of ultraviolet efficiency LAF and pass box

microbiological assay of cyanocobalamin or vitamin B12

gowning procedure for microbiological testing area

swab testing of various surfaces for bioburden determination

sop for endotoxin challenge test

Hold time study protocol for sterilized media

sop for personnel Qualification protocol for aseptic area

sop for sampling and testing of drain water

Sop for Operation of Airborne Particle Counter

sop for Inoculum Preparation

sop for Validation protocol of steam sterilizer autoclave

sop for pathogen detection from drain point

Sop for Analysis of Raw water Purified water water for injection and pure steam water

sop for preservatives efficacy test

sop for collection and preservation of in house isolated microorganisms

sop for Operation Calibration and Maintenance of Micropipette

Sop for UV Efficacy Test

sop for gram staining

Sop for swab testing

sop for microbiological testing of water

PROCEDURE FOR FUMIGATION

sop for depyrogenation of apparatus

sop for media preparation

sop for fertility test growth promotion test of media

sop for Operation and cleaning of moist heat sterilizer

sop for monitoring by active air sampler

sop for swab sampling and testing for clean rooms in production area

sop for monitoring in microbiology laboratory

sop for Fumigation of aseptic area and microbiology lab

sop for monitoring of personnel in aseptic area

sop for maintenance of cultures

sop for Operation and cleaning of laminar bench

preparation of settle plates

sop for monitoring of pure steam

sop for entry and exit procedure to m.l.t and b.e.t room

sop for storage of and use of media

sop for disposal of microbiological media and cleaning of microbiological glassware

Preparation Filtration Usage and Destruction of Disinfectant Solution

Entry Exit and gowning Procedure in Microbiology Laboratory

Safety measures to be followed in microbiology laboratory

Growth Promotion Test of Media

Preservation and Maintenance of Microbial Strain

Entry and Exit in Microbiology Sterile Area

Entry and Exit in Microbiology Limit testing area

Water sampling of raw water purified water WFI and PSG in microbiology Lab

 

 

 