Operation and Calibration of Air Sampler

 

Operation and Calibration of Air Sampler

 

1.0 Objective
To lay down a procedure for Operation and calibration of Air sampler.
2.0 Scope
This Standard Operating Procedure at Microbiology laboratory Quality control department of abc company.
3.0 Responsibility
3.1 Concerned Officer/Executive shall be responsible to follow this SOP.
3.2 Department Head or his designee shall be responsible to implement and ensuring to compliance of this SOP.
4.0 Abbreviations and Definitions
SOP : Standard Operating Procedure
QC : Quality Control
QA : Quality Assurance
SD : Standard deviation
Cal : Calibration
5.0 Procedure
5.1 Operation and calibration
5.1.1 Connect the power cable to instrument and power supply, Recharge the instrument for an one hour before for every usage.
5.1.2 Switch ON the instrument by once single pressing on the ‘ON/OFF’ button.
5.1.3 Set the time for 1000 liters by using ‘Set’ and ‘Arrow’ buttons.
5.1.4 Carry the sterile SS stain cup and Petri plates in a sterile condition to area where the Air sampling is scheduled.
5.1.5 Open the SS stain cup, and assemble the Petri plate in a direction to open side of air flow.
5.1.6 Prepare the media and Petri plate.
5.1.7 Once the Petri plates are assembled, press the start button.
5.1.8 After 1000 liters completed air vacuum pump will gets stops automatically.
5.1.9 Open the SS stain cup and remove the Petri plate from the instrument, close the Petri plate with upper part of Petri plate immediately.
5.1.10 Carry the Petri plate in a sterile condition to the Microbiology lab.
5.1.11 Keep the plates for incubation as per respective SOP No BBL/QC/SOP/028.
5.2 Cleaning and Sterilization
5.2.1 Before using the air sampler, wrapped SS stain cup in butter paper and keep it in Autoclave for Sterilization.
5.2.2 Sanitize the tubular metallic handle with 70 % v/v Isopropyl alcohol.
5.2.3 Clean the impeller with 70 % v/v Isopropyl alcohol between sampling cycles.
5.3 Precautions
5.3.1 Do not autoclave the tubular stainless body, it will damage the system.
5.3.2 Ensure the battery power is sufficient for conducting the test well in advance.
5.3.3 Test the power fall of the battery by switching on the control box and note the alarm, which should occur at any given timer selection.
5.3.4 For the charging the battery, bring the DC/AC switch on the neutral position and connect to 220V AC mains, using power cord.
5.3.5 After completion of the work put the air sampler into its carry box.
5.3.6 Calibration shall be performed for Motor /Battery efficiency, and for Air flow rate with calibrated anemometer by authorized out source calibration party once a year and when ever a major maintenance is done.
5.3.7 Acceptance criteria: 180 liters air per minute.
6.0 Forms and Records
6.1 Air Sampler-Usage log book – Annexure-1
7.0 Distribution
7.1 Master copy – Documentation Cell (Quality Assurance)
7.2 Control copy – Quality Control
8.0 History

Revision Number Details For Change
Reason for Revision
00 New SOP NA

sop for calibration and validation of micro autoclave

sop for Sterility failure investigation

cleaning and operation of discard autoclave

sop for operation of fogger machine

sop for Biological assay of lactic acid bacillus

sop for preparation of culture inoculum

STP for sterility testing of sterile gloves

sop for Operation and calibration of active air sampler

sop for transfer of material for testing and sampling in sterile area

entry & exit procedure in microbiology laboratory

Growth Promotion Test In Microbiology Laboratory

Operation of B.O.D in Microbiology Laboratory




Operation of Horizontal Laminar Air Flow in the microbiology laboratory

Operation and cleaning of Pass Box.

Operation and cleaning of air sampler

Cleaning and Sterilization of Glassware

Analysis of water for microbial load in microbiology lab

Operation and temperature monitoring of Refrigerator

Fumigation of Microbiology Laboratory.

Entry & Exit procedure In Sterility Area

SOP for Microbial analysis of Raw Material Finished Products

SOP for Operation & Calibration of pH meter in Micro Department

SOP for Operation & Calibration of pH meter in Micro Department

SOP Operation and calibration of Hot Air Oven In Microbiology

SOP for operation cleaning & calibration of Digital colony counter

SOP for Operation And Cleaning of Microscope

sop for Media Preparation and Consumption

sop for Receipt Storage and Usage of Culture Media

sop for Cleaning Sanitization And Disinfection In Microbiology

sop for Environmental monitoring of all the Classified area

sop for Handling and Sub culturing of Microbial cultures

sop for Media Growth Promotion Test and various Microbiological test

sop for BOD incubator operation and cleaning

sop sampling of water for microbiological analysis

sop for Disinfectant Efficacy Test

sop for for cleaning and operation of vortex mixture

sop for Temperature & Relative Humidity Monitoring

sop for Operation and Calibration of Heating Block

sop for Sterility Testing of Microbiology

sop for Disposal of Culture Media

sop for Drain point of Microbiology

sop for entry & exit procedure In Microbial limit test and Biosafety

sop for Gram Staining of Bacteria in Microbiology Laboratory

sop for Monitoring of Compressed Air/gases for microbiological analysis

sop for BET (Bacterial Endotoxin) test in Microbiology

sop for receipt storage and Determining the population of Biological indicators

sop for qualification of analyst microbiologist

sop for Bioburden test of Packing materials in Microbiology Laboratory

sop for microbiological assay of erythromycin antibiotic

sop for liquid particle counter

sop for operation and calibration of digital zone reader

sop for monitoring of ultraviolet efficiency LAF and pass box

microbiological assay of cyanocobalamin or vitamin B12

gowning procedure for microbiological testing area

swab testing of various surfaces for bioburden determination

sop for endotoxin challenge test

Hold time study protocol for sterilized media

sop for personnel Qualification protocol for aseptic area

sop for sampling and testing of drain water

Sop for Operation of Airborne Particle Counter

sop for Inoculum Preparation

sop for Validation protocol of steam sterilizer autoclave

sop for pathogen detection from drain point

Sop for Analysis of Raw water Purified water water for injection and pure steam water

sop for preservatives efficacy test

sop for collection and preservation of in house isolated microorganisms

sop for Operation Calibration and Maintenance of Micropipette

Sop for UV Efficacy Test

sop for gram staining

Sop for swab testing

sop for microbiological testing of water

PROCEDURE FOR FUMIGATION

sop for depyrogenation of apparatus

sop for media preparation

sop for fertility test growth promotion test of media

sop for Operation and cleaning of moist heat sterilizer

sop for monitoring by active air sampler

sop for swab sampling and testing for clean rooms in production area

sop for monitoring in microbiology laboratory

sop for Fumigation of aseptic area and microbiology lab

sop for monitoring of personnel in aseptic area

sop for maintenance of cultures

sop for Operation and cleaning of laminar bench

preparation of settle plates

sop for monitoring of pure steam

sop for entry and exit procedure to m.l.t and b.e.t room

sop for storage of and use of media

sop for disposal of microbiological media and cleaning of microbiological glassware

Preparation Filtration Usage and Destruction of Disinfectant Solution

Entry Exit and gowning Procedure in Microbiology Laboratory

Safety measures to be followed in microbiology laboratory

Growth Promotion Test of Media

Preservation and Maintenance of Microbial Strain

Entry and Exit in Microbiology Sterile Area

Entry and Exit in Microbiology Limit testing area

Water sampling of raw water purified water WFI and PSG in microbiology Lab

sop of Microbiological assay

Microbial monitoring of Drain points

sop for Operation of Double Door Autoclave

Operation and Calibration of Thermo block

Operation and Calibration of Micro pipette

Operation and calibration of BOD-Incubators

Operation and Calibration of Microscope

Operation and Calibration of Air Sampler

 

Leave a Comment

error: Content is protected !!